SPECIFICATIONS
1. Must fit within a 4-inch diameter conduit.
2. Must be able to measure turbidity range of 0.001 – 4,000 NTU, with output in NTU.
3. Accuracy 1% of reading or 0.001 NTU, whichever is greater.
4. Measurement frequency capability of at least 1 measurement per 5 minutes.
5. Must be able to operate in temperatures from 0 to 40°C.
6. Must have stainless steel or other corrosion- and oxidation-proof body.
7. Available data communications cable of at least 50 m.