Statement_of_Work-_01072019.docx

DOCX document 18 KB Posted

Attached to
Spectrograph Imaging System Federal contract opportunity
Solicitation number
FA8601-19-Q-A025
Issued by
Department of the Air Force Materiel Command Lifecycle Management Center Wright Patterson Air Force Base

About this file

SOW

View the file

Other files for this federal contract opportunity

Other files attached to Spectrograph Imaging System, newest first.
File Type Posted
Combined_Solicitation_Synopsis.docx DOCX document
Solicitation_-_FA860119QA025.pdf PDF
Attachment_2_-_Offeror_Reps_and_Certs.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work 500 mm Spectrograph Imaging System 7 January 2019

1. General: AFRL/RYMH Requires a 500 mm Czerny-Turner Imaging Spectrograph which is compatible with a Princeton Instruments PI-MAX 4 ICCD camera, has a fiber coupled adjustable slit entrance and has 6 gratings to be specified below.

2. Period of Performance: Delivery of all items shall be made no later than 45 days after contract award.

3. Building 620 address: 2241 Avionics Circle, Wright-Patterson AFB, OH 45433

4. Specific Requirements:

a. Minimum requirement: Brand Name Princeton Instruments HRS 500 Imaging Spectrograph System or Equal:

i. Focal Length of 500 mm

ii. Single Entrance Fiber coupled slit with 2 fibers:

1. Single leg fiber (SMA to vertical ferrule)

2. Multiple leg fiber (at least 3 SMA to one vertical ferrule)

iii. Single Exit with:

1. PI-MAX 4 compatible ICCD mount

iv. Acquisition software which is controllable using they Python programming language

v. Two grating turrets with 6 total gratings with the following characteristics:

1. Ruled Grating 1200 G/mm 750 nm Blaze (Turret 1)

2. Ruled Grating 1200 G/mm 500 nm Blaze (Turret 1)

3. Ruled Grating 1200 G/mm 300 nm Blaze (Turret 1)

4. Ruled Grating 150 G/mm 800 nm Blaze (Turret 2)

5. Ruled Grating 150 G/mm 500 nm Blaze (Turret 2)

6. Ruled Grating 150 G/mm 300 nm Blaze (Turret 2)

vi. Calibration lamp or lamps, preferable NIST traceable, capable of calibrating the spectrograph system in terms of absolute wavelength as well as relative intensity.

vii. Delivery Instructions: Complete spectrograph system delivered to Building 620. Delivery made Monday-Friday 0800-1500 hours, no federal holidays

5. Point of Contact: James Caplinger (937) 713-8550.