Purchase_Description.pdf
PDF 565 KB Posted
- Attached to
- Portable Automatic Test Equipment Calibrator (PATEC) Federal contract opportunity
- Solicitation number
- FA2263-17-Q-0028
About this file
Purchase Description
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Schedule_B_PATEC.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
17S-302A-OE
26 Jul 2017
Rev 0
Attachment 1 1
AFMETCAL
PURCHASE DESCRIPTION
FOR
PORTABLE AUTOMATIC TEST EQUIPMENT CALIBRATOR
1. Purpose
This purchase description contains descriptions of the equipment contained in the Portable Automatic Test
Equipment Calibrator (PATEC) standards set used to calibrate automatic test equipment.
2. Equipment List and Potential Suppliers
The following list shows the quantities of each item required per PATEC and identifies potential suppliers of each item.
Item Per PATEC Potential Sources of Supply Oscilloscope Calibrator 1 Fluke
Universal Counter / Rubidium 1 Spectracom
Angle Position Indicator 1 Clarke-Hess, North Atlantic
Power Meter Range Calibrator 1 Keysight
RF Microwave Synthesizer 1 Anritsu
Meter Calibrator 1 Fluke
TMA Controller 1 Dell, Gateway, Toshiba
Microwave Power Splitter 1 Keysight, Narda, Weinschel
Coaxial Step Attenuator 1 Keysight, Weinschel
10 dB Fixed Attenuator 1 Keysight, Narda, Weinschel
Operating Cases 2 ECS
TMA Case 1 Pelican
IEEE-488 cables 1 meter 2 L-Com, Keysight
IEEE-488 cables 2 meter 2 L-Com, Keysight
IEEE-488 cables 4 meter 2 L-Com, Keysight
GPIB-USB Interface Adapter 2 National Instruments
IEEE-488 Reversing Adapter 2 L-com, Keysight
BNC cables 36 inch 2 ITT Pomona
BNC cables 72 inch 2 ITT Pomona
Attachment 1 2
3. Limitation of Specifications and Salient Characteristics
Any features or additional capabilities of the brand name products listed in this purchase description which are not essential to meet the characteristics in this purchase description are not necessary and need not be provided.
The products listed in this purchase description must meet or exceed the following 1 year measurement uncertainties with a 95% confidence level and when operated within the operation temperature of 15 to 35 ºC. The uncertainties supplied with the documentation shall be calculated using a 95% confidence level (k = 2).
4. Meter Calibrator
4.1 Scope
This section of the purchase description describes a very high accuracy DC, AC, and Resistance calibration system.
4.2 General Description
The instrument described herein shall be Brand Name or equal to the following:
Fluke 5730A
Option: 5730A/03 Wideband
Accessories: Fluke 5700A-2083 Rack mount 19 inch, Qty 2
4.3 Electrical Performance Requirements
4.3.1 DC Voltage
4.3.1.1 Range, Resolution, Accuracy
The instrument resolution on each range shall be equal or better than the following specifications. The output voltage, within ±5 degrees C of the calibration temperature, shall be less than or equal to the following specifications:
Range Resolution Accuracy ±(ppm output + µV)
220 mV 10 nV 7.5 + 0.4
2.2 V 100 nV 5 + 0.7
11 V 1µV 3.5 + 2.5
22 V 1 µV 3.5 + 4
220 V 10 µV 5 + 40
1100 V 100 µV 6.5 + 400
4.3.2 AC Voltage
4.3.2.1 Range and Resolution
Range Resolution
2.2 mV 1 nV
22 mV 10 nV
220 mV 100 nV
Attachment 1 3
2.2 V 1 µV
22 V 10 µV
220 V 100 µV
1100 V 1 mV
4.3.2.2 Voltage Accuracy
The output voltage accuracy of the instrument, including but not limited to the effects of temperature, load regulation, noise, distortion, and line regulation, shall be equal to or better than the following specifications when operated within ±5 degrees C of the calibration temperature. The Voltage Ranges refer to the ranges of the instrument.
Range Frequency Accuracy: (ppm of output + V/mV) ppm V
2.2 and 10 to 20 Hz 240 4
22 mV 20 to 40 Hz 90 4
40 Hz to 20 kHz 80 4
20 to 50 kHz 200 4
50 to 100 kHz 500 5
100 to 300 kHz 1050 10
300 to 500 kHz 1400 20
500 kHz to 1 MHz 2700 20
220 mV 10 to 20 Hz 240 12
20 to 40 Hz 90 7
40 Hz to 20 kHz 57 7
20 to 50 kHz 120 7
50 to 100 kHz 310 17
100 to 300 kHz 655 20
300 to 500 kHz 1400 25
500 kHz to 1 MHz 2700 45
2.2 V 10 to 20 Hz 240 40
20 to 40 Hz 90 15
40 Hz to 20 kHz 42 8
20 to 50 kHz 67 10
50 to 100 kHz 85 30
100 to 300 kHz 336 80
300 to 500 kHz 1000 200
500 kHz to 1 MHz 1700 300
22 V 10 to 20 Hz 240 400
20 to 40 Hz 90 150
40 Hz to 20 kHz 42 50
20 to 50 kHz 67 100
22V (cont.) 50 to 100 kHz 83 200
100 to 300 kHz 254 600
300 to 500 kHz 1000 2000
500 kHz to 1 MHz 1500 3200
Range Frequency ppm mV
Attachment 1 4
220 V 10 to 20 Hz 240 4
20 to 40 Hz 90 1.5
40 Hz to 20 kHz 52 0.6
20 to 50 kHz 80 1
50 to 100 kHz 150 2.5
100 to 300 kHz 900 16
300 to 500 kHz 4400 40
500 kHz to 1 MHz 8000 80
1100 V (250 V maximum output @
15 to 50 Hz)
15 to 50 Hz 300 16
50 Hz to 1 kHz 70 3.5
4.3.2.3 Frequency Range and Accuracy
Range: 10 Hz to 1.1999 MHz
Wideband, 10 Hz to 30 MHz
Accuracy: 0.01%
4.3.3 DC Current
4.3.3.1 Range, Resolution, Accuracy
Range Resolution Accuracy (ppm of output + nA/A) ppm nA
220 A 0.1 nA 40 6
2.2 mA 1 nA 35 7
22 mA 10 nA 35 40
220 mA 0.1 µA 45 0.7A
2.2 A 1 µA 80 12A
11 10 µA 360 480 µA
4.3.4 AC Current
4.3.4.1 Range and Resolution
Range Resolution
220 A 1 nA
2.2 mA 10 nA
22 mA 100 nA
220 mA 1 µA
2.2 A 10 µA
11A 100µA
4.3.4.2 Accuracy
Attachment 1 5
The output current accuracy of the instrument, including but not limited to the effects of temperature, load regulation, noise, distortion, and line regulation, shall be equal to or better than the following specifications when operated within ±5 degrees C of the calibration temperature. The Current Ranges refer to the ranges of the instrument and NOT the range of currents over which the specification applies:
Range Frequency Accuracy: (ppm of output + nA/A) ppm nA
220 A 10 to 20 Hz 250 16
20 to 40 Hz 160 10
40 Hz to 1 kHz 103 8
1 to 5 kHz 280 12
5 to 10 kHz 1100 65
2.2 mA 10 to 20 Hz 250 40
20 to 40 Hz 160 35
40 Hz to 1 kHz 103 35
1 to 5 kHz 200 110
5 to 10 kHz 1100 650
22 mA 10 to 20 Hz 250 400
20 to 40 Hz 160 350
40 Hz to 1 kHz 103 350
1 to 5 kHz 200 550
5 to 10 kHz 1100 5000 ppm A
220 mA 10 to 20 Hz 250 4
20 to 40 Hz 160 3.5
40 Hz to 1 kHz 103 2.5
1 to 5 kHz 200 3.5
5 to 10 kHz 1100 10
2.2 A 20 Hz to 1 kHz 244 35
1 to 5 kHz 450 80
5 to 10 kHz 7000 160
4.3.5 Wideband Capability
This output port shall produce AC voltages over frequencies from 10 Hz to 30 Mhz. The output of the instrument, when in the wideband mode, shall be programmable from the front panel or from the I/O interface in volts or dBm
(0dBm = 1 mW into 50 ohms).
4.3.5.1 Range and Resolution
Range: 3 mV to at least 3.5 volts RMS
Resolution: 5 decades of resolution on all ranges
4.3.5.2 Accuracy
Attachment 1 6
The output voltage or dBm indications at 1 KHz into a precision 50 ohm load shall be accurate to within the following limits:
4.3.5.2.1 Wideband Amplitude
Range: 0 to 3.5 V, 30 Hz to 500 kHz
Range Accuracy: (% of output +
V)
1.1 mV 0.8 2 3 mV 0.7 3
11 mV 0.7 8 33 mV 0.6 16 110 mV 0.6 40
330 mV 0.5 100
1.1 V 0.5 400
3.5 V 0.4 500
4.3.5.2.2 Wideband Frequency Response
Range: 10 Hz to 30 MHz
Accuracy: (% of output + V) Amplitude Flatness, 1 kHz Reference
Frequency Voltage Range
(1.1 mV) (3 mV) (>3 mV)
Hz % V % V % V
10 to 30 0.3 0.3 0.3
30 to 120 k 0.1 0.1 0.1
120 k to 2 M 0.2 3 0.1 3 0.1 3
2 M to 12 M 0.4 3 0.3 3 0.2 3
12 M to 20 M 0.6 3 0.5 3 0.4 3
20 M to 30 M 1.5 15 1.5 3 1 3
4.3.6 Resistance
4.3.6.1 Range and Accuracy
Range: Accuracy: Two-Wire Adder (m)
() (ppm) Im = Current produced by
Ohmmeter
0 40 0.1 Lead Res 1 Lead Res
1 95 2 + 4 V/Im 4 + 4 V/Im
1.9 95 2 + 4 V/Im 4 + 4 V/Im
10 23 2 + 4 V/Im 4 + 4 V/Im
19 23 2 + 4 V/Im 4 + 4 V/Im
100 10 2 + 4 V/Im 4 + 4 V/Im
Attachment 1 7
190 10 2 + 4 V/Im 4 + 4 V/Im
1 k 6.5 2 + 4 V/Im 4 + 4 V/Im
1.9 k 6.5 10 15
10 k 6.5 10 15
19 k 6.5 50 60
100 k 8.5 100 120
190 k 8.5
1 M 13
1.9 M 18
10 M 40
19 M 47
100 M 100
4.3.7 Artifact Calibration
Calibration of the meter calibrator to full specified absolute uncertainty on all functions shall require only the following standards:
Standard
(nominal value)
Absolute Uncertainty
±(ppm)
1 ohm resistor 10.0
10 Kohm resistor 4.0
10 Volt 1.5
4.4 General Requirements
4.4.1 Interface Capability
The instrument shall be equipped with an IEEE-488.2-1992 interface. The interface shall incorporate all of the
IEEE-488.2 Common Commands and Queries. Any function that can be entered via the front panel shall be controllable via the interface, with the exception of power on/off. In addition, any or all of the functions which can be displayed, can be queried and sent over the interface. The GPIB Address shall be selectable from a system menu.
4.4.2 Dimensions
The instrument's dimensions shall not exceed:
Length: 22.75 inches; Height: 7 inches maximum; Width: Instrument plus rack mounting shall fit within a standard
EIA 19 inch rack.
4.4.3 Physical Mounting
The item shall be furnished with accessories necessary for mounting in a standard 19 inch Electronic Industries
Association (EIA) rack.
5. Power Meter Range Calibrator
5.1 Scope
Attachment 1 8
This section of the purchase description describes a microwave power meter range calibrator standard.
5.2 General Description
Keysight Range Calibrator Model 11683A
5.3 Electrical Performance Requirements
5.3.1 Range Calibrator Functions
The range calibrator functions outputs shall correspond to power meter readings of 3, 10, 30, 100, 300 microwatts and 1, 3, 10, 30, 100 milliwatts.
5.3.2 Calibration Uncertainty
The calibration uncertainty shall not exceed +/-0.25% on all ranges.
5.4 General Requirements
5.4.1 Dimensions
The instrument's dimensions shall not exceed:
Height 4.0 Inches
Width 5.7 Inches
Depth 9.0 Inches
6. RF Microwave Synthesizer
6.1 Scope
This section of the purchase description describes a high accuracy microwave frequency synthesizer.
6.2 General Description
Anritsu MG3693C Microwave Synthesizer with options:
MG3690C/2B 110 dB step attenuator
MG3690C/1B Rack mounting without chassis slides to install unit in a console with mounting shelves
MG3690C/16 High stability, ovenized 10 MHz time base
MG3690C/3 Ultra Low Phase Noise, main band
MG3690C/4 10 MHz to 2.2 GHz RF coverage, Ultra Low Phase Noise Version
Attachment 1 9 and with accessories:
34RKNF50 Ruggedized K(m) to Precision Type N(f) adapter. Qty 2
6.3 Electrical Performance Requirements
6.3.1 Frequency Range and Resolution
Frequency range of 10 MHz to 30 GHz with a minimum resolution of 0.1 Hz
6.3.2 Output Power
The synthesizer shall provide the following minimum output power capability:
+10 to -100 dBm from 10 MHz to 20 GHz
+ 3 to -100 dBm from 20 GHz to 30 GHz
6.3.3 Power Level Switching Time
The power level switching time shall not exceed 30ms with a change in the step attenuator.
6.3.4 Output Level Flatness
Variation of the output levels of the synthesizer shall not exceed +/-1.0 dB for 20 MHz to 30 GHz and +/-1.5 dB below 20MHz.
6.3.5 Output Level Resolution
The output power resolution shall have a minimum resolution of 0.01 dB.
6.3.6 Output On/Off Key
This switch shall enable the operator or remote program to toggle the RF output between an Off and On state.
During the Off state, the RF oscillator is turned off. The state shall be indicated by two LEDs located below the
Output On/Off key on the front panel.
6.3.7 Modulation
No modulation capability is required for the synthesizer.
6.3.8 Time Base
The synthesizer shall have an internal 10 MHz time base with an aging rate ≤5 X 10-10 parts per day. Time base variation shall not exceed 2 X 10-10 parts per degree C from 0 to 55 degrees C. The synthesizer shall also be capable of utilizing an external time base. The synthesizer shall have a rear panel BNC 50 Ohm connector capable of accepting a 10 MHz, 0 to + 20 dBm external time base.
6.3.9 Time Base Output
Attachment 1 10
The synthesizer shall have a rear panel type BNC 10 MHz time base output capable of providing at least 1 Vpp, AC coupled, into 50 Ohms.
6.3.10 Harmonics
The harmonic content of the output of the synthesizer shall not exceed the following specifications:
< -30 dBc for 10 MHz to 2 GHz
< -60 dBc for 2 GHz to 20 GHz
< -40 dBc for 20 GHz to 30 GHz
6.3.11 Sub-harmonics and Multiples
All sub-harmonics and multiples of sub-harmonics of the output signal of the synthesizer shall not exceed -60 dBc.
6.3.12 Spurious Outputs
The non-harmonically related spurious emissions shall not exceed -60dBc for RF frequencies from 10 MHz to 30
GHz.
6.3.13 Single Sideband Phase Noise
The single sideband phase noise of the output synthesizer shall not exceed the following specifications:
Frequency Range Frequency Offset (Hz) dBc (Hz) 10 MHz to 1050 MHz 100 -93 1050 MHz to 30 GHz 100 -63 10 MHz to 1050 MHz 1k -118 1050 MHz to 30 GHz 1k -88 10 MHz to 1050 MHz 10k -121 1050 MHz to 30 GHz 10k -98 10 MHz to 1050 MHz 100k -119 1050 MHz to 30 GHz 100k -92
6.4 General Requirements
6.4.1 Interface Capability
The instrument shall be equipped with an IEEE-488.2-1992 interface. The interface shall incorporate all of the
IEEE-488.2 Common Commands and Queries. Any function that can be entered via the front panel shall be controllable via the interface, with the exception of power on/off. In addition, any or all of the functions which can be displayed, can be queried and sent over the interface. The GPIB Address shall be selectable from a system
6.4.2 Dimensions
The synthesizer shall be designed for mounting in a standard 19 inch EIA electronic equipment rack. The depth of the synthesizer from the back of the front panel shall not exceed 22 inches. No controls or fittings shall project more than 1 1/2 inches from the face of the front panel. The height and width shall be as follows:
Height: 8.75 inches; Width: Instrument plus rack mounting shall fit within a standard EIA 19 inch rack
Attachment 1 11
6.4.3 Artifact Calibration
Calibration of the synthesizer shall be accomplished by comparison to external standards traceable to the National
Institute of Standards and Technology (NIST). Any unique accessories including software required for calibration of this item shall be furnished by the manufacturer.
7. Microwave Power Splitter
7.1 Scope
This section of the purchase description describes a microwave power splitter.
7.2 General Description
Weinschel Power Splitter Model 1870A or Keysight Power Splitter Model 11667A
7.3 Electrical Performance Requirements
7.3.1 Frequency Range
The operating frequency range shall equal or exceed DC to 18 GHz.
7.3.2 Impedance
The device input impedance shall be 50 ohms nominal.
7.3.3 Equivalent Output VSWR
The maximum equivalent output VSWR shall not exceed:
DC to 4 GHz 1.10:1
4 GHz to 8 GHz 1.20:1
8 GHz to 18 GHz 1.35:1
7.3.4 Output Tracking
The variance in the tracking between the output arms with change in frequency shall not exceed:
DC to 4 GHz 0.15 dB
4 GHz to 8 GHz 0.20 dB
8 GHz to 18 GHz 0.25 dB
7.3.5 Insertion Loss
The insertion loss from the input to either output shall be nominally 6.0 dB.
Attachment 1 12
7.3.6 Maximum Input Power
The device shall be capable of a continuous input power of +27 dBm (0.5 watt) without loss or degradation of function or accuracy.
7.3.7 Connectors
The connectors shall be precision Type N female on all ports.
8. Coaxial Step Attenuator
8.1 Scope
This section of the purchase description describes a microwave coaxial step attenuator.
8.2 General Description
Keysight Step Attenuator Model 8496B
Option 001: Type N(f) connectors
8.3 Electrical Performance Requirements
8.3.1 Frequency Range
The device operating frequency range shall be from DC to 18 GHz.
8.3.2 Incremental Attenuation
The device incremental attenuation shall be from 0 to 110 dB in 10 dB steps.
8.3.3 Attenuation Accuracy
The attenuation accuracy, referenced to 0 dB, shall be equal to or superior to the following (±dB):
Attenuation (dB) DC to 12.4 GHz 12.4 to 18.0 GHz
10 0.5 0.6
20 0.7 0.8
30 0.9 1.2
40 1.2 1.6
50 1.5 2.0
60 1.8 2.4
70 2.1 2.8
80 2.4 3.2
90 2.7 3.6
100 3.0 4.0
110 3.3 4.4
Attachment 1 13
8.3.4 VSWR Maximum
The VSWR maximum at 50 ohms nominal impedance shall not exceed:
DC to 8.0 GHz 1.5:1
8.0 GHz to 12.4 GHz 1.6:1
12.4 GHz to 18.0 GHz 2.0:1
8.3.5 Power Rating
The device power rating shall permit 1.0 Watt average, and 100 Watts peak without loss of accuracy or function.
8.3.6 Insertion Loss
The maximum insertion loss shall not exceed 0.6 dB + 0.1 dB/GHz at the 0 dB setting.
8.3.7 Connectors
The connectors shall be precision type N female at both ends.
9. 10dB Fixed Attenuator
9.1 Scope
This section of the purchase description describes a microwave 10 dB fixed attenuator.
9.2 General Description
The instrument described herein shall be Brand Name or equal to:
Narda 10dB Fixed Attenuator Model 779 or Keysight 10dB Fixed Attenuator Model 8491B-010
9.3 Electrical Performance Requirements
9.3.1 Attenuation
The attenuation shall be 10.0 dB +/-0.6 dB, bidirectional, over the frequency range of DC to 18 GHz.
9.3.2 Frequency Range
The device operating frequency range shall be DC to 18 GHz.
9.3.3 VSWR Maximum
The maximum VSWR at 50 Ohms nominal impedance shall not exceed:
DC to 4.0 GHz: 1.2:1
4.0 GHz to 12.4 GHz: 1.3:1
12.4 GHz to 18.0 GHz: 1.5:1
Attachment 1 14
9.3.4 Power Handling Capacity
The device shall be capable of dissipating 2 Watts continuous, and 100 watts peak without loss of accuracy or function.
9.3.5 Connectors
The device connectors shall be precision type N stainless steel with a Male Connector on one end and a Female
Connector on the other end.
10. Oscilloscope Calibrator
10.1 Scope
This section of the purchase description defines the requirements of an oscilloscope and digitizer calibration standard.
10.2 General Description
Fluke Oscilloscope Calibrator Model 9500B/3200 with options:
Opt. Name Quantity
9530 Fast Output Head to 3.2 GHz 4
9500-4955 9500 to 4950/DMM Calibration Interface Adapter 1
9500-90 Rack mount kit 1
10.3 Instrument Mainframe
10.3.1 General Description
The unit shall consist of a 3.2 GHz main frame capable of supporting up to 5 channels at the specifications of the active head described below.
10.3.2 Electrical Performance Requirements (See PD paragraphs 10.3.3 through
10.3.3.5.2.2)
10.3.3 3.2 GHz Fast Output Head
This active head shall be capable of supporting the following specifications.
10.3.3.1 Voltage Function
10.3.3.1.1 Amplitude Range
10.3.3.1.1.1 DC into 1 MOhms
Attachment 1 15
The amplitude range of the instrument shall include +/- 1 mV to +/- 200 V.
10.3.3.1.1.2 DC into 50 Ohms
The amplitude range of the instrument shall include +/- 1 mV to +/- 5.0 V.
10.3.3.1.1.3 Square Wave into 1 MOhms
The amplitude range of the instrument shall include 40 uV to 200 V pk-pk.
10.3.3.1.1.4 Square Wave into 50 Ohms
The amplitude range of the instrument shall include 40 uV to 5.0 V pk-pk.
10.3.3.1.2 Accuracy
10.3.3.1.2.1 DC into 1 MOhms
The accuracy of the instrument shall be equal to or superior to +/-(0.025% of setting + 25 uV).
10.3.3.1.2.2 DC into 50 Ohms
The accuracy of the instrument shall be equal to or superior to +/-(0.025% of setting + 25 uV).
10.3.3.1.2.3 Square Wave into 1 MOhms
The accuracy of the instrument shall be equal to or superior to +/-(0.1% of setting + 10 uV) for outputs greater than or equal to 1 mV pk-pk; and +/-(1% of setting + 10 uV) for outputs less than 1 mV pk-pk at frequencies less than or equal to 10 KHz.
10.3.3.1.2.4 Square Wave into 50 Ohms
The accuracy of the instrument shall be equal to or superior to +/-(0.1% of setting + 10 uV) for outputs greater than or equal to 1 mV pk-pk; and +/-(1% of setting + 10 uV) for outputs less than 1 mV pk-pk at frequencies less than or equal to 10 KHz.
10.3.3.1.3 Ranging
10.3.3.1.3.1 DC into 1 MOhms
The Volt/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously variable.
10.3.3.1.3.2 DC into 50 Ohms
The Volt/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously variable.
10.3.3.1.3.3 Square Wave into 1 MOhms
Attachment 1 16
The Volt/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously
10.3.3.1.3.4 Square Wave into 50 Ohms
The Volt/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously variable.
10.3.3.1.4 Deviation
10.3.3.1.4.1 DC into 1 MOhms
The output of the instrument at any given setting shall be adjustable over a range of at least +/-11% including over range and under range.
10.3.3.1.4.2 DC into 50 Ohms
The output of the instrument at any given setting shall be adjustable over a range of at least +/-11% including over
10.3.3.1.4.3 Square Wave into 1 MOhms
The output of the instrument at any given setting shall be adjustable over a range of at least +/-11% including over
10.3.3.1.4.4 Square Wave into 50 Ohms
The output of the instrument at any given setting shall be adjustable over a range of at least +/-11% including over
10.3.3.1.5 Rise/Fall Time
10.3.3.1.5.1 Square Wave into 1 MOhms
The rise and fall times of the square wave output shall be less than 150 nS for voltage outputs less than 100 V pk-pk, and less than 200 nS for voltage outputs greater than or equal to 100 V pk-pk.
10.3.3.1.5.2 Square Wave into 50 Ohms
The rise and fall times of the square wave output shall be less than 150 nS for all available output voltages.
10.3.3.1.6 Aberrations
10.3.3.1.6.1 Square Wave into 1 MOhms
The aberrations of the square wave output shall be less than 2% of output setting for the first 500 nS.
10.3.3.1.6.2 Square Wave into 50 Ohms
Attachment 1 17
The aberrations of the square wave output shall be less than 2% of output setting for the first 500 nS.
10.3.3.1.7 Frequency
The square wave frequency range shall include 10 Hz to 100 KHz.
10.3.3.2 Timing Marker Function
10.3.3.2.1 Square
10.3.3.2.1.1 Period
The periods available shall include 9.0091 nS to 55 S.
10.3.3.2.1.2 Ranging
The Time/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously
10.3.3.2.1.3 Deviation
The instrument shall be capable of adjustment from nominal output of at least +/-45% of setting including over range.
10.3.3.2.1.4 Timing Accuracy
The timing accuracy of the marker shall not exceed +/-0.25 ppm of output for settings less than 83 uS, and shall not exceed +/- 3 ppm of output for settings greater than or equal to 83 uS.
10.3.3.2.1.5 Amplitude
The amplitudes available shall include 100 mV to 1 V pk-pk.
10.3.3.2.2 Sine
10.3.3.2.2.1 Period
The periods available shall include 450.5 pS to 9.009 nS.
10.3.3.2.2.2 Ranging
The Time/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously
10.3.3.2.2.3 Deviation
The instrument shall be capable of adjustment from nominal output of at least +/-45% of setting including over
Attachment 1 18
10.3.3.2.2.4 Timing Accuracy
The timing accuracy of the marker shall not exceed +/-0.25 ppm of output for settings less than 83 uS, and shall not exceed
+/-3 ppm of output for settings greater than or equal to 83 uS.
10.3.3.2.2.5 Amplitude
10.3.3.2.3 Pulse
10.3.3.2.3.1 Period
The periods available shall include 900.91 nS to 55 S.
10.3.3.2.3.2 Ranging
The Time/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously
10.3.3.2.3.3 Deviation
The instrument shall be capable of adjustment from nominal output of at least +/-45% of setting including over
10.3.3.2.3.4 Timing Accuracy
The timing accuracy of the marker shall not exceed +/-0.25 ppm of output for settings less than 83 uS, and shall not
10.3.3.2.3.5 Amplitude
10.3.3.2.4 Narrow Triangle
10.3.3.2.4.1 Period
The periods available shall include 900.91 nS to 55 S.
10.3.3.2.4.2 Ranging
The Time/div scaling shall support all of the following sequences: 1, 2, 5; 1, 2, 2.5, 4, 5; and continuously
10.3.3.2.4.3 Deviation
Attachment 1 19
The instrument shall be capable of adjustment from nominal output of at least +/-45% of setting including over
10.3.3.2.4.4 Timing Accuracy
The timing accuracy of the marker shall not exceed +/-0.25 ppm of output for settings less than 83 uS, and shall not
10.3.3.2.4.5 Amplitude
10.3.3.3 Leveled Sine Function
10.3.3.3.1 Frequency Range
The frequency range available shall include 0.1 Hz to 3.2 GHz.
10.3.3.3.2 Amplitude (pk-pk into 50 Ohms)
The amplitude available shall include 5 mV to 5 V from 0.1 Hz to 550 MHz, 5 mv to 3 V from 550 MHz to 2.5
GHz, and 5 mV to 2 V from 2.5 to 3.2 GHz.
10.3.3.3.3 Accuracy
The amplitude accuracy of the instrument shall be equal or superior to +/-1.5% at single reference frequency (50
KHz to 10 MHz).
10.3.3.3.4 Flatness
The flatness of the output amplitude with respect to the reference frequency amplitude shall not exceed +/-2.0% from 0.1 to 300 MHz, +/-3% from 300 MHz to 550 MHz, +/-4% from 550 MHz to 1.1 GHz, and 5% from 1.1 GHz to 3.2 GHz.
10.3.3.3.5 Harmonic Purity
The 2nd harmonic of any output shall not exceed -35 dBc and the 3rd harmonic of any output shall not exceed -40 dBc in 12 GHz.
10.3.3.3.6 Nonharmonic and Subharmonic Purity
Nonharmonic and subharmonic signals for any output shall not exceed -40 dBc
10.3.3.4 Dual Sine Function
10.3.3.4.1 Frequency Range
The frequency range available shall include 0.1 Hz to 3.2 GHz.
Attachment 1 20
10.3.3.4.2 Time Alignment
The time alignment between any two channels shall not exceed +/-25 pS
10.3.3.5 Input Impedance
10.3.3.5.1 Resistance Measurement
10.3.3.5.1.1 Range
The range of resistances measured shall include 10 Ohms to
12 MOhms.
10.3.3.5.1.2 Accuracy
The accuracy with which the instrument measures the input resistances shall be equal to or superior to +/-0.5% over the entire range.
10.3.3.5.2 Capacitance Measurement
10.3.3.5.2.1 Range
The range of capacitances measured shall include 1 pF to 95 pF.
10.3.3.5.2.2 Accuracy
The accuracy with which the instrument measures the input capacitances shall be equal to or superior to +/-(3% of input+0.25pF) over the entire range.
10.3.4 General Requirements
10.3.4.1 Interface Capability
The instrument shall be equipped with an IEEE-488.2-1992 interface. The interface shall incorporate all of the
IEEE-488.2 Common Commands and Queries. Any function that can be entered via the front panel shall be controllable via the interface, with the exception of power on/off. In addition, any or all of the functions which can be displayed, can be queried and sent over the interface. The GPIB Address shall be selectable from a system
10.3.4.2 Dimensions
The dimensions of the mainframe shall not exceed:
Height : 5.25 Inches Maximum (excluding feet)
Depth : 20.0 Inches Maximum
Width : Instrument plus rack mounting shall fit within a standard EIA 19 inch rack
Attachment 1 21
10.3.4.3 Physical Mounting
The instrument shall be furnished with accessories necessary for mounting in a standard 19 inch Electronic
Industries Association (EIA) rack.
11. Universal Counter / Rubidium
11.1 Scope
This section of the purchase description defines the requirements of a universal counter and rubidium frequency standard used to measure timing and frequency parameters of complex waveforms.
11.2 General Description
Spectracom Universal Counter Model CNT-91R/AF
This particular Model comes with Options:
Opt 71 Rear Panel 0.1, 1, 5, and 10 MHz frequency outputs
Opt 10 Input C 100 MHz to 3.0 GHz
PM9622 Rack Flange Kit for use with supplied front handles
11.3 Electrical Performance Requirements
11.3.1 Channel A and B Specifications
11.3.1.1 Input Sensitivity (X1)
The input sensitivity of channels A and B shall be equal to or better than:
25 mV rms for sinewave inputs over their entire frequency spans
75 mV peak-to-peak pulse at minimum pulse width of 5 nS
11.3.1.2 Dynamic Range (X1)
The dynamic range of channels A and B shall be equal to or greater than:
75 mV to 5.0 V peak-to-peak, to 100 MHz, and
75 mV to 2.5 V peak-to-peak, for channel A inputs greater than 100 MHz
11.3.1.3 Trigger Level
11.3.1.3.1 Range
The trigger level range of channels A and B shall be equal to or greater than:
X1: -5 V to +5 V
X10: -50 V to +50 V
Attachment 1 22
11.3.1.3.1.1 Range Setting (X1)
11.3.1.3.1.1.1 Auto Trigger OFF
The trigger level range with auto trigger off shall be:
Preset: To 0 VDC nominal
Adjustable: From -5 VDC to +5 VDC
11.3.1.3.1.1.2 Auto Trigger ON
The trigger level range with auto trigger on shall be:
Preset: To nominal 50% point of input signal
Adjustable: Nominally between the positive and negative peaks of the input signal
11.3.1.3.2 Resolution
The trigger level resolution for channels A and B shall be equal to or better than:
X1: 10 mV
X10: 100 mV
11.3.1.3.2.1 Accuracy (X1)
The trigger level accuracy of the instrument for channels A and B on X1 attenuation and DC coupling shall be equal to or better than:
+/-(20 mV + 0.5% of reading)
11.3.1.3.2.2 Trigger Level Digital Voltmeter
The instrument shall include a trigger level digital voltmeter, which can measure and display the trigger levels of channels A and B.
11.3.1.3.3 Auto Trigger (X1)
11.3.1.3.3.1 Operation
The auto trigger function of the counter shall cause the counter to automatically select the 10% and 90% rise/fall time trigger points, 50% phase trigger points, or the operator's preset value for inputs to channels A and B.
11.3.1.3.3.2 Frequency Range (50% duty cycle)
The channels A and B auto trigger frequency range for a 50% duty cycle signal input shall be equal to or better than:
DC coupled: 30 Hz to 100 MHz
AC coupled and 50 ohms input impedance: 30 Hz to 100 MHz
AC coupled and 1 Megohm input impedance: 200 KHz to 100 MHz
Attachment 1 23
11.3.1.3.3.3 Minimum signal
The minimum signal input for auto trigger shall not exceed 100 mV rms.
11.3.1.3.3.4 Duty cycle range
The auto trigger duty cycle range shall include from 10% to 90% duty cycles.
11.3.1.3.4 Coupling
The coupling shall be AC or DC switchable.
11.3.1.3.5 Input Impedance
The input impedance of channels A and B shall be independently settable to 1 Megohm shunted by less than 35 pF and 50 ohms, when separate signals are applied to each channel; and 1 Megohms shunted by less than 50 pF, when connected for COMMON A input.
11.3.1.3.6 Attenuator
The instrument shall be capable of X1 and X10 signal attenuation of the input signals to channels A and B.
11.3.1.3.7 Slope
The slope of the trigger for inputs to channels A and B shall be positive and negative, independently selectable for each channel.
11.3.1.3.8 Channel Input
The instrument shall permit the operator to select the inputs to channels A and B as COMMON A or SEPARATE.
11.3.1.3.9 Maximum Operating Level
The instrument shall be capable of accepting signal inputs to channels A and B over a dynamic range of -50 VDC to +50 VDC with the appropriate attenuator setting.
11.3.1.4 Functions
11.3.1.4.1 Frequency
11.3.1.4.1.1 Range
The instrument shall be capable of measuring frequencies from 1 Hz to 300 MHz on channel A and from 1 Hz to
100 MHz on channel B.
11.3.1.4.1.2 Least Significant Digit Displayed
The LSD displayed for frequency measurements shall be equal to or better than:
Attachment 1 24
(Frequency X 1 nS)/Gate Time
11.3.1.4.1.3 Resolution
The attainable resolution of frequency measurements shall be equal to or better than:
+/-(2 X LSD + (1.4 X Frequency X Trigger Error)/Gate Time)
11.3.1.4.1.4 Accuracy
The total errors of the frequency measurements shall not exceed:
+/-(Resolution + (Time Base Error X Frequency))
11.3.1.4.2 Period
11.3.1.4.2.1 Range
The instrument shall be capable of measuring periods from 10 nS to 1000 S.
11.3.1.4.2.2 Least Significant Digit Displayed
The Least Significant Digit (LSD) displayed for period measurements shall be equal to or better than:
(Period X 1 nS)/Gate Time
11.3.1.4.2.3 Resolution
The attainable resolution of period measurements shall be equal to or better than:
+/-(2 X LSD + (1.4 X Period X Trigger Error)/Gate Time)
11.3.1.4.2.4 Accuracy
The total errors of the period measurements shall not exceed:
+/-(Resolution + (Time Base Error X Period))
11.3.1.4.3 Time Interval
11.3.1.4.3.1 Range
The instrument shall be capable of measuring time intervals (TI) from 10 nS to 1000 S.
11.3.1.4.3.2 Least Significant Digit Displayed
The LSD displayed for time interval measurements shall be equal to or better than 1 nS.
11.3.1.4.3.3 Resolution
Attachment 1 25
The resolution of time interval measurements shall be equal to or better than:
+/-(2 X LSD + START Trigger Error + STOP Trigger Error)
11.3.1.4.3.4 Accuracy
The sum of the errors in time interval measurements shall not exceed:
+/-((Resolution) + (Time Base Error X Time Interval) +
(Trigger Level Timing Error)+(2 nS))
11.3.1.4.4 Rise and Fall Time
11.3.1.4.4.1 Range
The instrument shall be capable of rise time and fall time measurements over the range of 20 nS to 10 mS.
11.3.1.4.4.2 Minimum Pulse Height
The minimum pulse height for rise time and fall time measurements shall not exceed 500 mV peak-to-peak.
11.3.1.4.4.3 Minimum Width at Peak of Signal
The minimum width at peak of signal required for rise time and fall time measurements shall not exceed 20 nS.
11.3.1.4.4.4 Duty Cycle Range
The instrument shall be capable of measuring rise time and fall time for input signals with a duty cycle range of between 20% and 80%.
11.3.1.4.4.5 Least Significant Digit Displayed
The LSD displayed for rise time and fall time measurements shall be equal to or better than 1 nS.
11.3.1.4.4.6 Resolution
The resolution of rise time and fall time measurements shall be equal to or better than:
+/-((2 X LSD) + (START Trigger Error) + (STOP Trigger Error))
11.3.1.4.4.7 Accuracy
The sum of the errors in rise time and fall time measurements shall not exceed:
+/-((Time Interval Accuracy)+ (Trigger Level Setting Error at 10% point)+
(Trigger Level Setting Error at 90% point))
11.3.1.4.4.8 Input Mode
Attachment 1 26
The instrument shall automatically set to COMMON A input with 10% and 90% trigger levels.
11.3.1.4.5 Ratio
11.3.1.4.5.1 Range
The instrument shall be capable of ratio A/B measurements at frequencies up to 100 MHz.
11.3.1.4.5.2 Least Significant Digit Displayed
The LSD displayed for ratio measurements shall be equal to:
Ratio/(Higher Frequency X Gate Time)
11.3.1.4.5.3 Resolution
The resolution of ratio A/B measurements shall be equal to or better than:
+/-(LSD + (Higher Frequency Trigger Error X Ratio)/Gate Time)
11.3.1.4.5.4 Accuracy
The accuracy of ratio A/B measurements shall be equal to or better than:
11.3.2 Channel C Specifications
11.3.2.1 Electrical Specifications
11.3.2.1.1 Sensitivity
Channel C of the instrument shall have an input sensitivity equal to or better than:
10 mV rms sine wave up to 1 GHz
100 mV rms sine wave from above 1 GHz to 1.3 GHz
11.3.2.1.2 Dynamic Range
The dynamic range of the signal inputs to channel C shall include:
10 mV to 1 V rms to 1 GHz
100 mV to 1 V rms from above 1 GHz to 1.3 GHz
11.3.2.1.3 Coupling
The input signals to channel C of the instrument shall be AC coupled.
11.3.2.1.4 Impedance
Attachment 1 27
The input impedance of the input to channel C of the instrument shall be 50 ohms nominally with a VSWR of less than 3:1.
11.3.2.1.5 Damage Level
The damage level of channel C of the instrument shall exceed +/-8 V (DC + AC peak). The channel C input shall be fuse protected.
11.3.2.2 Functions
11.3.2.2.1 Frequency
11.3.2.2.1.1 Range
The channel C input of the instrument shall enable it to measure frequencies between the 200 MHz and 2.7 GHz.
11.3.2.2.1.2 Least Significant Digit Displayed
The LSD displayed for channel C frequency measurements shall be equal to or better than:
(Frequency X 1 nS)/Gate Time
11.3.2.2.1.3 Resolution
The attainable resolution of channel C frequency measurements shall be equal to or better than:
+/-(2 X LSD + (1.4 X Frequency X Trigger Error)/Gate Time)
11.3.2.2.1.4 Accuracy
The total errors of the channel C frequency measurements shall not exceed:
+/-(Resolution + (Time Base Error X Frequency))
11.3.2.2.2 Ratio C/A
11.3.2.2.2.1 Range
The instrument shall be capable of Ratio C/A measurements where the input to channel A ranges from 1 to 160
MHz; and the input to channel C ranges from 100 MHz to 2.7 GHz.
11.3.2.2.2.2 Least Significant Digit Displayed
The LSD displayed for ratio measurements shall be equal to:
Ratio/(Higher Frequency X Gate Time)
11.3.2.2.2.3 Resolution
Attachment 1 28
The resolution of ratio C/A measurements shall be equal to or better than:
11.3.2.2.2.4 Accuracy
The accuracy of ratio C/A measurements shall be equal to or better than:
11.4 Rubidium High Stability Timebase and Frequency Reference Outputs
The instrument shall contain a rubidium high stability timebase.
11.4.1 Electrical Performance Requirements
11.4.1.1 Output Frequencies
The instrument shall provide rear panel outputs phase locked to the timebase at 10 MHz, 5 MHz, 1 MHz and 100
KHz.
11.4.1.2 Frequency Characteristics
11.4.1.2.1 Timebase Accuracy/Fractional Frequency Offset
The output frequency shall not change by more than 3 X 10 to the minus tenth per day after a 24 hour warm-up period and where the ambient temperature does not vary by more than +/-5 degrees C over that same period.
11.4.1.3 Phase Noise
The Single Side Band (SSB) phase noise in a 1 Hz Bandwidth shall be less than -140 dBc at 1000 Hz from the carrier frequency.
11.4.1.4 Retrace
The output frequency of the unit shall be within 1 X 10 to the minus 10 of the original set frequency within a 1 hour warm-up period after 24 hours off time. The output frequency of the unit shall be within 5 X 10 to the minus
11 of the original set frequency within a 24 hour warm-up period.
11.4.1.5 Output amplitude
The output amplitude available from all frequency outputs shall exceed 1 V p-p into 50
11.4.2 General Requirements
11.4.2.1 Interface Capability
Attachment 1 29
The instrument shall be equipped with an IEEE-488.2-1992 interface. The interface shall incorporate all of the
IEEE-488.2 Common Commands and Queries. Any function that can be entered via the front panel shall be controllable via the interface, with the exception of power on/off. In addition, any or all of the functions which can be displayed, can be queried and sent over the interface. The GPIB Address shall be selectable from a system
11.4.2.2 Measuring Functions
Measuring functions must include: phase, duty factor, rise/fall-time, peak voltage, frequency, period
11.4.2.3 Self-Tests
The instrument must run a self-test during every power-on or reset.
11.4.2.4 Calibration Requirements
The instrument must be capable of being calibrated with assets already in PMEL inventory. PMEL will perform the calibration.
11.4.2.5 Dimensions
The dimensions of the mainframe shall not exceed:
Height : 5.25 Inches Maximum (excluding feet)
Depth : 20.0 Inches Maximum
Width : Instrument plus rack mounting shall fit within a standard EIA 19 inch rack
11.4.2.6 Physical Mounting
The instrument shall be furnished with accessories necessary for mounting in a standard 19 inch Electronic
Industries Association (EIA) rack.
12. Test Module Adapter (TMA) Controller
12.1 Scope
A laptop controller is required that can efficiently operate under the Microsoft Windows 10 Professional operating system.
12.2 General Description
Controller: A laptop controller is required that can efficiently operate under Microsoft Windows 10 Professional
64 bit operating system. All major components of the system must be commercially available and fully supportable by commercial sources. Where specific brands are identified in this specification, no substitutes are allowed without prior approval from the AFMETCAL. The Controller must meet or exceed Microsoft Windows
10 Professional hardware requirements.
12.2.1 Salient Characteristics
12.2.1.1 Processor: Intel Pentium V or otherwise equivalent or better generation minimum, Attachment 1 30
2.0 GHz clock speed minimum.
12.2.1.2 Memory: Minimum 8 GB system Memory.
12.2.1.3 Input Output (I/O) External Ports, Slots and Connectors:
At minimum, the following types of I/O interface connections must be available on the controller or via port replication:
12.2.1.3.1 USB: Minimum of four integrated USB Ver 2.0 compliant ports.
12.2.1.3.2 Ethernet: One 100/1000 BaseT Ethernet RJ45 port.
12.2.1.3.3 Video: Areo-Capable Graphics card that supports Windows Display Driver Model (WDDM 1.0) with a minimum of 128 MB that supports Direct X 9.
12.2.1.4 Display Panel: Minimum 14.1-inch flat panel LCD Color Display Panel with at least 1600 x 900 resolution for a 16:9 aspect ratio.
12.2.1.5 Hard Drive: 80 GB (Minimum) Hard Disk Drive.
12.2.1.6 DVD-ROM
DVD/RW: at least 8X Speed
12.2.2 Operating System and Software:
All software must run under Microsoft Windows 10 (Professional operating system or greater 64bit, latest version release). All software must run Windows 10 as a non-elevated user. Shall be supplied with Microsoft Word and
Excel, latest release with a non-renewable license.Power, Mechanical & Physical Requirements
12.2.5 Additional Hardware: Additional hardware required to make this system function as identified by this purchase description shall be required.
13. Power, Mechanical & Physical Requirements
13.1 Applicability
The following paragraphs of this section on power, mechanical & physical requirements are applicable to every component of the purchase description to the extent that they can be made to apply, unless expressly overridden or additionally qualified by specifications in the individual device's description. For example, none of the cases described above have shafts, bearings, regulators, or switches as mentioned in the "Design and Construction" paragraph below, but indeed, they shall be designed and constructed of materials that meet the requirements of this specification and in accordance with the best commercial practices.
13.1.1 Stability
All instruments shall maintain accuracy as specified for at least 12 months, unless otherwise specified, when used in a laboratory environment within the environmental specifications of this purchase description.
13.1.2 Input Power
All AC powered instruments of this purchase description shall operate on power of 115/230 VRMS +/-10%, 50/60
Hz +/-5%. Detachable 115V power cords shall be provided with the system and shall meet the requirements of
MIL-PRF-28800F, paragraph 3.5.
Attachment 1 31
13.1.3 Commercial Parts
Commercial parts and materials that are used shall have properties consistent with a precision laboratory instrument.
13.1.4 Design and Construction
The equipment shall be designed and constructed of materials that meet the requirements of this specification and in accordance with the best commercial practices for industry-approved equipment. It shall be designed and manufactured to maximize reliability, maintainability, accuracy of operation, and ease of operation. All parts such as shafts, bearings, regulators, switches, controls, etc, shall have proper clearance and adjustments. They shall work together so that the equipment will supply the rated requirements without unnecessary strains, vibrations, or overheating. They shall be able to withstand the conditions met in shipping, storage, installation, and service. All material and parts of this equipment shall be capable of withstanding the environmental requirements of this specification.
13.1.5 Environmental Requirements
The equipment supplied shall meet the environmental requirements as specified in MIL-PRF-28800F paragraph 3.8 for Class 4 equipment with the following exceptions:
1. All electrical, electronic and electromechanical items shall be capable of operating in an ambient temperature of
15 to 35 degrees C without loss of required accuracy.
2. The equipment shall maintain specified accuracy in an atmosphere of from 20% to 60% relative humidity.
13.1.6 Enclosure Physical Characteristics
The equipment shall meet the enclosure physical characteristics requirements of MIL-PRF-28800F paragraph 3.6, except that paragraph 3.6.4.5 Handles does not apply.
13.2 Protective Covers
All exposed connectors shall be protected by dust covers or caps.
13.3 Safety
The equipment shall conform to MIL-PRF-28800F, paragraph 3.10.
14. Operating Cases
14.1 Scope
This section of the purchase description covers an operating case for the transportation and protection of the instrumentation used in the PATEC. The case shall permit the installation of a variety of rack mount configured equipment and drawers. The equipment shall be operated while installed in the case with the covers removed.
With the covers installed, the case shall provide environmental protection for the enclosed equipment.
Attachment 1 32
14.2 General Description
14.2.1 Case
The operating case described herein shall be Brand Name or equal to the following:
ECS P/N 00112306 Qty 1
ECS P/N 00112307 Qty 1
The above P/Ns include the following P/Ns and options:
7314 7000 Series Rackmount Case
202 Aluminum Drawer 3.47 inch (part of Case 1, P/N 00112306)
204 Aluminum Drawer 6.97 inch (part of Case 2, P/N 00112307)
208 AC Power Strip
209 Support Rails (Two Pairs)
213A Pressure Relief Valve
217 Color Desert Tan (33446)
235 Lid Hangers
236 Storage Pouch – Large Size for Rear Lid
326 Removable Casters with Brakes
All requirements listed are essentially those contained in the undated ECS Catalog for ECS 7314 Case.
14.3 Requirements
14.3.1 Case Body and Covers Design and Construction
14.3.1.1 Style and Dimensions
The case shall consist essentially of a case shell with front and rear removable covers, and lifting handles or rings.
The shell center shall include a frame suitable for mounting standard rack mountable instruments allowing the instruments to be operated in the case when the covers are removed. The frame shall be shock isolated.
The case maximum outside dimensions (excluding hardware or molded ribs) shall be 24 inches (6l0 mm) in width
(W), 30.0 inches (762.0 mm) in depth (D), 30.0 inches in height (H).
The clear inside rack height shall not be less than 24.5 inches.
14.3.1.2 Materials
The case body and covers shall be constructed of fire retardant, molded fiberglass reinforced plastic or other composite materials suitable for the purpose intended and meeting all other performance specifications specified herein.
14.3.1.3 Casters
Casters shall be provided to allow for moving the unit without lifting. The casters shall be removable and shall have a locking “brake” to prevent it from moving during use.
Attachment 1 33
14.3.1.4 Pressure Relief Valve
A simple, manually operated pressure relief valve or automatic pressure relief valve shall be provided to equalize the air pressure inside the case with the outside air pressure. The valve shall be designed, positioned, and recessed on the case to prevent damage and accidental opening. The valve assembly shall be replaceable without opening the case and shall not require the use of special tools.
14.3.1.5 Guides
Guides or alignment indicators shall be provided for ease in aligning the front and rear covers to the case body during closing of the case.
14.3.1.6 Handles
Four heavy-duty, large bail handles shall be provided with stops to hold the bail at right angle to the mounting plate when in carrying position and spring loaded to fold the bail into a down position when released. The handles shall be fitted two on each side of the main body of the case. The handles shall be recessed into the surface of the case body to prevent damage from direct or glancing blows during movement and shipment of the case. The bail of the handles shall have a grip diameter not less than 0.375 inch ((0.94 cm), an inside length not less than 4.5 inches
(11.43 cm) with a minimum clearance of 2.0 inches (5.08 cm) between the bail grip and the mounting plate. The handles shall be installed sufficiently above the center of gravity of the equipment to ensure carrying stability.
Protection shall be provided from all sides. The handles shall be capable of lifting a total case weight of 250 pounds (113.5 kg).
14.3.1.7 Mounting Rack Frame
Mounting Rack Frame Unit
The case shall incorporate an interior shock isolated, heavy duty, equipment mounting rack frame unit permitting the installation of standard EIA , 19-inch (482.6 mm) rack mounting equipment up to the maximum height of the rack frame. The internal shock mounting provisions shall suspend the mounting rack frame to isolate the equipment from the case shell.
14.3.1.7.1 Mounting Rack Frame Clearance
The case body shall have a depth of not less than 22 inches (558.8 mm). The front and rear covers shall have a depth of not less than 2.50 inches (63.5 mm). This will provide a minimum clearance of 3.50 inches (88.9 mm) between the rack faces and the insides of the covers. A clearance of not less than 2.00 inches (50.8 mm) shall be provided between the sides of the rack and the inside of the case body and between the top and bottom of the rack and the inside of the case body.
14.3.1.8 Instrument Supports
T-bar instrument supports shall be provided so that each mounting rack frame may be configured for any combination of instruments and/or drawers. The T-bar supports shall extend from the sides of the mounting rack frame to provide support for the instrument chassis. The edges of the installed T-bar supports shall be a maximum of 15 inches apart. The T-bar support mounting shall be designed to provide continuous adjustment of the supports at any position within the mounting rack height so as to provide instrument chassis support for both standard (1.75 inch height increment) chassis and non-standard chassis. The horizontal part of the T-bar must be cut back on both
Attachment 1 34 ends so that it is at least 1 1/2 inches from the rack face. Each case shall be provided with two sets of T-bar supports.
14.3.1.9 Lid Hangers
Lid hangers shall be provided so that each lid may be attached to the side of the case. Brackets on the top of the case enclosure mate with brackets mounted on the lid.
14.3.1.10 Storage Pouch
One storage pouch per case shall be provided. The pouch shall be mounted in the rear case lid, shall be removable via snaps and have a zippered opening. Pouch dimensions shall be at least 12”x12”x2”.
14.3.2 Drawer
The drawer shall be constructed of aluminum or molded glass fiber reinforced plastic with an aluminum face.
Mounting slides shall be supplied to install the drawer in the rack frame. The slides shall be furnished with all necessary hardware for mounting to the front and rear rack frame drilling pattern. It shall be possible to mount the drawer in any 1.75 inch incremental position within the rack frame. The drawer slides furnished shall have a positive stop feature to prevent accidental removal of the drawer from the slide. The drawer slides shall further have a finger operated quick release to permit deliberate removal of the drawer from the slide. The drawer shall be fitted with a lid or cover. The lid shall be secured to the rear of the drawer with an aluminum piano type hinge. To secure the drawer in the rack frame, two one-quarter turn latch buttons shall be provided. One shall be mounted on each side of the front panel of the drawer and engage the rack frame equipment mounting rail. A suitable handle shall be mounted in the center of the front panel. The drawer dimensions shall be as follows:
MINIMUM
DRAWER SIZE FRONT PANEL HEIGHT INSIDE HEIGHT INSIDE WIDTH INSIDE LENGTH
1 3.47 inches 3.00 inches 16.0 inches 18.0 inches
2 6.97 inches 6.5 inches 16.0 inches 18.0 inches
When installed in the rack frame, no portion of the…
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it. Updated .