Cowger lab requirements and SOW for Percival plant growth chamber
I. Statement of Work Dr. Christina Cowger conducts research on a variety of fungal pathogens of small grains, in particular, wheat and barley. Some experiments will be conducted under specific temperature and light conditions, which require the use of a growth chamber designed to accommodate plants at various heights.
We have determined that the Percival Scientific model E-41VL will meet our requirements for a growth chamber for this research.
II. The following criteria are absolute requirements for a growth chamber:
· IntellusUltra Control system with diurnal and 24 hour settings.
All of our other growth chambers are Percival models, and we understand the programming and maintenance of these chambers.
· At least 37 ft3 of interior space.
· 34 ft2 of space over five tiers of shelves
· Temperature range of 7 to 44C with lights on.
· Externally mounted vertical LED lamp banks.
· Lights that are dimmable and can irradiate to an intensity of up to 515 μmoles/m2/s.
· Electrical service requirements of 120/1/60.
III. Details of quote—see attached.
| Percival Scientific plant growth chamber E-41VL | $16,950 | ||
| Estimated freight | $ 1,400 | ||
| Total | $18,350 |
IV. Delivery address Rebecca Whetten 100 Derieux Place NCSU Gardner Hall 3407 Raleigh NC 27695
Rebecca Whetten phone 919-931-3668
File details come from the government source that posted it. Updated .