B08_ATCH_1_SALIENTS_S23030.pdf
PDF 181 KB Posted
- Attached to
- MULTI-PARAMETER WATER QUALITY SONDE IAW SALIENTS Federal contract opportunity
- Solicitation number
- 140G0123Q0071
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Sol_140G0123Q0071_Amd_0001.pdf | ||
| B08_ATCH_3_FISMA_S23030_EXO2S.pdf | ||
| Sol_140G0123Q0071.pdf | ||
| B08_ATCH_2_508_S23030_EXO2S.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics S23030
MULTI-PARAMETER WATER QUALITY SONDE
Salient Characteristics
10 ea. Multi-Parameter Water Quality Sondes:
1. Shall consist of a multi-parameter water quality sonde with at least 7 ports for compatible user-replaceable sensors and a central wiper system.
2. Shall be powered from an external 12 VDC power source only and require no internal batteries.
3. Shall be compatible with a range of sensors including, but not limited to, temperature, conductivity, pH, dissolved oxygen, turbidity, depth, rhodamine, ORP, total algae, FDOM, total dissolved and suspended solids, ISE ammonium, ISE chloride, and UV nitrate. Depth sensor can be added at a later time if needed.
4. No depth sensor shall be included with unit at time of purchase.
5. Shall be capable of supporting SDI-12, RS-232, and USB communications.
6. Shall include wireless Bluetooth communications.
7. Shall have internal data logging with ability to store at least 1,000,000 readings.
8. Shall have a sample rate of at least 4 Hz.
9. Shall operate over a temperature range from at least -5 to 50o C.
10. Shall be no larger than 3.0 inches in diameter and 20 inches in length.
11. Shall have the ability to auto-recognize any sensor in any port.
12. Shall include “wet-mateable” connectors allowing cable changes in wet conditions and providing protection against corrosion.
13. Shall include LED indicators for sonde status and Bluetooth status.
14. Shall include a removable bail for suspending the sonde.
15. Shall include compatible software that allows users to manage, visualize, and organize field data collected from a sonde. In addition, the software shall support:
- viewing live and recorded data,
- sensor calibration,
- managing deployment of a sonde (setup sonde for unattended logging),
- a graphical user interface for data analysis.
File details come from the government source that posted it. Updated .