B08_ATCH_1_Salients_Particle_Size_Analyzer.pdf

PDF 191 KB Posted

Attached to
PARTICLE SIZE ANALYZER IAW SALIENTS Federal contract opportunity
Solicitation number
140G0124Q0040
Issued by
Department of the Interior US Geological Survey Office of Acquisitions and Grants

About this file

This document outlines the salient characteristics for a particle size analyzer system and accessories solicited under solicitation number 140G0124Q0040 by the Department of the Interior US Geological Survey Office of Acquisitions and Grants. The system shall include a laser diffraction particle size analyzer with the ability to detect multiple wavelengths of horizontally and vertically radiated light across a range of 10nm to 3,500um. Accompanying modules are required to handle both dry powder and liquid samples, with the liquid module providing auto-dilution, filling, and rinsing capabilities. Additional requirements include a sonicator kit for dispersion of wet samples, HEPA vacuum for dry sample delivery, and a workstation with the specified software pre-installed. Support and training services such as preventative maintenance, system health checks, software upgrades and basic operator training must also be provided.

View the file

Other files for this federal contract opportunity

Other files attached to PARTICLE SIZE ANALYZER IAW SALIENTS, newest first.
File Type Posted
B08_ATCH_3_FISMA_S23147.pdf PDF
B08_ATCH_2_508_S23147.pdf PDF
Sol_140G0124Q0040.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics S23147

PARTICLE SIZE ANALYZER SYSTEM AND ACCESSORIES

SALIENT CHARACTERISTICS:

System shall consist of a laser diffraction particle size analyzer (optical bench type) for determining particle size distributions in sediment studies. Shall also include sample handling modules, a computer workstation, software, particle size standards, and other accessories.

1 EA, Laser Diffraction Particle Size Analyzer :

1. Shall consist of a multi-wavelength laser diffraction particle size analyzer.

2. Shall have a particle size and analysis measurement range from at least 10 nm to 3,500 um.

3. Shall include technology capable of detecting differences between horizontally & vertically radiated light for three different wavelengths.

4. Shall include a sample handing module, for measuring dry powder samples, with the following features:

a. Shall have a measurement range of at least 400 nm to 3,000 um.

b. Shall be capable of measuring entire dry samples in accordance with ISO Standard 13320.

c. Shall include a programmable obscuration setting for optimizing sample feed rates.

d. Shall include user selectable vacuum pressure.

5. Shall include a sample handling module, for analyzing liquid samples, with the following features:

a. Shall be capable of analyzing samples suspended in aqueous and non-aqueous diluents.

b. Shall have a measurement ranging from at least 0.010 um to 2,000 um.

c. Shall be fully automatic with auto-dilution, auto-filling, and auto-rinsing capability.

d. Shall include a variable speed pump for total dispersion control of samples.

6. Shall be capable of operating from 100 - 240 VAC power.

7. Shall include a USB interface.

8. Shall be capable of exporting data in XLSX, TSV, or PDF file formats.

9. Shall be capable of operating over a range of temperatures from at least -10 to 400 C.

10. Shall be no larger than 20” x 38” x 10” in size.

11. Shall include operational software with the following features:

a. Initiate measurements,

b. Create standard operating methods,

c. View current and historical data with the ability to create overlays,

d. Monitor user diagnostics during analysis,

e. Perform self-check diagnostics,

f. Perform quality control pass/fail checks,

g. Create customized reports, graph data, and export files,

12. Shall include all applicable accessories, including but not limited to, cables, hose, etc. for fully operating the system.

1 EA, Additional Maintenance/Training/Support:

13. Shall include one preventative maintenance visit and one annual system health check, in addition to the standard system warranty.

14. Shall include software and engineering upgrades.

15. Shall include basic operator training.

16. Shall include (as applicable) relocation support.

Salient Characteristics S23147

17. Shall include unlimited remote technical support.

1 EA, Sonicator Kit:

18. Shall be compatible with above the analyzer described above.

19. Shall consist of a device for dispersion control of wet samples.

20. Shall be capable of in-situ sonication before or during sample analysis.

21. Shall include fully adjustable power settings.

1 EA, HEPA Vacuum:

22. Shall be compatible with the dry sampling module described above.

23. Shall consist of high-capacity HEPA style vacuum.

24. Shall be capable of delivering samples to the sensing area of the analyzer.

25. Shall have fully adjustable pressure ranges.

1 EA, Workstation Assembly:

26. Shall be compatible with the analyzer described above.

27. Shall include a Windows 10 compatible computer system with software, display, keyboard, mouse, and required connectors.

28. Shall include a minimum 20” LCD display.

29. Applicable analyzer system software, described above, shall be pre-installed and configured on the computer.

1 EA, Particle Size Standards (15 um):

30. Shall consist of NIST traceable standards to verify instrument performance.

31. Shall have a mean diameter particle size of 15 um.

32. Shall include at least 10/pk.

1 EA, Particle Size Standards (35 um):

33. Shall consist of NIST traceable standards to verify instrument performance.

34. Shall have a mean diameter particle size of 35 um.

35. Shall include at least 5/pk.