B08_ATCH_1_SALIENTS.pdf
PDF 590 KB Posted
- Attached to
- REFRIGERATED WATER SAMPLERS & ACCESS IAW SALIENTS Federal contract opportunity
- Solicitation number
- 140G0124Q0178
About this file
This document is a set of Salient Characteristics for REFRIGERATED PORTABLE WATER SAMPLERS AND ACCESSORIES, which details the technical specifications and requirements for the products needed under federal contract opportunity 140G0124Q0178.
The requirements include 10 refrigerated portable water samplers with various features and capabilities such as corrosion resistance, power compatibility, bottle capacity, NEMA enclosure rating, and pump specifications. Additionally, the contract includes 10 sets of 14 sample bottles and accessories, 10 suction lines, 10 mobile carts, 8 flow interface modules, and 8 level sensors with sensor adapter plates. The contract is a total small business set-aside with a NAICS code of 334513 and a size standard of 750 employees. The Request for Quote (RFQ) will be issued on or about 5/29/2024, with written quotes due by 06/04/2024. The anticipated award date is 06/15/2024. Vendors must be registered in SAM.gov to be eligible for the purchase order.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Sol_140G0124Q0178.pdf | ||
| B08_ATCH_2_508_S24104.pdf | ||
| B08_ATCH_3_Fisma_form_S24104_OAG_12_14_22.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics S24104
REFRIGERATED PORTABLE WATER SAMPLERS AND ACCESSORIES
SALIENT CHARACTERISTICS:
10 EA, Wastewater Sampler:
1. Sampler shall consist of a pump, controller, top cover, center section, base, distributor arm and two pump tubes and power cord. Does not include bottle configuration kit.
2. Shall be made of corrosion resistant, impact resistant, UV resistant materials.
3. Shall be capable of being powered from external 120 VAC, 60 Hz or 12VDC power sources.
4. Shall be compatible with polypropylene 950 ml bottles and capable of holding 14 bottles for sample collection.
5. Enclosure shall have a National Electrical Manufacturers Association (NEMA) 4X, 6 (IP67) enclosure rating, at a minimum.
6. Shall be capable of interfacing with a cell phone modem.
7. Shall be portable and include handles for transporting.
8. Shall include all necessary parts, cables, manual, etc., and applicable software for system operation.
Pump:
9. Shall have a lifting capability of at least 28 feet.
10. Shall include a non-contact type liquid detector device that detects when liquid sample reaches the pump to automatically compensate for changes in head heights.
11. Shall support intake suction tubing up to 99 feet (1 to 30 m) long and be made of vinyl with inside dimension 3/8 inch (1 cm).
12. Shall be repeatable within ±5 ml or ±5% of the average volume in a set, or better.
13. Shall have a sample delivery velocity of at least 2.7 FT./s at 15 ft. (0.83 m/s at 4.6 m)
14. Shall have a pump tubing life of at least 1,000,000 pump counts.
Controller:
15. Shall include a user interface control panel with backlit display and keypad.
16. Shall have non-volatile ROM memory.
17. Shall have operational temperature range from 32° to at least 120°F (0° to 49°C).
18. Shall be SDI-12 compatible.
Operations:
19. Shall have storage capability for at least five (5) sampling programs.
20. Shall have sample frequency of 1 minute to at least 99 hours and 59 minutes, in at least 1-minute increments.
21. Shall allow for automatic rinsing of suction line up to at least three (3) rinses for each sample collection.
22. Shall be capable of running diagnostics test for RAM, ROM, pump, display, and the distributor.
23. Shall sample programmable sample volumes from 10 to at least 9,990 ml in 1 ml increments.
10 EA, Set of 14 Bottles and accessories:
24. Shall be compatible with the sampler described above.
25. Shall consist of fourteen (14), 950-ml polyethylene bottles, with caps.
26. Shall include at least one (1) vinyl suction line that has 3/8 inch inside diameter and is at least 25 ft. long with standard weighted polypropylene strainer, tubing coupler, and at least two (2) pump tubes.
10 EA, Suction Line:
27. Shall be compatible with the sampler described above.
28. Shall be at least 100 ft. long and made of vinyl.
29. Shall consist of a 3/8-inch tubing coupler made of stainless steel or vinyl.
30. Shall consist of a 3/8-inch stainless-steel strainer for low-flow applications.
10 EA, Mobile Cart:
31. Shall be compatible with the sampler described above.
32. Shall include frame with lifting handles, pneumatic tires and pull handles.
8 EA, Flow Interface Module:
33. Shall consist of a Flow Module device that interfaces between a compatible water sampler and the level sensor described above.
34. Shall be compatible with Isco Brand Avalanche and 6700 Series Samplers.
35. Shall be capable of converting water level sensor analog outputs to water level readings.
36. 4. Shall be capable of operating over a temperature range from at least 0 to 49o C.
37. Shall include non-volatile memory and have the ability to store data at various intervals ranging from 1 minute to 30 minutes, at a minimum.
38. Shall have a NEMA 4x Enclosure with an Ingress Protection (IP) rating of at least IP67, or better.
39. Shall be capable of operating from an external 12 VDC (nominal) power supply.
40. Enclosure shall be no larger than 5” x 6” x 2” and weigh no more than 1.5 lbs.
41. Shall include a rechargeable desiccant cartridge, or other moisture absorbent device, to prevent moisture from entering the sensor vent tube.
8 EA, Level Sensor:
42. Shall consist of a vented submersible pressure sensor for measuring water level.
43. Shall be compatible with the flow interface module described below.
44. Shall have a measurement range from at least 0.1 to 10 ft. of water.
45. Shall include a fixed, vented cable of 75 ft. in length.
46. Shall be capable of operating over a temperature range from at least 0 to 49o C.
47. Shall have a measurement accuracy of +/-0.012 ft./ft., or better.
48. Shall be no larger than 1” in diameter and 10” in length and weigh no more than 5 lbs.
49. Shall be capable of operating in NEC (National Electric Code) Hazardous Area Classifications Class 1, Division 1 Groups A, B, C, and D.
50. Shall be capable of producing an analog output signal that is compatible with the flow interface module described below.
8 EA, Sensor Adapter Plate:
51. Shall consist of a metal adapter plate for installing the sensor (described above) to a compatible mounting system in a downhole round sewer pipe.
52. Plate shall be constructed of non-corrosive metal.
53. Shall include tabs and a threaded screw (with nuts, and washers) for attaching the sensor.
54. Shall also include side clamps to facilitate holding the sensor in place.