Attachment D - Procedure for Residual Measurement - Pages 1-4.docx
DOCX document 30 KB Posted
- Attached to
- Drinking Water Samp. -Bacteriol State and local contract opportunity
- Solicitation number
- 3000024507
- Issued by
- East Baton Rouge Parish, Baton Rouge City, Louisiana
About this file
This document is an attachment detailing the Procedure for Residual Measurement for the Louisiana Department of Health, Office of Public Health (LDH/OPH) under RFx #3000024507. The procedure specifically outlines the precise steps for collecting and measuring chlorine residuals when sampling water for bacteriological analysis in Public Water Systems (PWS). The protocol requires meticulous sampling techniques, including disinfecting sample taps, flushing water lines, and using approved measurement methods such as N,N-diethyl-p-phenylenediamine (DPD) colorimeters to assess free or total chlorine levels.
The documentation provides comprehensive technical guidance for chlorine residual measurement, including specific meter types, test procedures, and dilution methods. It mandates weekly meter verification using standardized chlorine standards, details multiple measurement ranges and methods (including low and high-range tests), and specifies strict documentation requirements. Analysts must record sample collection times, residual measurements, and enter data into the STARLIMS system by 8:00 AM the day following sample collection, with initial data entry to be performed by LDH/OPH staff until the contractor is set up for system access.
View the file
Other files for this state and local contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment A - Special Terms & Conditions - Pages 1-10.docx | DOCX document | |
| Attachment B - Specifications - Pages 1-2.docx | DOCX document | |
| Attachment E - Sample Truck Routes & Delivery Locations - Page 1.pdf | ||
| Attachment G - Online Bidding Instructions - Pages 1-10.pdf | ||
| Attachment C - Procedure for the Collection of Bacteriological S.docx | DOCX document | |
| 3000024507.pdf | ||
| Attachment F - Price Sheet - Page 1.xls | XLS spreadsheet |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment D - Procedure for Residual Measurement Louisiana Department of Health, Office of Public Health (LDH/OPH) RFx #3000024507
PURPOSE: The following procedure must be followed for the Residual Measurement (free or total chlorine) when collecting routine samples for bacteriological (total coliform/E. coli) analysis.
I. Sampling Procedure:
1. Remove any attachments from the sample tap, such as hoses, filters, screens, or aerators, and disinfect the sample tap by spraying the sample tap thoroughly with Isopropyl alcohol (91%). Allow to air dry for 2 minutes.
2. Flush the sample tap for 5 minutes until the water temperature reaches a constant temperature, and then reduce water flow to a smooth (laminar) stream about the size of a pencil if possible.
3. The disinfectant (chlorine) residual must be measured at the time of collection for the total coliform sample at the LDH/OPH-approved sample sites (e.g., TCR-001, MRT-001, etc.) in the Public Water Systems (PWS) distribution system.
4. The disinfectant (chlorine) residual must be measured using an LDH/OPH-approved N,N-diethyl-p-phenylenediamine (DPD) colorimeter (e.g., HACH Colorimeter).
5. Measure the residual type (free or total) that is specified for the water system. Total chlorine is measured for systems that use chloramines. Measure the residual using an appropriate method, taking into account any interfering substances that may be present. For example, both manganese and chloramine can interfere with the DPD colorimetric method, but do not interfere with the indophenol method. As a result, for systems that operate using free chlorine, that have manganese and/or naturally occurring ammonia, the indophenol method or a DPD method that specifically accounts for the interference of concern should be selected
6. Only EPA [see 40 CFR Part 136.3, Table II] or State-approved methods for potable water shall be used. When running the test, the analyst should follow the specific directions provided with the test kit. In general, the analyst will need to measure out a known volume of sample and will need to add the reagent(s) in the order described, wait a specific reaction time, and then measure the color that develops in the sample. The intensity of the color that develops after the addition of a reagent is measured using a spectrophotometer and relates directly to the amount of disinfection residual present in the sample.
7. Record the sample collection date (if changed) and time and residual measurement (and Meter#) on the sample chain-of-custody form (COC). The COC must be signed and included with the samples to the LDH/OPH Lab or Depot.
8. Enter the sample collection date (if changed) and time and residual measurement into STARLIMS by 8:00 A.M. (CT) on the day after sample was collected. Initially this will be done by LDH/OPH staff until the Contractor is setup to access STARLIMS.
II. Meter and Methods:
| Meter |
| Test |
| Sample Size |
| Sample Container |
| Reagents |
| Test Procedure |
| Range |
Pocket Colorimeter™ II, Chlorine (Free and Total) #5870000 DPD Free (Low Range)
| 10-mL |
| Glass |
#2427606
DPD
10 mL Powder Pillows
| Refer to directions provided with test kit. |
| 0.02 – 2.00 mg/l (will max out and flash at 2.20 mg/l) |
DPD Total (Low Range)
DPD Free (High Range)
| 5-mL |
| Plastic |
#4864302
DPD
2-10 mL or 1-25 mL Powder Pillows
| Refer to directions provided with test kit. |
| 0.1-8.0 mg/l for 2-10 mL |
0.1-10.0 mg/l for 1-25 mL
DPD Total
Pocket Colorimeter™ II, Chloramine, Mono and Free Ammonia (Free) #5870026 Indophenol Free Cl Hach Method 10241
| 10-mL |
| Plastic |
#4864302 Freechlor F Reagent Solution (5 drops)
Monochlor F Reagent Powder Pillows
| Refer to Hach Indophenol Method 10241 |
| 0.04 - 4.50 mg/L |
DR900 Multiparameter Portable Colorimeter #9385100 DPD Free (Low & High) 5-mL 10-mL Plastic #4864302 Glass #2401906 DPD Powder Pillows Free 10069, 10245 Total 10070, 10250
| Refer to directions provided with test kit. |
| High 0.1-10 mg/L |
Mid 0.05 – 4.00 mg/L
DPD Total (Low & High)
Indophenol Free Cl Hach Method 10241
| 10-mL |
| Glass |
#2401906 Freechlor F Reagent Solution (5 drops) Monochlor F Reagent Powder Pillows
| Refer to Hach Indophenol Method 10241 |
| 0.04 – 4.50 mg/l |
| Pocket Colorimeter™ II, Chlorine, (Free and Total) plus pH– not recommended for compliance reporting. |
| DPD Free |
5-mil
Plastic #4864302
DPD 1-25 mL Powder Pillow
Refer to directions provided with test kit.
0.2 – 10.0 mg/L Requires LR procedure when result is < 2 mg/L; instrument does not support LR procedure.
DPD Total
Notes:
1. Under Range – flashing number below test range; Verify instrument cap is correctly seated and Check zero by measuring a blank
2. Over Range - flashing number above test range; Check for Light Blockage; Switch meter to HR and retest sample; if the sample is still over range, dilute the sample to bring it within the test range. See Section IV. HOW TO DILUTE SAMPLES
3. Field test results that are suspected of being incorrect or are unexpected may be resampled for confirmation.
III. METER VERIFICATION:
1. The disinfectant (chlorine) residual meter must be verified weekly with chlorine standards. There are 2 sets, Low Range and High Range; each with 1 blank and 3 standards.
2. Check that the chlorine standards are not expired.
3. Verify the meter with both Low Range and High Range by zeroing with the blank and reading each standard. See below examples.
4. If the meter fails by reading outside the range of the standard, do not use the meter for chlorine measurement. Contact the central office for another meter.
| Meter: SN 13080E228638 | Verification Date: 1/6/17 | |
| Standard Lot: A3242 | ||
| Exp Date: 8/2017 |
| Low Range |
| Standard Value |
| Range |
| LR Meter Reading |
| Pass/Fail |
| Standard 1 |
| 0.22 ± 0.09 |
| 0.13-0.31 |
| 0.21 |
| Pass |
| Standard 2 |
| 0.92 ± 0.10 |
| 0.82-1.02 |
| 0.93 |
| Pass |
| Standard 3 |
| 1.61 ± 0.14 |
| 1.47-1.75 |
| 1.63 |
| Pass |
| Standard Lot: A3229 |
| Exp Date: 8/2017 |
| High Range |
| Standard Value |
| Range |
| HR Meter Reading |
| Pass/Fail |
| Standard 1 |
| 2.1 ± 0.2 |
| 1.9-2.3 |
| 2 |
| Pass |
| Standard 2 |
| 3.7 ± 0.3 |
| 3.4-4.0 |
| 3.7 |
| Pass |
| Standard 3 |
| 6.7 ± 0.6 |
| 6.1-7.3 |
| 6.8 |
| Pass |
IV. HOW TO DILUTE A SAMPLE WHEN THE RESULT IS OVER RANGE:
The simplest dilution is a one-to-one or 2-fold dilution, where one part sample is mixed with one part deionized or distilled water. This reduces the concentration of anything in the sample by half. Follow the usual test procedure using the diluted sample, and then multiply the result by 2 to find the concentration of the sample before it was diluted.
Example: Your sample reads off-scale on your kit or instrument display. Mix 5 mL sample with 5 mL deionized water and repeat the test. The result using this diluted sample is 1.9 mg/L. The concentration of the sample before dilution is (1.9 x 2) = 3.8 mg/L.
If the sample is still over-range after dilution, a larger dilution will need to be made. For example, mix one part sample to two parts deionized water for a 3-fold dilution (multiply result by 3), one part sample to three parts deionized water for a 4-fold dilution (multiply result by 4), and so on. If your test uses the sample as a blank, use the diluted sample as the blank.
It is very important to measure the amount of sample and deionized water accurately when making dilutions. An error in these measurements will lead to an error in the result, which will then be multiplied by the dilution factor. At minimum, measure the sample and deionized water in a graduated cylinder. For best results use a TenSette pipet for measuring the sample, add it to a Class A volumetric flask, and fill to the mark with deionized water. Mix thoroughly before testing.
You can adjust the measured volumes to accommodate the pipets and flasks available to you. For instance, if you need to make a 20-fold dilution, you can measure 1 mL of sample and add 19 mL of dilution water. Or you can measure 5 mL of sample and add 95 mL of water (or dilute to the mark on a 100-mL volumetric flask) to get the same dilution. If only 10 mL is required for running the test, use just 10 mL of this mixture. Use the extra for the blank or duplicate measurements if desired.