About this file

This specifications document outlines requirements for WR-6 thermoelectric microcalorimeter standards to be procured by the National Institute of Standards and Technology. The standards are needed to support research and measurement services in microwave power calibrations for 5G communication, spectrum sensing, and modulated-wave metrology. Three units are required initially with an option to purchase up to three additional units. Key specifications include an operating frequency range of 110-170 GHz, maximum input power of at least 20 dBm, thermopile sensitivity of at least 100 mV/W, dimensions less than 100x50x50 mm, and total mass less than 100g. The related federal contract opportunity is solicitation number NB6720102302197 for WR-6 thermoelectric microcalorimeter standards from the Department of Commerce National Institute of Standards and Technology. Response and award dates are not provided.

View the file

Other files for this federal contract opportunity

Other files attached to WR-6 Thermoelectric Microcalorimeter Standards, newest first.
File Type Posted
Attachment B - Experience Project Data Sheet.docx DOCX document
Attachment C - QAs-1.xls XLS spreadsheet

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Requirements / Specification Document

TITLE: WR-6 thermoelectric microcalorimeter standards

Requesting Laboratory / Division / Group:

Communications Technology Laboratory RF Technology Division (672) Guided Wave Electromagnetics Group (672.01)

General Statement of Need:

The Communications Technology Laboratory (CTL) at the National Institute of Standards and Technology (NIST) enables science and industry by providing essential measurement methods, instrumentation, and standards to support all phases of communications technology development, from fundamental characterization and discovery to production.

To support research and measurement services in microwave power calibrations, the Radio Frequency Technology Division in CTL requires WR-6 thermoelectric power sensor standards that are compatible with microcalorimeters. This instrumentation is needed for CTL to advance the state-of-the-art in microwave power measurements and to support 5G communication, spectrum sensing, and modulated-wave metrology.

Item Specifications and Quantities:

Item 1: WR-6 thermoelectric microcalorimeter standard Quantity: 3 Specifications: 1.1 to 1.13

Optional line items Item 2: WR-6 thermoelectric microcalorimeter standard Quantity: Option to purchase 1 to 3 additional units Specifications: 1.1 to 1.13

1.1. Shall have a WR-6 rectangular waveguide interface that is compliant with the precision-dowel interface IEEE 1785-2a, as specified in IEEE Std 1785.2-2016.

1.2. Shall have a minimal operating frequency range of 110 GHz to 170 GHz.

1.3. Shall have a radiofrequency (RF) termination with a small reflection coefficient magnitude throughout the operating frequency range. A reflection coefficient magnitude of less than 0.2 is ideal. Smaller reflection coefficient magnitudes are considered advantageous in comparison to larger reflection coefficient magnitudes.

1.4. The RF termination shall have a maximum input power of at least 20 dBm.

1.5. Shall have a DC heater with a maximum input power of at least 20 dBm.

1.6. The DC heater shall have a four-wire lead configuration.

1.7. The DC heater shall have a DC resistance in the range of 50 Ω to 2 kΩ.

1.8. Shall have a thermopile to sense heating from RF power dissipation with a sensitivity of at least 100 mV/W and a thermopile to sense heating from DC power dissipation with a sensitivity of at least 100 mV/W. Heating from both RF and DC power may be sensed with a single thermopile or with one thermopile for the RF termination and one for the DC heater. Higher sensitivity is considered advantageous in comparison to lower sensitivity.

1.9. Shall have a DC connector that allows direct electrical connections to the DC heater, the thermopile(s), and any auxiliary temperature sensors.

1.10. Low RF leakage from the standard throughout the operating frequency range is advantageous, with lower leakage considered advantageous in comparison to higher leakage. RF leakage at least 40 dBc below the RF input power is ideal.

1.11. If the standard contains active readout or data storage electronics, it shall be possible to operate the standard without supplying power to these electronics. The RF termination, DC heater, thermopile(s), and any auxiliary temperature sensors shall be operable without supplying power to active readout or data storage electronics.

1.12. Shall have dimensions of less than 100 mm × 50 mm × 50 mm (length × width × height).

1.13. Shall have a total mass of less than 100 g.

Procurement Support Information and Ancillary Services:

Section Requirements/Specifications

Planning Considerations ☒ N/A

☐ List any visits required to allow potential vendors to see site specific constraints.

[Explain Requirements.] ☐ US citizen required on-site (foreign nationals require pre-registration 30 days prior to visit.)

☐ Vendor Representative On-Site more than three (3) days.

☐ Other Considerations:

[List other considerations to ensure an accurate quote.]

Shipping & Shipping Address:

Delivery ☐ N/A

☒ NIST-Boulder Mail Stop: 672.01 325 Broadway Boulder, CO 80305

Shipping Criteria:

☒ Partial Delivery Acceptable

☐ [Enter Shipping and Delivery information by item.] ☐ Direct Delivery to OU Building Required

[Explain Need for Requirement] Building /Room Number: [Enter Building/Room Number] ☐ Building has Loading Dock Containerization Preference: [Enter Container Type] Other Requirements:

[List Any Other Requirement]

Delivery Date Criteria:

☒ Delivery Shall Be Competed No Later Than:

6/1/2024

☐ Delivery Shall Be Made No Earlier Than:

Click or tap to enter a date.

☐ Scheduled Deliveries:

Item Number Due Date

Electronic Media ☒ N/A

[Enter Electronic Media Requirements by Item.]

☐ Software

☐ Electronic Manuals ☐ License Key

☐ Other:

[List Any Other Electronic Media Requirements]

☐ US Government Email Address for Software Delivery:

[Enter Address]

Installation ☒ N/A

[Enter Installation Requirements by Item]

☐ Vendor Technician on Site

☐ Rigging

☐ Uncrating / Unpackaging

☐ Removal of Packaging Material

☐ Equipment Set Up

☐ Start-Up Services

☐ Turnkey Installation

☐ Other:

[Enter Details]

NOTE: TPOC shall be the coordination POC.

Facility / Utility Considerations ☒ N/A

[List Site Specific Utility Considerations by Item]

☐ OFPM Work Order Numbers [List Work Order Numbers]

☐ Power Available Voltage: [Enter voltage] (V) Maximum Current: [Enter max current draw] (A) Phase: [Choose Phase] Frequency: [Typically 60Hz] (Hz) NEMA Plug Type: [Enter Plug Type] (i.e.: NEMA L5-20P) Environment: [Choose Environment] Other Power Considerations:

[List Any Other Power Considerations.]

☐ Utility Connections:

[List Relevant Available Utilities, Pressure(s) and Flowrate(s).]

☐ Other Considerations:

[List Other Considerations to Ensure an Accurate Quote.]

Inspection and Acceptance

[Inspection Requirements Per Item (Who, When, What, Where)]

☐ On-Site

☐ N/A ☒ Acceptance Expected to Not to Exceed 7 Business Days ☐ Acceptance Expected to Exceed 7 Business Days:

☐ Inspection and Acceptance Plan:

[Explanation] ☐ Other:

[Enter Other Requirements]

Training ☒ N/A

Training Requirements (Who, When, What, Where)

☐ Number of Trainees: [Enter Number of Trainees] ☐ On-Site ☐ Off-Site Location: [Enter Location(s)] Is Trainee Travel Required: [Choose Answer] Scope of Training ☐ Operation ☐ Maintenance ☐ Troubleshooting

☐ Safety ☐ Other [Enter Other Scope of Training Requirements] ☐ Training Materials to be Provided by Government:

[Describe Materials] ☐ Training Materials to be Provided by Vendor:

[Describe Materials] ☐ Other [Enter Other Training Requirements] NOTE: TPOC shall be coordination POC.

Warranty ☐ N/A

[Choose Warranty Type] Warranty Length: Standard manufacturer’s warrantee

What is included?

☐ Labor ☐ Parts ☐ Travel

☐ Cost and Liability for Returns ☐ Other

[Enter Other Requirements]

☐ 1 Option Period ☐ 2 Option Periods ☐ 3 Option Periods ☐ 4 Option Periods (Maximum)

Calibration ☒ N/A

[Describe Calibration Requirements Per Item] Calibration Plan Length: [i.e. 3 - Years]

What is Included?

☐ Labor ☐ Spare Parts ☐ Consumables

☐ Travel ☐ Cost and Liability for Shipment ☐ Software Updates ☐ Manuals

☐ Other [Enter Other Requirements] ☐ Requirements:

[Enter Requirements] (i.e.: On-Site Within 72 hours, etc.)

Length:

[i.e. 1 – Year / 6 - Months]

☐ 3 Option Periods ☐ 4 Option Periods (Maximum)

Maintenance ☒ N/A

[Describe Maintenance Requirements Per Item] Maintenance Plan Length: [i.e. 3 - Years]

What is Included?

☐ Labor ☐ Spare Parts ☐ Consumables

☐ Travel ☐ Cost and Liability for Shipment ☐ Software Updates ☐ Manuals

☐ Other [Enter Other Requirements] ☐ Requirements:

[Enter Requirements] (i.e.: On-Site Within 72 hours, etc.)

Length:

[i.e. 1 – Year / 6 - Months]

☐ 3 Option Periods ☐ 4 Option Periods (Maximum)

Government Furnished Material / Equipment ☒ N/A

List Any Government Furnished Material / Equipment Per Item] ☐ Samples for Testing ☐ Equipment which will leave Government Site

Description:

Serial Number:

NIST Property Number:

[Enter Description] [Enter Serial Number] [Enter NIST Property Number]

☐ Other: [Enter Other Details]

Travel ☒ N/A

[Describe any Travel Requirements Per Item] ☐ Explain:

[Enter Justification / Requirement] ☐ Other:

[Enter Other Requirements]

NOTE TO VENDOR: To the maximum extent possible, include travel requirement estimates as part of pertinent sections above (Installation, Maintenance, Repair, etc.). Otherwise, Travel will be a separate line item, reimbursable per the Federal Travel Regulation.

File details come from the government source that posted it. Updated .