This document provides salient characteristics for an ENT Guided Navigation System procurement. The solicitation seeks a navigation system that utilizes EM and/or optical tracking methods, reusable and disposable instruments, patient trackers, and registration capabilities. Key requirements include mobile cart functionality, touchscreen monitor, storage, instrument interfaces, and support for side-mount or flat EM emitter designs. The related federal contract opportunity is solicitation 36C26223Q1766 for an ENT Guided Navigation System through the Department of Veterans Affairs Veterans Health Administration VISN 22.

View the file

Other files for this federal contract opportunity

Other files attached to 6515--ENT Guided Navigation System, newest first.
File Type Posted
ATTACHMENT B - SOW-Specifications.docx DOCX document
36C26223Q1766.docx DOCX document
Certificate of Compliance (return with offer) NMR Waiver Granted.pdf PDF
RFQ Description (ENT Guided Navigation System).docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

36C26223Q1766

ATTACHMENT A- SALIENT CHARACTERISTICS

ENT Guided Navigation System

Salient Characteristics:

· Must have EM type of navigation at a minimum but can also have other types of navigation capabilities

· Must have hind-tracked and tip-tracked instrumentation and tracking methods

· Must have reusable probes, suctions, and necessary single use items for instruments

· Must be able to track other (plug & play) instrumentation

· Must support use of Optical and EM designed instruments

· Must have disposable type patient trackers

· Must have registration scan capabilities for MR/CT

· Vendor must provide in-service delivery, set-up, functionality checks/tests, and continued support for repairs/warranty issues

· Wheeled cart:

· Mobile capability to allow unit to move from room to room

· High-definition touch screen monitor, flat panel design, that is adjustable via attached arm to allow multiple configurations for user needs

· Lockable wheeled casters

· At least one or more storage bins

· Input/Output (I/O) panel for connections to optical instruments and/or microscopes

· USB keyboard/mouse

· EM instrument interface

· Horizontal sliding work surface

· Infrared camera to enable geometrically triangulation of the spatial coordinates for each optical marker on the instrument and the reference frame

· Electromagnetic emitters/fields:

· Must support use of 2 emitter designs (one at a time is adequate): side-mount and flat, via EM emitter port(s) on an (I/O) panel

· Must support use of Side-mount emitter navigation field

· Must support use of Flat-emitter navigation field