ATTACHMENT A - SALIENT CHARACTERISTICS - CHEMISTRY PIPETTING SYSTEM.docx

DOCX document 17 KB Posted

Attached to
6630--CHEMISTRY PIPETTING SYSTEM Federal contract opportunity
Solicitation number
36C26223Q1752
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22

About this file

This document outlines the salient characteristics for a chemistry pipetting system procurement. The Phoenix VA Healthcare System requires a pipetting calibration and software system that can dispense microliter volumes between 5-1000 microliters. The system must include software that allows users to view calibration schedules and due dates, operator assessments, digital reviews and approval workflows, generate reports including audit trails, and schedule competency assessments. The software should be 21 CFR Part 11 compliant. The system must generate results traceable to national and international standards. It must also comply with regulatory requirements of the Joint Commission.

View the file

Other files for this federal contract opportunity

Other files attached to 6630--CHEMISTRY PIPETTING SYSTEM, newest first.
File Type Posted
ATTACHMENT B - SCHEDULE OF PRICING - CHEMISTY PIPETTING SYSTEM.xlsx XLSX spreadsheet
36C26223Q1752.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics – Chemistry Pipetting System Phoenix VA Healthcare System

1. Brand Name or Equal Description

BRAND
PRODUCT DESCRIPTION
PRODUCT #
QTY
Artel
Pipette Calibration System (PCS)
PCS-512
1
Artel
ArtelWare Software
AW-200
1
Artel
PCS Laptop (Win 10),
PCS-793
1
Artel
Label Printer
PCS-746
1
Artel
QualAssure Bulk Kit
PCS-211
1
Artel
ArtelWare PCS Instrument Calibrator Kit
AW-600
1

2. Salient Characteristics

· Must be able to dispense in the following microliter volumes;

a. 5, 10, 20, 25, 50, 70, 80, 100, 200, 250, 300, 500, and 1000

· Must include software.

· Software should provide a dashboard that allows users to;

a. View due calibrations

b. View and schedule operator assessments

c. View digital reviews and approval workflows.

d. Generate reports, including automatic report generation for audit trails.

e. Print calibration labels

f. Schedule competency assessment

g. Print calibration labels

· Must be 21 CFR Part 11 compliance ready.

· Must generate results traceable to national (NIST) and international standards (SI units).

· Must comply with regulatory requirements of the Joint Commission (JC) standards.

File details come from the government source that posted it. Updated .