This document outlines technical requirements for an imaging spectrometer to be procured under solicitation number FA910124QB012 by the Department of the Air Force Materiel Command Test Center. The imaging spectrometer must have a CCD imaging sensor measuring less than 700 by 500 by 250 millimeters, a focal length greater than 500 mm, and capabilities to eliminate re-diffracted light. It must include a kinematic triple grating turret with 1800, 1200, and 2400 gr/mm gratings and have wavelength accuracy of +/- 0.2 nm. Additional requirements include computer-controlled slits and filter wheel, as well as purge port, cabling for control and data acquisition, and software compatible with Windows 10. Relevant filters must also be provided.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| FA9101-24-Q-B012 Sources Sought.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
FA9101-24-Q-B012
UV-NIR System Spectral Response (SSR) Optical Bench
Technical Requirements
Provide an Imaging Spectrometer with the following specifications.
• Charge-coupled device (CCD) imaging sensor
• Dimensions o Length: less than 700 millimeters o Width: less than 500 millimeters o Height: less than 250 millimeters
• Focal length greater than 500 mm
• Designed to eliminate re-diffracted light
• Computer controlled, kinematic triple grating turret with on-axis grating rotation
• Gratings to be included:
o 1800 gr/mm classically ruled grating at 300 nm o 1200 gr/mm blazed holographic grating at 500 nm o 2400 gr/mm blazed holographic grating at 400 nm
• Wavelength accuracy +/- 0.2 nm
• Purge port
• Computer controlled front, side, and exit slits that can be varied from 0 to 7 millimeters
• Motorized, computer controlled 6 position filter wheel for up to 6 each 1" diameter filters.
• Three filters to be included:
o Order sorting long pass filter with 325 nm (+/- 2 nm) cut-on; colored glass; 1" diameter o Order sorting long pass filter with 385 nm (+/- 2 nm) cut-on; colored glass; 1" diameter o Order sorting long pass filter with 630 nm (+/- 2 nm) cut-on; colored glass; 1" diameter
• Software to provide complete instrument control and data acquisition. Must be compatible with Windows 10.
• Appropriate cabling to facilitate instrument control and data acquisition between instrument and AEDC provided computer.