This statement of need is for the procurement of two liquid handling systems capable of positive displacement liquid dispensing with six dispensing syringes, and air-displacement pipetting with a minimum range of 0.5uL to 125uL across pipetting tip formats including 384, 96, 8 and 16 channel arrays. The dispensing system must accommodate a wide range of liquids from 80% ethanol to 50% glycerol without classification requirements. The pipetting system must use liquid classes to compensate for different liquid types. Additional requirements include a small footprint of 26 by 22 inches, at least 16 labware positions with space for ancillary equipment, a fully configurable user interface, protocol authoring on any Windows device, and fixed base configurations to share protocols between systems. The related federal contract opportunity is a brand name or equal solicitation from the National Institute of Allergy and Infectious Diseases for two liquid handling systems meeting the specifications in the statement of need.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment 2 - Bill of Materials (BOM).pdf | ||
| Attachment 3 - Full Text Clauses.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Needs –Liquid Handling System
This statement of need is for the procurement of two instruments capable of the following:
1. Liquid handling systems
a. Dispensing system: the system must use liquid dispensers with TRUE positive displacement of liquids for aspiration and non-contact dispensing with 6 dispensing syringes.
b. Pipetting system: the system must include at least a 384-tip air-displacement pipetting head, working in a minimum range from 0.5uL to 125uL. Plastic consumable tips must be automatically exchanged by the system and available in 384, 96, 8, and 16 channel arrays, for flexibility to work with a wide variety of 96 and 384 well plates. The system must be able to automatically switch between these formats for flexibility.
2. Liquid Class management
a. Dispenser Liquid classes: the dispense system must be capable of handling a wide range liquid type from 80% ethanol through to 50% (w/v) glycerol, with NO requirement for liquid classification. Common biological assay components such as aqueous buffers, DMSO, beads, cells, or glycerol protein stocks must all be dispensed using the same standard settings as well as particulates such as SPRI beads.
b. Pipetting liquid classes: The pipetting system must use liquid classes to compensate for different liquid types and is well suited to a range of liquid types.
It uses an air-displacement pipetting system.
3. NIAID has limited space and requires a small footprint 26 inches wide by 22 inches deep.
The instrument must have at least 16 positions for labware and additional space for a plate shaker, incubator, and reagent storage.
4. User Interface must be completely configurable to tailor the user experience from those requiring restricted access to run specific protocols only, to full open access to all features for Super users. Must be able to install the software on any windows device to author protocols virtually anywhere and send them to the system via a network or over the cloud.
5. The system must include fixed base configurations to share protocols between systems.
Configuration must also include a microplate shaker and incubation capabilities.
6. Must be application ready with off the shelf protocols ready to use.
File details come from the government source that posted it. Updated .