This document contains specifications for a near-infrared spectroscopy sample analyzer to be procured by the United States Department of Agriculture Agricultural Research Service under solicitation number 12905B23Q0244. The analyzer must be capable of determining protein, moisture, fat, ash, fibre, starch and color levels in biological samples weighing as little as 10 grams or 5 grams at a rate of at least one sample per minute. It must operate in reflectance mode across a wavelength range of 850 to 2500 nanometers with a signal to noise ratio under 50 micro absorbance units in the 400 to 700 nanometer range. Additional requirements include a data interval and spectral resolution under 0.5 nanometers, wavelength precision under 0.005 nanometers and accuracy within 0.05 nanometers of an internal reference standard, and an absorbance range less than or equal to 2 absorbance units. The supplied software must support discriminant analysis, local predictions, and display and selection of individual subscans.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Combined Synopsis Solicitation 12905B23Q0244 RFQ.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Near-infrared spectroscopy (NIRS) sample analyzer specifications
• Determine protein, moisture, fat, ash, fibre, starch and colour of biological samples including ground, unground or liquid material such as compound feed, grains/cereals, vegetal protein meals, vegetal by-products, animal by-product/rendering, forages.
• Analyze at least one sample per minute.
• Must be able to handle samples as small as 10 grams or up to 5 grams.
• Must cover a wavelength range from 850 to 2500 nanometers (nm), or beyond, and operate in reflectance measurement mode.
• Signal to noise ration must be less than 50 micro absorbance units (au) in the 400 to 700 nm range.
• Data interval and spectral resolution must be a under 0.5 nm.
• Wavelength precision must be less than 0.005 nm and accuracy must be within 0.05 nm of internal reference standard.
• Optical bandwidth must be 8.75 +/- 0.1 nm.
• Absorbance range must be less then or equal to 2 au.
• Software must support discriminant analysis and local preditions.
• Software must be capable of displaying and selecting individual subscans of each sample.