Attachment 1 Specifications.pdf
PDF 402 KB Posted
- Attached to
- Autoclave (Wenatchee, WA) Federal contract opportunity
- Solicitation number
- 12905B23Q0141
About this file
This document includes specifications for an autoclave and details of a related federal contract opportunity. The specifications call for a STERIS AMSCO 250LS autoclave or equal model with pre-vacuum capability, electric steam generation, freestanding cabinet, 120V/50-60Hz and 208V/60Hz electrical power, power doors, at least 250L capacity, options for two shelves and a loading rack, single door touch display for setting cycles, control lockout switch, gravity and liquid cycles capable of 130C, cabinet depth of 50" or less from front to back, and an auto flush and drain system able to connect to a 1 1/2" ODT drain terminal, 1/2" ODT generator drain, 1" NPT cold feed water at 30-50 psig, and 1/2" NPT hot feed water at 20-50 psig. Proposals should include delivery and installation. The related federal contract opportunity is solicitation number 12905B23Q0141 for an autoclave in Wenatchee, Washington issued by the US Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| RFQ_12905B23Q0141_Terms and Conditions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications
Brand Name or Equal to STERIS AMSCO 250LS autoclave.
Autoclave specifications and salient characteristics:
Pre-vacuum capability Electric steam generation Freestanding cabinet (not to be recessed into wall) 120V, 50/60Hz electrical (for controls) 208V, 60Hz (for steam generator) Power doors At least 250L capacity Options for two shelves and loading rack Single door Touch Display for setting cycles Control Lockout switch for safety Must have Gravity and Liquid cycles Must get up to 130C at least Cabinet depth from front to back (not including handle) no greater than 50" Auto Flush and Drain System
Must be able to connect to/feed into:
1 1/2" ODT drain terminal 1/2" ODT generator drain 1" NPT (30-50 psig dynamic) cold feed water 1/2" NPT (20-50 psig dynamic) hot feed water
Proposals should include delivery and installation.
File details come from the government source that posted it. Updated .