This document includes specifications for biological incubators required by the US Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area. The agency requires biological incubators to rear insect species and conduct research experiments, which must maintain consistent temperatures and humidity under a timed lighting schedule to replicate natural environments. The specifications call for incubators equivalent to the Percival Scientific Incubator I-22L, with two separate controllable chambers per unit, independent programmable temperature and humidity controls for each chamber, temperature ranges of 4-44°C with lights on and 2-44°C with lights off, LED lighting that is programmable and dimmable, and an IntellusUltra control system for monitoring and programming all variables with step and ramp modes. The related federal contract opportunity is solicitation number 12905B23Q0152 for biological incubators to be located in Parlier, California from the same agency.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| RFQ_12905B23Q0152_Terms and Conditions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATIONS and Salient Characteristics: Biological Incubator
Brand Name or Equal
USDA ARS SJVASC-CPQ requires biological incubators to rear insect species and conduct ongoing research experiments. These incubators must be held at consistent temperatures and humidity under a timed lighting schedule to replicate natural environments in the laboratory.
The required incubator must be equivalent to the Percival Scientific Incubator – I-22L, which has two separate controllable chambers per unit to maximize space and expedite experiments.
The incubators have 24-hour chamber monitoring with a programable controller for all variables (temperature, humidity and lighting) allowing up to 500 different programable steps and ramps.
The compressors for each chamber have a range of 4°- 44°C and will hold temperatures at a 0.05-degree precision. Percival also offers support for the entire life on the incubator.
22 gauge interior and 18 gauge exterior electro-zinc plated steel construction.
Two independent programable chambers per unit.
Each chamber has own compressor, observation door with window and access port.
Temperature range of 4°- 44°C with lights ON (±0.5°), 2°- 44°C with lights OFF (±0.5°).
Adjustable high and low temperature controls with audible alarms and visual indicators.
Linear LEDs with intensity programable up to 120 micromoles/m2/sec, and dimmable between 5-100% output.
New IntellusUltra™ control system with 24-hour monitoring and programing of all variables with step and ramp modes up to 500 steps and 50 total programs.
Ultrasonic humidifier & dehumidifier with electronic RH sensor ± 10% RH.
NEMA 5-15R 1A convivence receptacles, programable via Intellus Control System.
Warranty of two (2) years parts and labor.
File details come from the government source that posted it. Updated .