Attachment 1 SOW- XRF Analyzers FY21 FINAL.pdf
PDF 97 KB Posted
- Attached to
- Amendment 1 - Combined Synopsis/Solicitation - XRFs Brand Name or Equal Federal contract opportunity
- Solicitation number
- CPS-2120-21-0001
- Issued by
- Consumer Product Safety Commission
About this file
This document is a combined synopsis and solicitation from the Consumer Product Safety Commission seeking quotes for handheld X-ray fluorescence analyzers. The CPSC requires sixteen analyzers to screen consumer products for lead and other elements to ensure compliance with safety standards. Offerors must demonstrate that their analyzers meet technical requirements for detecting low levels of targeted elements in materials like plastic, metal, glass, and paint. Quotes are due by August 10th and must include pricing for twelve base analyzers and four optional additional units, with optional pricing for up to sixteen on-site training sessions. Live demonstrations of analyzer performance will take place the week of August 2nd to confirm conformance with mandatory technical criteria. Award will be made to the lowest priced technically acceptable offeror by September 10th with delivery required within thirty days.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment 2 XRF Technical Evaluation Plan FY21.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Consumer Product Safety Commission Statement of Work
Handheld X-ray Fluorescence (XRF) Analyzer
SECTION I: Description of XRF Requirements
The portable Handheld XRF analyzers must be capable of measuring lead (Pb) content in children’s products to determine compliance with the Consumer Product Safety Improvement
Act (CPSIA). The XRF must meet the requirements outlined for the testing apparatus in the
ASTM F2617 Standard Test Method for Identification and Quantification of Chromium, Bromine, Cadmium, Mercury, and Lead in Polymeric Material Using Energy Dispersive X-ray Spectrometry. Units must be suitable for use as a handheld analyzer for the measurement of lead, cadmium, arsenic, antimony, selenium, barium, chromium, and mercury in consumer products including metals, plastic, glass, and paint. The XRF shall be capable of effectively measuring elements with atomic number at least as low as phosphorus (P) without the use of vacuum or helium purge.
SECTION II: Background
The CPSC laboratory, field, and import surveillance staff require XRF analyzers that are critically needed to screen consumer products for lead (Pb) and other elements of interest.
SECTION III: Equipment Requirements and Performance
The XRF analyzer shall be new and a model currently in production. Refurbished or demonstrator instruments are not acceptable. The XRF analyzer shall be robust and durable for daily use in both a laboratory and field setting and must be suitable for continuous local and/or long-distance transport.
The XRF analyzer system shall meet or exceed the following functions and specifications:
A. Instrument shall operate with a 50 kV X-ray tube, and shall not employ a radioactive isotope source. X-ray detector with energy resolution equal to or better than 185 eV full width at half maximum of the Mn Kα line.
B. Instrument shall contain a Silicon Drift Detector (SDD) capable of effectively measuring elements with atomic number at least as low as phosphorus (P) without the use of vacuum or helium purge. Should have the ability to process over 180,000 detector events per second.
C. Instrument shall include a built-in camera to provide close up view of analysis area.
D. Instrument shall be able to achieve a minimum 3mm small spot size with the capability of a larger spot size for sample data acquisition.
E. Instrument shall include a minimum of two (2) Li-ion batteries, each capable of powering the XRF analyzer at least 4 hrs. without external power and include a battery charger and accessories needed to recharge batteries; AC adapter.
F. Instrument shall include a portable fully shielded test stand and a clip-on backscatter shield. The stand should collapse easily without tools to allow hand-carrying into the field. The stand shall allow tool free setup, attachment and removal of analyzer.
G. Instrument shall include preprogrammed software modes that will accurately quantitate lead and other elements in a variety of homogeneous matrices such as metal, plastic, and glass and a mode for user-definable calibrations and analyses. The analyzer must include modes for quantitative determination of concentrations in percent (%) and parts per million (ppm) of target elements in plastic, metal, and glass.
H. The preprogrammed software modes shall include a specific “test all” mode (or equivalent) for sample analysis. The “test all” mode (or equivalent) will automatically select the mode for the user when the user is uncertain of the appropriate mode to choose. The “test all” mode (or equivalent) will determine the sample matrix (i.e.
metal, plastic, painted metal, PVC plastic) and apply the appropriate algorithms for elemental detection and quantification.
I. Instrument shall be capable of detecting Pb, Sb, As, Ba, Cd, Cr, Hg, Se, Br, Cl in paint layers on painted consumer products. The analyzer must be able to detect < 80 parts per million (ppm) Pb on 20-micron thick paint layer. The analyzer must be able to detect and measure Pb on mass per unit area basis using the National Institute and Standards Technology (NIST) Standard Reference Material® 2569 Level 2 and
Level 3 Lead Paint Films.
J. Instrument shall be capable of quantifying and detecting Pb, Sb, As, Ba, Cd, Cr, Hg, Se, Br, Cl in metal, plastic, glass, rubber, textiles. The analyzer must be able to detect
< 20 ppm Pb in plastics. The analyzer must be able to detect and accurately measure
Pb on mass fraction basis in plastic certified reference materials that contain low and high levels of Pb such as the European Reference Material ERM®-EC680k, ERM®-
EC681k and the NIST Standard Reference Material® 2861 PVC plastic. The analyzer must be able to detect < 70 ppm Pb in metal alloys such as zinc, tin, copper, steel.
K. Instrument shall be able to detect and accurately measure Pb on mass fraction basis in glass certified reference materials that contain high levels of Pb such as the
Restriction of Hazardous Substances (RoHS) 2-3 Glass Standard.
L. The instrument display shall include a touch screen display which allows user input and reading of results (within seconds) at the point of use. The analyzer shall contain user-friendly, graphical interface software compatible with Microsoft Windows 10 or later. For data handling, the analyzer must include means for assigning sample numbers to test data, store, transfer, and retrieve reports and sample numbers securely including transfer of data to Windows PC.
M. Instrument shall operate with a stand-alone laptop computer and include a PC connection cable.
N. Instrument shall include calibration check standards (a minimum of one metal alloy standard and one plastic standard material) for calibration check.
O. Instrument shall have password protection capabilities in the software for added security for appropriate personnel use.
P. Instrument shall include a carrying case to safely transport XRF analyzer.
Q. Maximum weight of analyzer with battery shall not exceed 4 lbs.
SECTION IV: Requirements
The contractor shall perform the following:
A. Technical documentation:
• The contractor shall provide a Certificate of Calibration with each analyzer purchased. The Certificate of Calibration shall contain all the measurement readings of reference materials the vendor used for calibration verification.
• The contractor shall provide a radiation safety certificate with the instrument stating that the instrument meets all federal and state radiation safety requirements.
• The contractor shall provide a User Manual with the analyzer.
B. Maintenance:
• The contractor shall provide a quote for the cost of repair and/or calibration of the XRF analyzer. In addition, a loaner XRF analyzer from the contractor shall be made available to CPSC free of charge during any repair incident and shall be the similar make/model of the CPSC XRF analyzer being repaired.
• The vendor must be able to provide local service support.
C. (Optional Item) Training: The contractor shall provide a minimum ½ day on-site training that includes: operation of unit, routine maintenance, basic application, and safety. On-site training may require a maximum of sixteen (16) separate locations throughout the continental United States. If exercised, this training will be scheduled and delivered within one year of contract award.
D. Warranty: The contractor shall provide a minimum two (2) year warranty for the instrument that covers all parts, labor, and travel. The government reserves the option to negotiate an extended warranty plan.
SECTION V: Live Demonstration
If the contractor confirms they can satisfy all of the requirements noted in the Statement of
Work, they shall provide the CPSC with a live demonstration to validate the performance of the
XRF analyzer based on Technical Evaluation Factors A and B described in Attachment 2. The demonstration will take place at the CPSC’s National Product Testing and Evaluation Center
(NPTEC) located at 5 Research Place (5RP) in Rockville, MD. All technical and performance requirements are mandatory and must be met in order to be eligible for award. Interested vendors must email Greg Grayson at ggrayson@cpsc.gov to schedule their live demonstration.
Demonstrations will be held on or between August 2nd and August 6th 2020. This demonstration will test Technical Evaluation Factors A and B as described in Attachment 2.
SECTION VI: Instructions to Offerors
Information requested herein must be furnished in writing and be fully and completely in compliance with RFQ instructions. The information requested and the manner of submission is essential to permit the prompt evaluation of all quotes on a fair and uniform basis. Simple statements of compliance without detailed specifications that allow the Government to compare a quoted product to the Government’s requirement may not be considered sufficient evidence that the Quoter’s product can meet the requirement.
mailto:ggrayson@cpsc.gov
Any additional clauses required by public law, executive order, or acquisition regulation, in effect at the time of execution of the proposed contract, will be included. It is understood that your quote will become part of the official contract file.
This solicitation does not commit the Government to pay any cost for the preparation and submission of a quote. In addition, the Contracting Officer is the only individual who can legally commit the Government to the expenditure of public funds in connection with the proposed acquisition.
Each offeror should recognize that its initial price and technical quote may be used as the sole and final basis for award and should quote accordingly. The contractor must submit their quote in two separate volumes:
Volume I: Technical Specifications
The contractor must provide technical specifications for their quoted analyzer showing that it meets or exceeds the standards outlined in this Statement of Work. Vendors must also provide brief descriptions of available on-site and remote trainings for end users.
Volume II: Pricing
The vendor must provide pricing for sixteen (16) quoted handheld XRFs. It must provide firm fixed price optional pricing for up to sixteen (16) on-site training sessions (see section IV(c) of this SOW). Discounts against established prices and rates are requested.