PURCHASE DESCRIPTION

1. Statement of Need and Purpose: Purchase of bench top Flow Cytometer laboratory system with a high sensitivity detector is required for analysis of aspect ratio, size, and fluorescent intensity of cells and isolated mitochondria in a 96 well format, at a high flow rate (up to 20,000 events/s). The unique design required for this instrument should include use of a CCD camera with lasers, as opposed to the less sensitive photomultiplier tubes. The increased sensitivity is needed in particular for analysis of isolated mitochondria in which signals are dim. This CCD camera design enables high sensitivity detection which is a requirement for this instrument. The instrument is needed to determine the effect of Parkinson’s disease-causing genes on mitochondrial function, consistent with the mission of the Inherited Movement Disorders Unit in the Neurogenetics Branch of the NIH.

2. Background Information and Objective: Will need an instrument that can analyze aspect ratio, size, and fluorescent intensity of cells and isolated mitochondria in a 96 well format, at a high flow rate (up to 20,000 events/s). It will enable assessment of mitophagy, mitochondrial reactive oxygen species production, ATP production, among other critical mitochondrial functions. This will aid the research being done by the Inherited Movement Disorders Unit (IMDU) in the Neurogenetics Branch of the NINDS.

They have found that specific genes causing Parkinson’s disease influence the degradation and function of mitochondria. These processes can be monitored in an unbiased way by flow cytometry. Some assays require analysis of isolated mitochondria.

These place special demands on the flow cytometer as the signal is too dim to be detected by conventional flow cytometers that use photo-multiplier tubes. By contrast, the systems with CCD based camera detection have the necessary sensitivity to make these measurements reliably. Additionally, the CCD-camera based system can provide information on parameters like particle size and aspect ratio which will greatly facilitate IMDUs line of research.

3. Generic Name of Product: Flow Cytometry System

4. Purchase Description: Brand Name or Equal

(1) Specify trade name: Amnis CellStream Flow Cytometry System

(2) Specify product’s brand name: Amnis CellStream Flow Cytometry System

(3) Specify manufacturer's name: Luminex

(3) Specify model, type, catalog, and/or part number(s):

Model, trade name: Amnis CellStream Flow Cytometry System

Catalog #:CS-100196, CS-200561, CS-200405, CS-500200, CS-600200, 2 Year Warranty.

5. Salient characteristics:

The vendor shall provide a Flow Cytometry System brand name or equal to Amnis CellStream Flow Cytometry System. The instrument shall have the following specifications:

Version 11-30-2016-1 FAR Part 11—Describing Agency Needs

• Three lasers exciting at 405 nm, 488 nm, and 561 nm and be able to detect 15 fluorescent parameters.

• Use CCD camera

• Must be able to detect small particles (< 100 nm for fluorescent particles)

• Must have a flow rates capable of reaching 20,000 event/sec.

• Ability to sample single tubes and 96 well plates must be incorporated into the instrument.

• Must fit on a benchtop with a total footprint no larger than 440 mm x 625 mm x

495 mm.

• Must be able to detect particle area and aspect ratio as parameters.

• Must allow real time visualization of samples in each of the detector’s channels.

• Must be able to provide absolute cell counts without reference beads.

• Must have built in capability of analyzing 96 well plates.

• Must have a detector with sensitivity less than 10 molecules of equivalent soluble fluorochrome (MESF) for FITC and less than 5 MESF for PE.

• Detector much have 7 decades of dynamic range.

• Must use PBS buffer as sheath fluid.

• Must allow one lick operation for start, calibration, and cleaning.

• Must be field upgradable.

• Unit must have a 2-year parts and labor warranty

4. Quantity: 1 System

5. Delivery Date: delivery date 8 weeks after receipt of order.

6. Period of Performance: N/A

7. Option(s): N/A