ATCH Salient Characteristics.pdf
PDF 129 KB Posted
- Attached to
- Hyperion Ion Source Federal contract opportunity
- Solicitation number
- FA7022-21-Q-0001
About this file
This document contains details on a sole source federal contract opportunity for a Hyperion Ion Source from Cameca Instruments, Inc. The Department of the Air Force Air Combat Command is seeking to procure a high brightness RF-plasma oxygen ion source to replace the duoplasmatron on several of its Cameca IMS 1280 mass spectrometer instruments. The ion source and associated equipment, installation, training, and warranty are being sourced directly from Cameca Instruments as the sole manufacturer and provider of this equipment for their IMS-1280 line. Pricing and other details on the ion source and its capabilities are provided to describe the requirements of this non-competitive sole source procurement.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Public Release Combined Synopsis.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
CAMECA IMS 1280 / 1280-HR/1270 E7
Item QTY Unit Price Total Price
O- RF PLASMA ION SOURCE 1 $ $
Salient Characteristics below
High brightness RF-plasma oxygen ion source to be installed in place of the duoplasmatron on the CAMECA IMS 1280 / 1280-HR/1270 E7 Including ion source, leak valve, power supplies and rack, software, users guide, and dedicated cooling system.
SPECIFICATIONS:
High brightness RF-plasma oxygen ion source delivering beams of O-, O2- and O2+.
Note: all specifications given below are performed with source HV = -13 kV and sample HV = +10 kV.
Maximum beam current:
O-: Ip > 1 μA Ip: primary beam intensity (measured in the primary Faraday cup)
Beam current stability (peak to peak):
O-: ΔI/I < 3 % over 10 hours, with 3σ rejection
Beam density:
For a beam diameter in the range [5 μm, 30 μm]:
O-: > 100 mA/cm2 O2-: > 20 mA/cm2
Minimum beam diameter (16 - 84 % definition):
O-: ≤ 0.7 μm
Installation of the source and 3 days of training included in price.
1 year warranty included in price.