This document contains a draft technical specification for a UV-NIR spectrometer. The specification includes requirements for the spectral range, resolving power, CCD resolution, dynamic range, sample rate, time stamp accuracy, A/D converter, integration time range, wavelength repeatability and accuracy, radiometric repeatability and sensitivity, linearity, stray light, order sorting filter, optical ND filter, lens, operating voltage, electronic emission compliance, operating environment, detector cooling, instrument mounting, and sighting scope. Compliance with these specifications would be verified through vendor acceptance testing according to an acceptance test plan and report to be provided two weeks and with delivery respectively. The related federal contract opportunity is a sources sought notice for a JSIS UV-NIR spectrometer issued by the Department of the Air Force Materiel Command Test Center.

View the file

Other files for this federal contract opportunity

Other files attached to JSIS UV-NIR Spectrometer, newest first.
File Type Posted
Atch 1 - UV-NIR Spectrometer Draft Spec.pdf PDF
FA9101-22-Q-B028 Spectrometer Sources Sought.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

APPENDIX A TECHNICAL SUMMARY

Table 1: Spectrometer Specifications

Detector and Wavelength Characteristics Compliance

(YES/NO)

Spectral Range Delivered suite: 255 – 1100 nm Capability of Optics and detectors: 190 – 1100 nm

Resolving Power (w/ a 48 µm entrance slit)

≥ 2760 at 315.7nm ≥ 780 at 589.3nm

CCD Resolution (w/ a 48 µm entrance slit)

≤ 0.12 nm at 315.7nm ≤ 0.76 nm at 589.3nm and 768.2nm

Dynamic Range ≥ 10,000

Sample Rate 500 Hz at full spectral range

Electronics

Time Stamp ≤ 1msec accuracy

A/D Converter ≥ 16-bit

Integration Time independently adjustable from 1msec – 1 sec /frame rate

Performance

Wavelength Repeatability ≤ 0.03 nm

Wavelength Accuracy ≤ ± 0.2 nm

Radiometric Repeatability ≤ ±1% of full-scale reading

Radiometric Sensitivity Maximum NESI for 2ms integration and 16um slit width

Wavelength (nm) NESI (W/m2/nm)

260 1.60×10-6

360 1.80×10-7

400 5.20×10-8

600 4.00×10-8

700 4.00×10-8

1100 4.00×10-7

Linearity ±5% of reading from 5% to 95% full A/D scale

Stray Light ≤ 1×10-4 of total transmission for assembly

Order Sorting Filter Include if necessary

Optical

ND Filter Integrated filter wheel with ND open, 1, 2, 3, and closed

Lens Government provided C- or T-mount lens

General

Operating Voltage 120VAC ± 10% at 60Hz ± 3 Hz with a maximum current of 6 amperes

Electronic Emission CE or FCC (Part 15) compliant

Operating Environment -12°C to 20°C at 1 to 80% relative humidity, non-condensing

Detector Cooling electrically cooled, operable within 15min upon initial power

Interchangeability Changing of slits, gratings, and optics possible by user

Instrument Mounting integrated 1/4in Ø, 20 TPI holes, capable of tripod mounting

Sighting Scope Adjustable co-aligned visible display or integrated sighting scope to accurately aim the unit. Nominal boresight distance 280m.

Statement of Volatility include Statement of Volatility or procedure for memory wipe

Atch 2 FA9101-22-Q-B028 Sources Sought

Verification Compliance

(YES/NO)

Vendor Acceptance Testing

Verify compliance with the specification before delivery through test, demonstration, analysis, or inspection

Acceptance Test Plan Provide Acceptance Test Plan 2 weeks before test

Acceptance Test Report Provide Acceptance Test Report with Instrument delivery

Atch 2 FA9101-22-Q-B028 Sources Sought