Atch 1 JSIS UV Calibration Gap Specifications.pdf

PDF 156 KB Posted

Attached to
FA9101-24-Q-B012 – UV-NIR System Spectral Response (SSR) Optical Bench Federal contract opportunity
Solicitation number
FA9101-24-Q-B012
Issued by
Department of the Air Force Materiel Command Test Center

About this file

This document contains specifications for an imaging spectrometer system and details of a related federal contract opportunity. The specifications require a charge-coupled device imaging sensor with dimensions under 700 by 500 by 250 millimeters, focal length over 500 millimeters, and capabilities to eliminate re-diffracted light. The system must include computer-controlled grating turrets and slits, as well as filter wheels holding order sorting long pass filters with cut-ons at 325, 385, and 630 nanometers. Control and data acquisition software compatible with Windows 10 is required along with cabling. The related solicitation seeks proposals for a UV-NIR system spectral response optical bench and references an attachment for details. The opportunity is issued by the Department of the Air Force Materiel Command Test Center.

View the file

Other files for this federal contract opportunity

Other files attached to FA9101-24-Q-B012 – UV-NIR System Spectral Response (SSR) Optical Bench, newest first.
File Type Posted
Atch 2 FA910124QB012 Model Contract.pdf PDF
FA9101-24-Q-B012 - Combined Synopsis Solicitation.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

JSIS UV Calibra�on Gap Equipment

Provide an Imaging Spectrometer with the following specifica�ons.

• Charge-coupled device (CCD) imaging sensor

• Dimensions o Length: less than 700 millimeters o Width: less than 500 millimeters o Height: less than 250 millimeters

• Focal length greater than 500 mm

• Designed to eliminate re-diffracted light

• Computer controlled, kinema�c triple gra�ng turret with on-axis gra�ng rota�on

• Gra�ngs to be included:

o 1800 gr/mm classically ruled gra�ng at 300 nm o 1200 gr/mm blazed holographic gra�ng at 500 nm o 2400 gr/mm blazed holographic gra�ng at 400 nm

• Wavelength accuracy +/- 0.2 nm

• Purge port

• Computer controlled front, side, and exit slits that can be varied from 0 to 7 millimeters

• Motorized, computer controlled 6 posi�on filter wheel for up to 6 each 1" diameter filters.

• Three filters to be included:

o Order sor�ng long pass filter with 325 nm (+/- 2 nm) cut-on; colored glass; 1" diameter o Order sor�ng long pass filter with 385 nm (+/- 2 nm) cut-on; colored glass; 1" diameter o Order sor�ng long pass filter with 630 nm (+/- 2 nm) cut-on; colored glass; 1" diameter

• So�ware to provide complete instrument control and data acquisi�on. Must be compa�ble with Windows 10.

• Appropriate cabling to facilitate instrument control and data acquisi�on between instrument and AEDC provided computer.

JSIS UV Calibration Gap Equipment