This document outlines the salient characteristics for a digital radiography system solicited under federal contract opportunity 140G0223Q0191 by the Department of the Interior US Geological Survey Office of Acquisitions and Grants. The system must include an x-ray generator with a minimum 32kW power and 125kVp peak voltage capable of exposures from 3-6300ms. The detector must be a minimum 17"x17" flat panel with at least 3072x3072 pixels and 14-bit depth. Additional requirements include a minimum 21" touchscreen display, DICOM compatibility, anatomical programming functions, and table dimensions sufficient to support over 400 pounds. The system assembly and quality control must be performed in the United States and be capable of diagnostic-quality images from 10-100 pounds.

View the file

Other files for this federal contract opportunity

Other files attached to DIGITAL RADIOGRAPHY, newest first.
File Type Posted
B08_Atch_A_0002.pdf PDF
Sol_140G0223Q0191_Amd_0002.pdf PDF
Sol_140G0223Q0191_Amd_0001.pdf PDF
Sol_140G0223Q0191.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

1) System

a. Assembly and final Quality Control approval MUST be done in the United States

b. MUST provide detailed, diagnostic-quality images of small patients 10 – 25 grams and also medium sized patients up to 100 lb.

2) Table

a. Hands-free collimator light control

b. 4-way float top table

c. Minimum Patient table Dimensions: 60” x 24”

d. Minimum total dimensions: 60”W x 36”D x 77”H

e. Weight: can withstand >400 lb

f. CE certification

3) Generator - High Frequency

a. 32kW / 400mA 125kVp

b. Adjustable between 25-400mA

c. Adjustable between 40-125kVp in 1kvP increments

d. Adjustable between 3-6300ms

e. Generator consolidated into table base. No satellite components

f. Must be Mono-Bloc technology:

i. 32kW rotating anode monobloc

ii. 320 mA–100 kV @100ms

g. Operates from 110V, 20A power. Does not require dedicated 220V electrical

h. Frequency: 50/60 Hz

i. 2-point and 3 -point Anatomical Programming Radiography (APR) capabilities

j. Capacitor load voltage and inverter control

k. Overheat Alarm

l. X-ray emission warning light and alarm

4) Detector - Flat Panel

a. at least 17” x 17” Tethered

b. Cesium CSi (NOT GADOX)

c. Amorphous Silicon

d. Pixels: minimum 3072x3072

e. Maximum Pixel Pitch:145 um

f. Effective Area: minimum 16” x 16”

g. Detective Quantum efficacy (DQE): minimum 65

h. Bit depth: Minimum bit depth 14

i. Dynamic Range: minimum capability of 0-16000

j. Auto Exposure Detection

k. CANNOT be charge coupled

6) Software

a. Built-in lookup table

b. Built-in Histogram charting

c. Built-in Gray scale graphing

d. DICOM 3.0 compatible

e. Compatible with Microsoft operating system (PC workstation not included)

f. Stitching functionality

g. GPR (Grid Pattern Removal) function

h. Anatomical Programming (APR)

i. Built-in Exotic species template with user-adjustable customization and memory

7) Touchscreen display -

a. Minimum 21” Full HD “Heads Up” Touchscreen

b. Resolution: minimum 1920x1080p

d. Minimum 50M:1 Dynamic Contrast Ratio

e. Can be operated with nitrile, and/or vinyl medical gloves

8) Operation

a. Allows a single person can take radiographs without stepping away from table or patient

i. Touchscreen selection of anatomical programming

ii. Hands-free collimation

iii. Easy patient positioning using the 4-way float function

iv. Preview images in <10 second processing time

v. Save image directly from touchscreen

9) Support MUST be:

i. Based in the US

ii. Communication to be exclusively in English, available by voice, email and/or chat and in-person when required

10) Trade-in – Total cost of bid may include trade-in value credit for the following equipment (subject to cosmetic and operational condition as determined by inspection by bidder):

a. Poskom PXP-40HF Portable X-Ray (USGS Property #494671)

b. Portable X-Ray Mobile Stand

c. Summit Portable X-Ray Mobile Table

d. Fire CR X-Ray Processor and 2 14” cassettes

File details come from the government source that posted it. Updated .