80TECH22Q0037 Statement Of Work CW22 Optimus Rev1.pdf
PDF 123 KB Posted
- Attached to
- Electronically Scanned Pressured System Components Federal contract opportunity
- Solicitation number
- 80TECH22Q0037
About this file
This document is a solicitation from the National Aeronautics and Space Administration (NASA) for an electronically scanned pressure system and components. NASA requires an optimized pressure measurement system for its Cleveland, OH research facility including Measurement Specialties 32HD pressure scanners, an 8432 pressure calibrator controller, an OSPR system processor, OFIU fiber optic interfaces, RP remote processors, RPS remote power supplies, MSDI scanner digitizing interfaces, and associated cables. The vendor must be able to provide all listed components and specifications. Quotes are due by September 19, 2022 with an anticipated award date of a firm fixed price contract. Delivery is requested by July 30, 2023. The solicitation will be evaluated based on technical acceptability and price, with the lowest priced technically acceptable proposal to receive the award.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SF30_80TECH22Q0037-9.9.22.pdf | ||
| JOFOC PR805755_807255 signed 8.23.pdf | ||
| 1707-4200805755-Hammett-signedbyCIO.pdf | ||
| 80TECH22Q0037 RFQ Order Enclosure 4- Pricing Exhibit.xlsx | XLSX spreadsheet | |
| 80TECH22Q0037 RFQ Order Enclosures-1_2_3.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK CW22 Optimus
1. Objective/Description The Optimus system is meant to replace the existing pressure measurement system in the CW22 test facility at NASA Glenn Research Center in Cleveland, Oh.
2. Characteristics, Scope, and Specs/Requirements
System shall support the following pressure scanners of range 15 psid
a. Intelligent Pressure Scanner, TE Connectivity Model 32HD Mini Module or fully equivalent with the following salient specifications:
Compatible for use with TE Connectivity MSDI (mini scanner digitizing unit) via OSCB cable.
Scanner shall contain an on board microprocessor and 32 pressure transducers, each with internal EEPROM memory chip for storing factory calibration data.
Unit shall be digitally compensated for zero, span, nonlinearity and temperature effects.
Scanner shall be configured for full scale measurement range from a common reference input.
Transducer accuracy, after re-zero, +/-0.03% full scale or better.
Transducers shall be able to withstand 2X over pressure of full scale range without reducing sensor performance (Proof Pressure).
Able to withstand 3X over pressure of full scale range without damaging sensor (Burst Pressure).
Scanner shall have an operating range of 0 to 50 degrees C, the transducer reading shall be internally compensated to correct for temperature effects over this operating range of 0 to 50 deg C with a total thermal error of +/-0.001% F.S./deg C.
Scanner shall provide front access for input of the pressure measurements.
Scanner shall have 0.063” bulged tubulation fittings mounted on a quick disconnect plate for input of pressure measurements.
Scanners shall have internal manifold valving and pneumatic calibration and control pressure inputs that allow for on-line pneumatic re-zero & span calibrations, and on-line leak checks of the pressure measurement transducers.
Scanners shall be capable of reading rack source and purge pressures.
Scanners shall support dual range calibration.
System shall support the following controllers of range 30 psia
b. Intelligent Pressure Calibrator, TE Connectivity Model 8432 or fully equivalent with the following salient specifications:
Compatible for use with Optimus and 8400 Systems via the Remote Processor unit (part number 8404.)
Able to perform calibration of Optimus mini modules (32HD and 64HD models) When used to calibrate Optimus mini modules, able to provide system accuracy of greater than +/- 0.03% FS.
Able to perform Auto Accuracy Tests on the mini modules.
Controller shall utilize resonant quartz pressure standards for absolute measurements and digital pressure standards for differential measurements.
Contain a pressure relief valve designed to vent at 110% of the instruments full scale pressure.
Controller shall provide known stable pressures to electronic pressure scanners for periodic calibration or validation.
c. System shall support the following system controller System Processor, TE
Connectivity Model OSPR or fully equivalent with the following salient specifications:
Compatible for use with TE Connectivity 32HD and 64HD mini modules Able to control up to 32 mini modules for a maximum of 2048 data channels.
Able to interface with TE Connectivity MSDI unit via OFIU (Optical Fiber Interface Unit) installed at the factory.
Able to control TE Connectivity PCU to calibrate Optimus mini modules, able to provide system accuracy of greater than +/- 0.03% FS.
Interface with existing COBRA data systems in GRC facilities via in-house software and hardware trigger input on back panel.
d. System shall support the following fiber interface units:
Optical Fiber Interface Unit, TE Connectivity Model OFIU or fully equivalent with the following salient specifications:
Able to provide noise free data communication via 62.5 micron multimode fiber cable between the OSP and the MSDI scanner interface.
Compatible for use with TE Connectivity 32HD and 64HD mini modules vis MSDI over fiber connection.
Able to support simultaneous pressure and force balance measurements.
e. System shall support the following remote processing units:
Remote Processor, TE Connectivity Model RP or fully equivalent with the following salient specifications:
Able to interface with TE Connectivity OSPR System Processor via RS-485/ GPIB interface.
Up to three RP units can be controlled by one OSPR unit.
Able to interface with TE Connectivity PCU Pressure Controller Unit.
Rack mountable unit. Can be placed in close proximity to mini modules to allow for short calibration tubing runs and therefore reduce calibration pressure settling time.
Able to support simultaneous pressure and force balance measurements.
f. System shall support the following remote power supply units:
Remote Power Supply, TE Connectivity Model RPS or fully equivalent with the following salient specifications:
Able to provide closely regulated power to TE Connectivity MSDI unit.
Able to provide +12 VDC, -12 VCD, and +5 VDC rail to TE Connectivity/ 32HD and 64HD mini modules.
Also compatible with legacy 8400 system.
g. System shall support the following miniature scanner digitizing interfaces:
Miniature Scanner Digitizing interface, TE Connectivity Model MSDI or fully equivalent with the following salient specifications:
Able to provide regulated power to TE Connectivity 32HD pressure scanning mini modules.
Able to provide excitation feedback measurement on the mini modules.
Able to conduct analog to digital (A/D) conversion of signals coming from TE Connectivity models 32HD and or 64HD miniature pressure scanning units.
Small size and low power consumption to allow for placement inside rack mounted system near the mini modules.
Cross sectional area designed to match the 32HD and 64HD mini modules.
Able to interface with TE Connectivity OFIU via fiber optic cable.
Able to support connection of up to eight mini modules and able to scan a maximum of 512 pressure data channels.
h. System shall require the following cables:
8476 MSDI power cable or fully equivalent with the following salient specifications:
Able to provide regulated power to TE Connectivity MSDI, miniature scanner digitizing interface
OPCY power cable splitter or fully equivalent with the following salient specifications:
Able to split 8476 power cable to accommodate two MSDI units
8484 GPIB interface cable or fully equivalent with the following salient specifications:
Able to connect Optimus OSPR with Optimus remote rack Able to send serial GPIB communications between the two units
OSCB scanner cable or fully equivalent with the following salient specifications:
Able to connect 32HD miniature pressure scanner to MSDI miniature scanner digitizing interface Able to send regulated power to the 32HD miniature pressure scanner Able to send analog signals back to the MSDI from the 32HD Able to carry digital channel selection signal to the 32HD miniature pressure scanner.
-Warranty 12 months from date of shipment
-Install
Install to be handled in-house by facility engineers and technicians
-Maintenance Services Maintenance to be handled in-house by facility engineers and technicians
The requirement will be used in a pressure measurement system.
List of the part numbers, quantity, and a brief description listed below:
Item Description Quantity OSPR-0100000000 Optimus System Processor 1 OFIU-01000000 Optimus Fiber Optic Interface 4 MSDI-00000000 Miniature Scanner Interface 4 8432-0300000000 PCU 30 psia 1 8482-0000000000 Blanking Unit 6 8404-0000010000 Remote Processor 1 8491-0100000000 Remote Power Supply 2
8476-0010000001 Power Cable, 8491 to mSDI, 100 ft 2
OPCY-
0000000303 Power Cable for two mSDI 2 8484-0000300000 Interface Cable, OSP to RP, 3 ft 1
OSCB-
0000000030 Optimus Scanner Cable, 2 ft 12 32HD-0723001220 Pressure Scanner 32 Ports 6
OPCY-
0000000303
Power Cable splitter for two mSDI 1
8484-0000300000 Interface Cable, OSP to RP, 3 ft 1
File details come from the government source that posted it. Updated .