Statement of Work: Microanalysis system with Silicon Drift Detector for FE-SEM
We are looking to purchase microanalysis systems for our Hitachi S-4700 FE-SEMS. System shall meet or exceed specifications for the AZtecEnergy Standard Microanalysis System with X-MaxN 20 large area Analytical Silicon Drift Detector. One system must include large area mapping capability (for the 2003 Hitachi S-4700).
Microanalysis system shall meet or exceed:
• Be compatible with Hitachi S-4700 (1997, serial 5703-03 and 2003, serial 6059-03)
• Silicon Drift Detector size of at least 20mm2
• Excellent low energy analysis, including detection of Be and above
• Solid angle of 0.00626sr or greater
• Resolution specs shown at installation (FWHM) in accordance with ISO15632:2002 o Mn Ka – 127eV at 50kcps o F Ka – 58eV at 50kcps o C Ka – 56eV at 50kcps
• Ability to acquire images up to 8k x 8k resolution.
• Can acquire multiple electron images for region of interest.
o Collect multiple fields over large map areas, set up though intelligent set up wizard o Designed for flat and tilted samples, both stage tilt and pre-tilt o Collect up to 1500 individual maps, at up to 4k x4k EBSD resolution and align o them automatically o Field alignment through image correlation routine o Montage into a single dataset with up to 64 million data points o Investigate montage data as a single site of interest
• Incorporates Filtered Least Squares spectrum processing without the user making background fitting adjustments
• Utilizes XPP matrix correction
• Can perform un-normalized quantitative spectrum analysis.
• Can automatically correct for non-linearity of peak resolution and position.
• Sum peak correction not only eliminates sum peaks, but replaces counts to the corresponding component peaks in the spectrum.
• Ability to collect element maps at resolutions up to 4k x 4k.