Specifications for Space Simulation Facility 35” Cryopump
Cryo Pump Specification
Inlet Flange
Roughing Flange
Regeneration Purge Tube
Volume Flow Rate (Liters/second)
Capacity at 5x10^-6 Torr
Crossover
Cool down Time
Instrumentation
Mounting Orientation
Helium compressor compatibility
Helium Compressor Specification
Input Power Required
Air Cooled
Remote Start/Stop capability
Feed System Specification
Supply Liquid N2 to Cryo pumps
LN2 insulated lines
35” ANSI
2” ANSI / 2”ASA
0.5”
N2 – 30,000 H2 – 20,000
N2 – 6,000 std Liters H2 – 30 std Liters
2,000 Torr Liters
240 min to 20K
Silicon diode temperature sensor
Any orientation must be compatible with CBST6.0 Helium compressor
480V, 3 Phase, 5.5 kW
<95°F Ambient
Provide adequate LN2 flow to insure robust operation of cryo pumps
Vacuum insulated lines to insure minimal product loss.
Specifications for Space Simulation Facilities 48" Cryogenic Pumps
48" Cryopump Specification
High Vacuum Flange
Roughing Flange
Regeneration Purge Tube
Volume Flow Rate (Liters/second)
Capacity at 5x10^-6 Torr
Crossover
Cool down Time
Instrumentation
Mounting Orientation
Helium compressor compatibility
Helium Compressor Specification
Input Power Required
Water Cooled
Remote Start/Stop capability
Feed System Specification
Supply Liquid N2 to Cryo pumps
LN2 insulated lines
48” ANSI
2” ANSI / 2”ASA
0.5”
N2–60,000 H2 – 30,000
N2 – 6,000 std Liters H2 – 30 std Liters
2,000 Torr Liters
255 min to 20K
Silicon diode temperature sensor
Any orientation must be compatible with CBST6.0 Helium compressor
480V, 3 Phase, 5.5 kW
1.5 GPM @ 75F.
Provide adequate LN2 flow to insure robust operation of cryo pumps
Vacuum insulated lines to insure minimal product loss.
| 35in Pump Specs |
| 48in Pump Specs |