PURCHASE DESCRIPTION X-Ray Irradiator revised.pdf

PDF 671 KB Posted

Attached to
SOURCES SOUGHT: X-Ray Irradiator Federal contract opportunity
Solicitation number
75N95022Q00226
Issued by
Department of Health and Human Services National Institutes of Health National Institute on Drug Abuse

About this file

This sources sought notice from the Department of Health and Human Services National Institutes of Health National Institute on Aging seeks capability statements from small businesses for the acquisition of an x-ray irradiator. The irradiator must be able to control dosage, target radiation to a specific part of a mouse, provide lethal and sub-lethal radiation, have an environmental chamber and programmable shelf and turntable. Capability statements are due by May 19th, 2022 and must address each technical requirement and provide company information including number of employees, professional qualifications, and government contract vehicles if applicable. The notice seeks to determine availability of qualified small business sources to potentially set aside the requirement and inform acquisition planning.

View the file

Other files for this federal contract opportunity

Other files attached to SOURCES SOUGHT: X-Ray Irradiator, newest first.
File Type Posted
PURCHASE DESCRIPTION X-Ray Irradiator.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Version 11-30-2016-1 FAR Part 11—Describing Agency Needs

PURCHASE DESCRIPTION

1. Statement of Need and Purpose: The National Institute on Aging (NIA) is looking to acquire an x-ray irradiator for irradiating mice and cells for the Translational Gerontology Branch (TGB), Laboratory of Genetics and Genomics (LGG), and Laboratory of Molecular Biology & Immunology (LMBI).

2. Background Information and Objective: The acquisition of an x-ray irradiator would supplement our cesium irradiator which has several security protocols making it a challenge for some individuals to perform work. The x-ray irradiator would free up CMS staff since more users would be able to operate on their own due to less stringent security protocols in place. Having both an x-ray irradiator and cesium irradiator will allow for researchers to do comparison studies and determine if they could ultimately switch over to an x-ray irradiator for their experiments. The irradiator should have the ability to control dosage, target radiation to a specific part of a mouse, provide lethal and sub-lethal radiation, have an environmental chamber and programmable shelf and turntable.

3. Generic Name of Product: X-ray irradiator

4. Purchase Description: Precision X-Ray, Inc. X-Rad320 System including TouchRad Dose Measurement & Control (ASY-01045), adjustable variable collimator (XD1601- 0000), environmental chamber X-Rad 220VAC (ASSY355), external dose measurement (XD1610-0000) & water tank phantom (XD1610-0100), and programmable mototrized shelf with turntable (XD1602-2203)

5. Salient characteristics: The x-ray irradiator shall include a shielded x-ray cabinet with a 4 kW 320 kV tube, choice of coolant, multi-user control, full screen monitoring of specimens, remote diagnostics and software updating, adjustable shelf, dose measurement and control, adjustable variable collimator, environmental chamber, external dose measurement and a programmable motorized shelf with turntable.

6. Quantity: 1

7. Delivery Date: 90 days after receipt of order

8. Period of Performance: N/A

9. Option(s): N/A