Acoustic Doppler Profiler Salient Characteristics
Transducer frequency: 2 mHZ
Must have the capacity for pulse coherent doppler measurements (either included or available as an upgrade to this unit).
Minimum cells: 40
Cell size: option to include 0.1 – 1.0 m cells
Required range (distance): 4 m range with 0.1 m cells; 10 m range with 0.25 m cells
Required range (horizontal current velocity): - 5 m/s to 5 m/s
Maximum blanking distance: 10 cm
Minimum pressure resolution: 0.005 meters
Must be able to be horizontally mounted with a right angled, upward looking sensor head to allow for deployment in very shallow water with minimal flow disturbance.
Maximum case diameter: 85 mm
Able to provide power simultaneously to 2 external sensors: 7-20 VDC, 3.5 mA avg., 6 mA peak
Must be able to record 0-5 volt, 16-bit A/D data from 2 external sensors.