This document outlines the requirements for a stereotaxic alignment system procurement. The federal opportunity solicits a Model 1900 Stereotaxic Alignment Instrument including a head holder assembly with sagittal plate and non-rupture mouse ear bars, a Model 1940 digital manipulator with 1 micron resolution scales and 3-axis control, a universal off-plane adapter, cannula holder, and custom holder. Required documentation includes maintenance manuals, user manuals, and cleaning instructions. Delivery is required by January 31, 2024. The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 23 awarded contract number 36C26324P0316 to David Kopf Instruments for $27,291 on January 17, 2024 to fulfill this requirement.

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Functional/Performance/Specification General Equipment Name: Stereotaxic Alignment System Description & Salient Characteristics Model 1900 Stereotaxic Alignment System

1. Ability to manipulate and refine the position of small animal head.

2. Include stereotaxic head holder with sagittal plate

3. Mouse gas anesthesia head holder

4. Non rupture 60º tip mouse ear bars

5. Two large rails for mounting manipulators and accessories

6. Centering height gauge Model 1940 digital manipulator

1. 1 micron resolution linear glass scales with digital display

2. 3 Axis

3. Quick change tool holder to allow attachment of any stereotaxic alignment tool and maintain exact common vertical centerline position tool-to-tool.

4. Vertical and lateral travel of at least 128 mm

5. Fine adjustment via A/P control knob Required Individual Line Items & Quantity of Items

· 1 EA Model 1900 Stereotaxic Alignment Instrument

a. 1 each Head holder assembly with sagittal plate

b. 1 each Non-rupture 60º tip mouse ear bars

· 1 EADigital Manipulator
· 1 EAUniversal off Plane Adapter
· 1 EACannula Holder
· 1 EACustom holder (2.5 mm)

· (1 each) Maintenance Manuals (CD, PDF or Word)

· (1 each) User Manuals

· (1 each) Cleaning Instructions for all equipment, supplies and accessories Optional Line Items

· Warranty Information

· Delivery Requirements:

· All products should be ordered to arrive on or about January 31, 2024. Please include any shipping costs that are not included in the cost of the line items.