Microscope Operating Endodontic Requirement Specifications.docx

DOCX document 71 KB Posted

Attached to
Microscope, Operating Endocotic Federal contract opportunity
Solicitation number
36C26124Q0228
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 21

About this file

This requirement specifications document outlines the need for a wall mounted endodontic microscope for the new Akaka VA Clinic in Hawaii. The VA Pacific Islands Health Care System requires one Global Surgical A-Series microscope or equal, including all necessary mounting equipment, extension arms, light sources, and accessories. Product delivery must occur by the specified date to the clinic address provided. The selected vendor must coordinate delivery logistics with the VA points of contact and provide on-site installation, training, and waste removal services. Clinical training is mandatory for staff on microscope setup, operation, troubleshooting, and shutdown procedures. The related federal contract opportunity is a solicitation from the Department of Veterans Affairs Veterans Health Administration VISN 21 for one microscope, operating endodontic to fulfill this requirement.

View the file

Other files for this federal contract opportunity

Other files attached to Microscope, Operating Endocotic, newest first.
File Type Posted
36C26124Q0228.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

FY23-108 M8541 Microscope, Operating Endodontic

VA Pacific Islands Health Care System Initial Outfitting, Transition & Activation Requirement Specifications Microscope, Operating Endodontic (JSN: M8541)

VA Pacific Islands Health Care System New Akaka VA Clinic 91-1051 Franklin D. Roosevelt Avenue Kapolei, Hawaii 96707 Introduction The VA Pacific Islands Health Care System has a requirement to purchase Microscope, Operating Endodontic (JSN: M8541) with an authorized vendor to supply and deliver the item(s) identified in Table 1 below. See salient characteristics below. This is a Brand Name or Equal Requirement. This requirement is in support of the activation of New Akaka VA Clinic. The new Akaka VA Clinic has the need for one (1) wall mounted endodontic microscope to be used for performing oral surgery. Global Surgical A- Series microscope is the basis for design.

Table 1 – Package Items Summary

Item #
JSN
Item Description
Part # (basis of design)
Quantity
1
M8541
6-Step A-Series Microscope, AXIS System, Angle arm and couple
M A1006
1
2
M8541
Inclinable binocular, 0-220 degrees, F:160, 10X eyepieces
M A1022-10
1
3
M8541
Multi-focal Lens, 200-350mm optical range
M 1028ML
1
4
M8541
Wall mount, high or low mounting plate, 16" studs, shaft bracket, plate cover
M A730W
1
5
M8541
45 Degree arm, spring arm, extension arm (for use with high wall and ceiling mount)
M A730-45A
1
6
M8541
McDonald Arm,17" support extension arm for microscope system
M A730-717
1
7
M8541
LED light source for A Series Microscope, green and amber (UV) filter
A M801-LED
1
8
M8541
Carr binocular adapter, 45 degree binocular extender
M A1017
1
9
Prof Services
Install, Site Visit, Wall Mount Structure
Prof Services
1

Additional Item Requirements The vendor will provide service manuals that includes theory of operation, maintenance requirements, repair information, schematics, part lists and troubleshooting information in electronic format for use by the VA Biomed Department.

Date of manufacture must not be greater than 2 years before the purchase order date. The equipment must not be used or refurbished and must have the latest manufacturer’s software/hardware updates installed. Any batteries provided with the unit will not be manufactured more than 12 months before the purchase order issue date.

General Conditions

Site Address The address of the Akaka VA Clinic is provided below.

91-1051 Franklin D. Roosevelt Avenue Kapolei, Hawaii 96707

Delivery Schedule Please provide your estimated delivery time frame in your quote.

Schedule and Order Management Post-award, the vendor will coordinate delivery identified in this document with the VA prior to beginning any work (See Section 2.4). Once the dates have been confirmed, modifications to the schedule are subject to written approval by the CO, COR, and the vendor.

Delivery and Receiving Delivery and receipt of the proposed items is anticipated to be directed to the location identified below. Confirmation of delivery location will be provided post-award. To coordinate delivery, contact VA POC, Peter “Kelii” Won at (808)433-0167, peter.won@va.gov; SME-Solutions POC, Soren Nicholson at (719) 761-4387, Soren.Nicholson@sme-solutions.co, and Nathan Chong at (210) 268-7678, nathan.chong@sme-solutions.co.

☒ Akaka VA Clinic (Direct Delivery to address detailed in paragraph 2.1) There is no loading dock available. All delivery vehicles must be equipped with a “Tommy Gate” type lift to facilitate unloading to ground level.

☐ IOT&A Contractor Provided warehouse: Address TBD

FY24-108 JSN: M8541 Microscope, Operating Endodontic

Statement of Work Page 1 of 8

The vendor is required to define the need for a staging area to accommodate item delivery, if needed. Post-award, this information will be confirmed with the vendor.

The delivery of items identified in this document shall take place during normal loading dock business hours which are defined as: 8:15 AM to 3:00 PM Hawaii Standard Time, Monday through Friday, and excluding Federal Holidays.

Delivery trucks will not be permitted to remain at the receiving area. Trucks shall be unloaded, moved from the receiving area, and then brought back to the receiving area if required to haul out any waste, tools, or excess materials.

Labeling of delivered items shall include the VA facilities contract number and VA purchase order number for identification and reference upon receipt of product.

NOTE: Due to the COVID19 situation, the vendor must follow VA guidelines when on VA property. Any individuals who need to enter the building for installation &/or training will be required to wear a mask and be screened by the site staff prior to entry. Please allow time for this in your schedule.

Clean Up and Disposal There are no dumpsters available for vendor use.

The removal of waste and/or excess material shall be conducted through the loading dock area.

Assembly, Installation, and Calibration Required (Yes ☒ No ☐) The vendor is required to provide live clinical training on set up, operation, shut down, and troubleshooting procedures to clinical staff.

2.7.2 The vendor shall coordinate training dates with the on-site Project Manager. Post-award, these dates will be confirmed with the vendor to meet the clinical schedule. The proposed training date will occur after installation of equipment.

2.7.3 The vendor shall provide documentation and/or certification to each staff member that has successfully completed training.

Training Required (Yes ☐ No ☒)

2.7.1 The vendor is not required to provide live clinical training on set up, operation, shut down, and troubleshooting procedures to clinical staff.

Confidentiality and Non-Disclosure This equipment does not connect to a VA network. Therefore, in accordance with Appendix A, Handbook 6500.6 Checklist, block 6, the C&A requirements do not apply to this requirement, and a Security Accreditation Package is not required.

This equipment does not hold or store patient information. Therefore, in accordance with Appendix A, Handbook 6500.6 Checklist block 7, Appendix B and C provisions do not apply.

Salient Characteristics Microscope, Operating Endodontic (JSN: M8541) Total Quantity: 1 ☐ Network Required if Checked ☒ Manufacturer Certified Installation Required if Checked ☐ BMET Training Required if Checked ☐ End User Training Required if Checked

Basis of Design is Global Surgical A- Series MFG P/N: M A1006 Brand Name or Equal Requirement Salient Characteristics:

· System shall include endoscopic microscope, extension arm and wall mounts

· Microscope shall provide magnification range of 2.1 x 19.2

· Microscope shall provide for entire arch and increase magnification for precision and close inspection

· Microscope shall provide for easy movement of head for easy viewing of mouth

· Microscope shall provide for modularity to allow for upgrades and retrogrades as required

· Microscope shall provide inclinable binocular apparatus from 0-220 degrees

· Microscope shall provide for binocular extender to maintain comfortable upright position for reducing strain and fatigue

· Microscope shall provide adjustable (helicoid) eyecups

· Microscope shall provide for Fine Focus capability

· Microscope shall provide multi-focal lens with minimum of 200-350mm optical range

· Microscope shall provide a minimum binocular optical of 160mm

· Must include necessary options to be mounted on a wall

· Shall include extension arm to allow microscope to be mounted on the wall and reach patient

· Light source shall be LED

· Light source shall include green & UV filter options

· Light source shall provide a minimum/average 500 hours of bulb life

Statement of Work Page 1 of 8