36C25518Q9596-000.docx
DOCX document 15 KB Posted
- Attached to
- Topeka Steam Sterilizers Federal contract opportunity
- Solicitation number
- 36C25518Q9596
About this file
36C25518Q9596 36C25518Q9596_1.docx
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
FedBizOpps Sources Sought Notice
CLASSIFICATION CODE
SUBJECT
CONTRACTING OFFICE'S
ZIP-CODE
SOLICITATION NUMBER
RESPONSE DATE (MM-DD-YYYY)
ARCHIVE
DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
SET-ASIDE
NAICS CODE
CONTRACTING OFFICE
ADDRESS
POINT OF CONTACT
(POC Information Automatically Filled from User Profile Unless Entered)
DESCRIPTION
See Attachment
AGENCY'S URL
URL DESCRIPTION
AGENCY CONTACT'S EMAIL
ADDRESS
EMAIL DESCRIPTION
ADDRESS
POSTAL CODE
COUNTRY
ADDITIONAL INFORMATION
GENERAL INFORMATION
PLACE OF PERFORMANCE
* = Required Field FedBizOpps Sources Sought Notice Rev. March 2010 Topeka Steam Sterilizers 66048 36C25518Q9596 07-16-2018 N 339113 Department of Veterans Affairs Network Contracting Office (NCO) 15 3450 S 4th Street Trafficway Leavenworth KS 66048 Colmery O'Neil VA Medical Center 2200 Gage Blvd.
Topeka, KS 66622
USA
James.Hogue@va.gov Contract Specialist The Colmery – O’Neil VA Medical Center is wanting to procure a sterilizer. This solicitation uses a Brand Name or Equal Description of the product required. This permits prospective contractors to offer products other than those specifically referenced by brand name (Steris).
| Item # |
| Description |
| Qty. |
| SR2222211011 |
| Amsco 400 26X37.5X48 Prevacuum Single Hinge Recessed, 1- Wall Lh Steam 208V/240V |
| 2 |
| AX0650230830000013 |
| Atlas Transfer Carriage 26X37.5X48" (Amsco 400) |
| 3 |
| AX068013081 |
| 48" Atlas Loading Car |
| 3 |
| SE601852 |
| Install - Medium Sterilizer |
| 2 |
The vendor must provide pre-vacuum, gravity, liquid, leak test and daily air removal test (dart) cycles and accommodate a variety of adjustable sterilization parameter settings and instrument sets weighing up to twenty-five (25) pounds.
1. Requirements:
a. Must be no bigger than 26x37.5x48 to fit in designed location
b. Meet Current AAMI ST 79 Guidelines for processing 25 pounds (lbs.) trays
c. FDA Validated to process up to 15 wrapped trays per cycle for a total of 375 lbs. per load
d. New style easy open door
e. 12 Programmable Cycle Selections featuring a Color Touch Screen visible from 15 feet away
f. Customizable cycle options
g. Eye level ergonomic control
h. Easy to use Sterilization Cart & Carriage for loading and unloading the steam sterilizer with removable and adjustable shelf height for maximum flexibility of chamber configuration
i. Non-lubricated, steam-activated door gasket ensures no air leaks
j. Door stops automatically if there is an obstruction
k. 15-year chamber warranty
l. Elliptical or round chamber, jacket and door constructed with 316L stainless steel
m. Automatic utilities start-up and shut-down
n. Electronic water saving device
o. Capability of Pro-Connect Remote Monitoring
p. Power slide doors controlled with touch of a button
2. Site Specific:
a. Steam, drain, electrical controls, electrical-vacuum pump, feed water, and back-flow prevention
b. Contractor shall deliver and perform certified installation providing all labor, material, and equipment to install and coordinate with the general construction contractor responsible for constructing surrounding and connecting facility/infrastructure
c. Contractor shall remove all shipping packaging, and installation debris from the site
d. Contractor shall perform start up testing and validation and provide a report to VA SPS staff
3. Trade-in Information
a. ETO
b. Sterilizers x3