FedBizOpps Amendment to a Previous Combined Solicitation

CLASSIFICATION CODE

SUBJECT

CONTRACTING OFFICE'S

ZIP-CODE

SOLICITATION NUMBER

BASE NOTICE TYPE

RESPONSE DATE (MM-DD-YYYY)

ARCHIVE

DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS

SET-ASIDE

NAICS CODE

CONTRACTING OFFICE

ADDRESS

POINT OF CONTACT

(POC Information Automatically Filled from User Profile Unless Entered)

DESCRIPTION

See Attachment

AGENCY'S URL

URL DESCRIPTION

AGENCY CONTACT'S EMAIL

ADDRESS

EMAIL DESCRIPTION

ADDRESS

POSTAL CODE

COUNTRY

ADDITIONAL INFORMATION

GENERAL INFORMATION

PLACE OF PERFORMANCE

* = Required Field FedBizOpps Amendment to a Previous Combined Solicitation Rev. March 2010 Zeiss LSM 800 microscope system, or equal.

Questions/Answers - one (1) day extension 15212-5319 36C24E19Q0079 08-08-2019 N 333314 US Department of Veterans Affairs Veterans Health Administration Regional Procurement Office (RPO) East 323 North Shore Drive, Suite 500 Pittsburgh PA 15212-5319 Contracting Officer: Michael Haydo Email: michael.haydo@va.gov VA Greater Los Angeles Healthcare System Los Angeles, CA Questions/Answers

1. Solicitation states requirement of “Monochrome CCD for widefield imaging”. If widefield imaging can be accomplish with confocal scanner, is a monochrome camera still required?

A Monochrome CCD is required.

1. Solicitation states requirement of “capability for DIC”. Does the proposal require DIC optics or merely compatibility with DIC?

A transmitted light DIC set, or equivalent, is required.

1. Solicitation states “This is a BRAND NAME OR EQUAL REQUIREMENT for a Zeiss LSM 800 system with Airyscan, Multi-photon Confocal Microscopic system”; however, “AiryScan” and “multiphoton” are not listed in the minimum requirements section. Are “AiryScan” and/or “multiphoton” a requirement of this proposal?

Both Airyscan and Multi-photon are part of the LSM 800 system, so systems that meet or exceed what they provide would be required.