FedBizOpps Sources Sought Notice
CLASSIFICATION CODE
SUBJECT
CONTRACTING OFFICE'S
ZIP-CODE
SOLICITATION NUMBER
RESPONSE DATE (MM-DD-YYYY)
ARCHIVE
DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
SET-ASIDE
NAICS CODE
CONTRACTING OFFICE
ADDRESS
POINT OF CONTACT
(POC Information Automatically Filled from User Profile Unless Entered)
DESCRIPTION
See Attachment
AGENCY'S URL
URL DESCRIPTION
AGENCY CONTACT'S EMAIL
ADDRESS
EMAIL DESCRIPTION
ADDRESS
POSTAL CODE
COUNTRY
ADDITIONAL INFORMATION
GENERAL INFORMATION
PLACE OF PERFORMANCE
* = Required Field FedBizOpps Sources Sought Notice Rev. March 2010 Sources Sought - cage and rack washer 15212-5319 36C24E18Q0077 N 334516 US Department of Veterans Affairs Veterans Health Administration Service Area Office (SAO) East North Shore Drive, Suite 500 Pittsburgh PA 15212-5319 lynn.portman@va.gov The Hines VAMC research service needs a replacement cage and rack washer for its Veterinary Medical Unit (VMU). This anticipated contract is for the equipment, freight and installation. The cage and rack washer will sanitize animal cage racks, large cages, carts, specialized rat "condos" and other animal husbandry devices that are too large to fit in the tunnel washer. Sanitization is required to meet Federal regulations and is required to maintain AAALAC accreditation. The VA has conducted market research to determine that a unit that will meet the VA’s need is equipped with an oscillating jet system to provide improved coverage over systems that utilize rotary spray arms or stationary jets. The unit is constructed of stainless steel and has all stainless-steel piping, headers, jets and drain valve. The unit is designed with seam flanges on the exterior of the cabinet in order to eliminate protrusions, bolts, nuts within the chamber, which results in an interior that does not accumulate debris and can be readily cleaned to minimize cross-contamination. The unit is equipped with self-cleaning primary debris screens and an exclusive sump bottom flush that acts to back flush to the drain any debris that settles on the tank floor. The unit also comes equipped with a water hammer eliminator as well as a steam moisture separator that prevents water slugging in the steam supply line. The unit also comes with the capability to either return the acid solution to its reservoir for re-use or dump to drain after neutralization. Circulating pump has variable frequency drive to vary water output and spray pressure. Operation is controlled through a microprocessor that controls programmable cycles of automatic Operation and comes equipped with a diagnostic program for system calibration and assurance of proper operation.
Other relevant criteria/information
· Stainless steel construction (doors, coils, piping, pumps, headers, drain valve, jets, floor grid); flangeless interior cabinet design
· Interior load clearance 48" x 85" x 92"
· Microprocessor controlled operation plus diagnostic program
· 16" x 18" safety glass view window in each door
· Circulating pump pressure gauge
· Electromechanical header drive system
· Automatic vent damper
· Automatic self-cleaning in-line screen
· Three tank reservoir system for storage re-circulated and non-recirculated treatments
· Drain cool-down system- direct cold water injection
· Emergency stop button at each end
· Pneumatic valves and controls
· Neutralization of acid solutions prior to draining
· Oscillating jet system
· Water hammer eliminator
· Friction release safety door hardware
· 2 inside safety stop cables (one on either side of chamber)
· All controls and valves on left hand side
· The washer is a two door pass-through model. The dirty side outer cabinet dimensions are: 106” tall x 96” long x 69” wide (exclusive of left-side controls and valves) with inclusion of the valves and controls the width is 97”
· The clean side is essentially a flush-mounted wall mount with door.
· Minimal downtime required
The anticipated contract will be firm fixed price. This notice is not a request for competitive proposals, however; any responsible source who believes it is capable of meeting the requirement may submit a capability statement to the contracting office no later than Wednesday 3/21/18 at 12 PM, EST. Interest/capability statements may be sent to Lynn Portman at lynn.portman@va.gov and should include company name, address, point of contact, business size, SDVOSB/VOSB status, and description on how the vendor can meet the requirements. All submissions must include a narrative description of how the proposed cage and rack washer will meet the requirements stated above AND product literature to include pictures. No telephone responses will be accepted. This notice is to assist the Government in determining sources only. A solicitation is not currently available. If a solicitation is issued, all interested parties must respond to that solicitation announcement separately from the responses to this announcement.