36C24820Q0160-006.pdf
PDF Posted
- Attached to
- Equipment: Dermatology Laser System Federal contract opportunity
- Solicitation number
- 36C24820Q0160
About this file
36C24820Q0160 Attachment B - Cost and Pricing Schedule -Derm Laser 2.pdf
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C24820Q0160-005.pdf | ||
| 36C24820Q0160-007.pdf | ||
| 36C24820Q0160-004.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment B
RFQ: 36C24820Q0160
GradeDermatology Laser to include Installation and training.
Line Item
Description
Qty
Unit
Unit Price
Total Price
Dermatology Laser Delivery System
EA
Module - Full field ablation
Arm Broadband Articulated Delivery Arm with Hand pieces
Module - BBL or IPL (Broad Band light or Intense Pulsed light) within 300-1400nm range
Module- 308 nm (UVB)- treatment psoriasis, eczema, vitiligo
Module- Fractionate ablation- Er:YAG
Square Treatment Adapter small smooth edge 15x15mm or equivalent for laser type
11mm Round Treatment Adapter or equivalent for laser type
7mm Round Treatment Adapter or equivalent for
45 x 15mm Large Smoothedge Treatment Adapter or equivalent for laser type
Air Chiller – with hose
Smoke Evacuator -Surgical Plume Evacuator
1 Day Training at Installation and 1 Day follow up training within 1 year
5 (five) Year Maintenance and Service Plan
All equipment shall meet the following specifications:
GradeDermatology Laser to include
Specifications/Features:
Line Item 0001. - Dermatology Laser Delivery System
a. Must be a multi-application/platform medical/research grade (not Spa grade) laser system.
b. Must use/support the following light/laser-based technologies:
Ability to adjust the depth and density of ablation, and coagulation independently.
a. For coagulation, vaporization, ablation or cutting of soft tissue (skin) in
Dermatology
b. Laser skin resurfacing procedures for the treatment of Acne scars; Actinic chelitis; Keloids; Verrucae; Actinic Keratoses, Seborrheic Keratoses, lichenoid keratoses; Scar revision, acne keloidalis nuchae
c. basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis lesions
d. Erbium:YAG laser ablation effectively removes lesions of both Hailey-Hailey
Disease and Darier disease and can also yield excellent long-term results in chronic, recalcitrant cases.
e. Superficial Hyperpigmentation and Vascular Lesions, Active Acne/Rosacea;
Painless Har Removal for all Skin Types
f. Tattoo Removal, Melasma, Deep Pigmented Lesions, Nail Fungus
g. treatment psoriasis, eczema, vitiligo
Required Energy/Joules Range: (Based on Treatment – see below):
Wavelengths:
Erbium 2904nm
BBL/IPL - 300-1400nm ranges QS Yag - 1064nm UVB - 308 nm
c. Must include all safety equipment for staff and patients to include (3) sets of safety Glasses/Goggles
Line Item 0002. - Module - Full field ablation – Er:YAG 2940nm
Line Item 0003. - Arm Broadband Articulated Delivery Arm with Hand pieces
a. Various adapters or handpieces to be able to adjust spot size for treatment
Line Item 0004. - Module - BBL or IPL (Broad Band light or Intense Pulsed light) within 300- 1400nm range Variety of filters in order to treat:
a. Superficial Hyperpigmentation
b. Vascular Lesions
GradeDermatology Laser to include
c. Acne/rosacea
d. Painless Har Removal
Line Item 0005. - Module- 308 nm (UVB)- treatment psoriasis, eczema, vitiligo
Line Item 0006. - Module- Fractionate ablation- Er:YAG
a. Variety of depth and density coverage to allow treatment various conditions and skin types
• For example shallow (up to 4 microns of ablation) to deep resurfacing (up to 200 microns of ablations per pass, with or without variable depth coagulation)
Line Item 0007. - Square Treatment Adapter small smooth edge 15x15mm or equivalent for
Line Item 0008. - 11mm Round Treatment Adapter or equivalent for laser type
Line Item 0009. - 7mm Round Treatment Adapter or equivalent for laser type
Line Item 0010. - 45 x 15mm Large Smoothedge Treatment Adapter or equivalent for laser type
Line Item 0011. - Air Chiller – with hose
Line Item 0012. - Smoke Evacuator -Surgical Plume Evacuator
Line Item 0013. - 1 Day Training at Installation and 1 Day follow up training within 1 year
Line Item 0014. - 5 (five) Year Maintenance and Service Plan
Output Range:
Laser Output Class IV, 0.7-3 um, 100W, 75J, 200 ms Light Output 300-1400 nm, 0-300J
Pulse Range:
Pulse Duration of individual pulse 0.5 to 1.5 ms Pulse Interval 0.2 to 200 ms Broadband pulsed light with 300-1400 nm Fluence and the duration of each pulse or pulse sequence can be varied to suit the application needed.
File details come from the government source that posted it. Updated .