Bay Pines VA Health Care System 36C24819Q1211 Salient Characteristics
SALIENT CHARACTERISTICS
Vascular Imaging Ultrasound System *Brand Name or Equal*
Equal must meet these salient characteristics. Explanation of how the equal meets each salient characteristic must be included with quote.
Preferred Brand: ZONARE
ZS3 7.x General Imaging Ultrasound System
· Must be able to utilize all the information contained in the returning echo data set and as such cover the field of view in much fewer transmit / receive cycles.
· Must have 2D/ B-mode imaging, color flow, angio/velocity calculations
· Must be compatible with ZONARE transducers we currently use
· Must be portable, light weight, easily maneuverable to fit in tight Operating Room suites and have a battery life of 4 hours or greater and a plug-in mode.
· Must have at least 19” high resolution HD viewing screen for multiple physicians, to avoid sterile field contamination.
· Must have multi-transducer port and at least 2 USB ports
· Must have a touchscreen for high visibility and wide viewing angle
· Must have high definition imaging, that is applicable to multiple ultrasound applications, for example: Abd sonography, Vascular sonography and intra-operative studies.
C-4 Curved Array Transducer
· Must be light weight
· Must be compatible ZS3 General Imaging Ultrasound System (or similar system)
· Must have good penetration for indicated frequency
· Must have the ability to image in between costal spaces.
· Must have adjustable frequencies to image patients of varying sizes and medical conditions.
L8-3 Linear Array Transducer
· Must utilize high frequency imaging of at least 8mgHz and above
· Must use linear array technology to provide accurate vascular exams.
· Must be light weight
· Must be compatible ZS3 General Imaging Ultrasound System (or similar system)
· Must have good penetration for indicated frequency
· Must have the ability to image in between costal spaces.
· Must have adjustable frequencies to image patients of varying sizes and medical conditions.
L14-5sp (Special Procedures) Linear Array Transducer
· Must be HD-high frequency of at least 14mgHz and above for intra-operative use
· Must be light weight
· Must be compatible ZS3 General Imaging Ultrasound System (or similar system)
· Must have good penetration for indicated frequency
· Must have the ability to image in between costal spaces.
· Must have adjustable frequencies to image patients of varying sizes and medical conditions.