Attachment B 36C24818Q0702
SALIENT CHARACTERISTICS
1. Oxygenator Pump & Table Pack/Kit
· Carmeda coated
· Integrated arterial filter
· 25 µm filtration
· 260 mL prime
· Enhanced gas transfer
· Plastic heat exchanger (PET)
· Recirculation and cardioplegia ports
· ½" quick-connect water ports
· Ambidextrous design
· BPA-free plastic
· Clear housing
· Volume-displacing leur caps
· Built-in temp probe sleeve
· 1-7 L/min flow
2. Specifications
| Membrane type: |
| Microporous polypropylene hollow fiber |
| Membrane surface area: |
| 2.5m2 |
| Heat exchange surface material: |
| Polyethylene terephthalate (PET) |
| Static priming volume: |
| 260 mL |
| Recommended blood flow rate: |
| 1-7 L/min |
| Maximum water side pressure: |
| 30 psi |
| Maximum blood pressure: |
| 750 mm Hg |
| Arterial outlet port: |
| 3/8" |
| Venous inlet port: |
| 3/8" |
| Arterial sample port: |
| Female luer port |
| Recirculation port: |
| 1/4" |
| Cardioplegia port: |
| 1/4" |
| Gas inlet port: |
| 1/4" nonbarbed |
| Gas outlet port: |
| 3/8" nonbarbed |
| Water ports: |
| 1/2" quick disconnects |
| Filtration: |
| 25 µm |