Text version

1© 2007 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1.800.643.6275 | 1.937.526.3662 | midmark.com

Step 1: Cut cable ties.

Remove shipping brace.

Procedure Chair Installation Applies to Models:

230 (all) 630 (-001 thru -009) 641 (all) Language or origin: English

(630 shown)

Equipment Alert Voltage setting that shows thru window must match supply voltage. Incorrect setting will result in damage to the chair.

Voltage Conversion Procedure:

A. Open fuse holder cover.

B. Remove fuse holder .

C. Flip fuse holder 180°.

(630 shown)

003-1406-00 Rev. LA2 (8/10/22)Style A

Models w/Rotational Base only Step 2: Remove shipping bracket.

Step 3: Remove shipping bolts.

Remove chair from skid.

NOTE: Shipping brace may not be present on some models.

Does the voltage setting that appears in the fuse holder window match the facility supply voltage?

If YES, go to Step 4.

If NO, you must perform Voltage Conversion Procedure.

2© 2007 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1.800.643.6275 | 1.937.526.3662 | midmark.com

MA642802i

Step 4: Connect hand/foot control to cord inlet.

Tighten locking ring.

Note: The cord connectors are “keyed” and must be oriented properly to connect.

Rear Cord Inlet

(630 shown)

Front Cord Inlet

003-1406-00

MA720301i

Before connecting the Midmark device power cord into an electrical power source outlet, it is recommended to contact a local licensed electrician or a local Authority Having Jurisdiction over the internal building power distribution to verify that the device when powered up will comply with all local electrical codes. Use the electrical power rating marked on the device when determining the appropriate electrical branch circuit and outlet requirements.

For 115V rated device models with a current rating at or above 12 Amps, a dedicated branch circuit may be required if the branch circuit protection is provided by a 15 Amp breaker.

Step 5: Connect power cord.

Step 6: Adjust four leveling feet until chair is level.

Step 7: Install hand control mounting discs (2), and holster.

Midmark models only Step 8: Install magnetic base covers (3 pcs.)

Equipment Alert The chair must be level to ensure proper operation.

Attention Slide narrow portion of cord connector into bracket as shown.

Locking Ring