This document is a Sources Sought Notice from the Department of Veterans Affairs Network Contracting Office regarding a water purification system for the VA Chesapeake Community Based Outpatient Clinic. The notice seeks a system capable of producing at least 200 liters of water per hour with deionization capabilities, similar to the Medica-R 200 model. Interested parties should respond by February 13, 2024 at 10:00 AM Eastern Time. The system will be located at the VA Medical Center in Hampton, Virginia and must meet specifications of 200 liters/hour maximum production, 21 liters/minute maximum flow rate at 45 PSI, and resistivity and total organic carbon measurements provided.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VA
Chesapeake
Community
Based
Outpatient
Clinic
(CBOC)
Chesapeake
HCC1
| SUBJECT* |
| WATER PURFICATION SYSTEM HCC1 LAB |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 23667 |
| SOLICITATION NUMBER* |
| 36C24624Q0414 |
| RESPONSE DATE/TIME/ZONE |
| 02-13-2024 10am EASTERN TIME, NEW YORK, USA |
| ARCHIVE |
| 10 DAYS AFTER THE RESPONSE DATE |
| RECOVERY ACT FUNDS |
| N |
SET-ASIDE
| PRODUCT SERVICE CODE* |
| 6515 |
| NAICS CODE* |
| 339113 |
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667
POINT OF CONTACT*
PLACE OF PERFORMANCE
| ADDRESS |
| VAMC Hampton |
100 Emancipation Dr
Hampton VA
| POSTAL CODE |
| 23667 |
COUNTRY
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
AGENCY CONTACT’S EMAIL ADDRESS
EMAIL DESCRIPTION
Sources Sought Notice Sources Sought Notice
DESCRIPTION
| *= Required Field |
| Sources Sought Notice |
Sources Sought Notice
Salient Characteristics
Water Purification System that is capable of producing not less than 200 liters of water per hour, fitted with deionization capabilities
MODEL MEDICA-R 200
· Make-up rate @15°C 200 l/hr
· Delivery flow rate – maximum115V, 60Hz 21 l/min @ 45psi (3bar) Inorganics (resistivity @25°C) >10 MΩ-cm (1)
· Organics (TOC) - typical <30ppb -500 PPB (1) Bacteria - typical <10 CFU/ml (2)
· Particles 0.2 μm filter Silica <0.05 mg/l
· Notes: 1 with Ion-exchange cylinder installed (Nuclear or Hypex grade resin), 2 subject to correct operating and maintenance procedures
· Feedwater Requirement:
· Source: Tap water as detailed below Conductivity: <1000 μS/cm
· Contaminant:
| · Measure: | Range: | Pretreatment: |
| · Calcium: Ca ppm as CaCO3 | <250 | None3 |
| · >250 | Softener or use very low RO recovery3 | |
| · Free chlorine: Cl2ppm | <0.1 | None |
| · 0.1 – 0.5 | 20 inch carbon block | |
| · >0.5 | Cylinder of carbon sized correctly to obtain <0.1 ppm | |
| · Chloramine: Cl2 ppm | <0.1 | None |
| · 0.1 – 0.5 | 20 inch high efficiency carbon (5) | |
| · >0.5 | Cylinder of carbon sized correctly to obtain <0.1 ppm | |
| · Silica: SiO2 ppm | <30 | None |
| · >30 | 20 inch cartridge depth filter or use very low RO recovery (4) | |
| · Fouling index: FI | <10 | None |
| · 10 to 20 | 20 inch cartridge depth filter | |
| · >20 | Back washable media filter with a minimum flow rate of 20 l/min | |
| · Iron/manganese: Fe/Mn ppm | <0.05 | None |
| · >0.05 | 20 inch cartridge depth filter (5) | |
| · >0.1 | Back-washable Fe filter (6) | |
| · Organics: TOC ppm C | <2 | None |
| · 2 to 3 | 20 inch carbon block (4) | |
| - | >3 | Cylinder of carbon sized correctly for TOC demand (4) Temperature: 4 - 40°C (Recommended 15 - 25°C) |
· Flowrate: (requirement at 15°C) 20 l/min
· Drain requirements: (gravity fall with air gap) 45 l/min
· Feedwater pressure: 60psi (4bar) maximum, 30psi (2bar) minimum
· Notes: 3) Check LSI, increase frequency of acid cleaning, 4) Increase frequency of alkaline cleaning,
· 5) Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter), 6) Increase frequency of acid cleaning Wall mount canister filter will need (Wall backing required)
· Must include: Installation, Commissioning, Training
· Electrical Requirements Mains Input: 115V, 60Hz
· System control voltage (not including pumps and UV) 24V dc Power consumption (peak demand): 2000VA
· Electrical protection rating 20 amps Noise level during recirculation <70dBA
· ***Note: System requires approximately 4' x 6' of floor space for tanks, filtration Unit, and filters.
· -The floor drain must be a minimum of 3" and meet all applicable building codes and system requirements
· -Actual Unit measurement 32" Wide x35"Deep x72"High