36C24624Q0414.docx

DOCX document 66 KB Posted

Attached to
6515-- WATER PURFICATION SYSTEM HCC1 LAB Federal contract opportunity
Solicitation number
36C24624Q0414
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 6

About this file

This document is a Sources Sought Notice from the Department of Veterans Affairs Network Contracting Office regarding a water purification system for the VA Chesapeake Community Based Outpatient Clinic. The notice seeks a system capable of producing at least 200 liters of water per hour with deionization capabilities, similar to the Medica-R 200 model. Interested parties should respond by February 13, 2024 at 10:00 AM Eastern Time. The system will be located at the VA Medical Center in Hampton, Virginia and must meet specifications of 200 liters/hour maximum production, 21 liters/minute maximum flow rate at 45 PSI, and resistivity and total organic carbon measurements provided.

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

VA

Chesapeake

Community

Based

Outpatient

Clinic

(CBOC)

Chesapeake

HCC1

SUBJECT*
WATER PURFICATION SYSTEM HCC1 LAB

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
23667
SOLICITATION NUMBER*
36C24624Q0414
RESPONSE DATE/TIME/ZONE
02-13-2024 10am EASTERN TIME, NEW YORK, USA
ARCHIVE
10 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
6515
NAICS CODE*
339113

CONTRACTING OFFICE ADDRESS

Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667

POINT OF CONTACT*

PLACE OF PERFORMANCE

ADDRESS
VAMC Hampton

100 Emancipation Dr

Hampton VA

POSTAL CODE
23667

COUNTRY

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

Sources Sought Notice Sources Sought Notice

DESCRIPTION

*= Required Field
Sources Sought Notice

Sources Sought Notice

Salient Characteristics

Water Purification System that is capable of producing not less than 200 liters of water per hour, fitted with deionization capabilities

MODEL MEDICA-R 200

· Make-up rate @15°C 200 l/hr

· Delivery flow rate – maximum115V, 60Hz 21 l/min @ 45psi (3bar) Inorganics (resistivity @25°C) >10 MΩ-cm (1)

· Organics (TOC) - typical <30ppb -500 PPB (1) Bacteria - typical <10 CFU/ml (2)

· Particles 0.2 μm filter Silica <0.05 mg/l

· Notes: 1 with Ion-exchange cylinder installed (Nuclear or Hypex grade resin), 2 subject to correct operating and maintenance procedures

· Feedwater Requirement:

· Source: Tap water as detailed below Conductivity: <1000 μS/cm

· Contaminant:

· Measure:Range:Pretreatment:
· Calcium: Ca ppm as CaCO3<250None3
· >250Softener or use very low RO recovery3
· Free chlorine: Cl2ppm<0.1None
· 0.1 – 0.520 inch carbon block
· >0.5Cylinder of carbon sized correctly to obtain <0.1 ppm
· Chloramine: Cl2 ppm<0.1None
· 0.1 – 0.520 inch high efficiency carbon (5)
· >0.5Cylinder of carbon sized correctly to obtain <0.1 ppm
· Silica: SiO2 ppm<30None
· >3020 inch cartridge depth filter or use very low RO recovery (4)
· Fouling index: FI<10None
· 10 to 2020 inch cartridge depth filter
· >20Back washable media filter with a minimum flow rate of 20 l/min
· Iron/manganese: Fe/Mn ppm<0.05None
· >0.0520 inch cartridge depth filter (5)
· >0.1Back-washable Fe filter (6)
· Organics: TOC ppm C<2None
· 2 to 320 inch carbon block (4)
->3Cylinder of carbon sized correctly for TOC demand (4) Temperature: 4 - 40°C (Recommended 15 - 25°C)

· Flowrate: (requirement at 15°C) 20 l/min

· Drain requirements: (gravity fall with air gap) 45 l/min

· Feedwater pressure: 60psi (4bar) maximum, 30psi (2bar) minimum

· Notes: 3) Check LSI, increase frequency of acid cleaning, 4) Increase frequency of alkaline cleaning,

· 5) Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter), 6) Increase frequency of acid cleaning Wall mount canister filter will need (Wall backing required)

· Must include: Installation, Commissioning, Training

· Electrical Requirements Mains Input: 115V, 60Hz

· System control voltage (not including pumps and UV) 24V dc Power consumption (peak demand): 2000VA

· Electrical protection rating 20 amps Noise level during recirculation <70dBA

· ***Note: System requires approximately 4' x 6' of floor space for tanks, filtration Unit, and filters.

· -The floor drain must be a minimum of 3" and meet all applicable building codes and system requirements

· -Actual Unit measurement 32" Wide x35"Deep x72"High

VA

Chesapeake

Community

Based

Outpatient

Clinic

(CBOC)

Chesapeake

HCC1

Qty:

Finishes/Options

Image

PATHOLOGY

AND

LABORATORY

LAB-

C1012(1) Department, Room, Quantity Page of image1.jpeg

File details come from the government source that posted it. Updated .