36C24624Q0414_2.docx
DOCX document 66 KB Posted
- Attached to
- 4610-- WATER PURFICATION SYSTEM HCC1 LAB Federal contract opportunity
- Solicitation number
- 36C24624Q0414
About this file
This document provides details for a pre-solicitation federal contract opportunity for a water purification system. The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 6 is seeking a water purification system capable of producing not less than 200 liters of water per hour with deionization capabilities and a maximum delivery flow rate of 21 liters per minute at 45 pounds per square inch. The system should be a MEDICA-R 200 model with a makeup rate of 200 liters per hour and have inorganic resistivity greater than 10 megohm-centimeters and typical organic total organic carbon levels. The opportunity number is 36C24624Q0414 for the Chesapeake Community Based Outpatient Clinic in Chesapeake, Virginia.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VA
Chesapeake
Community
Based
Outpatient
Clinic
(CBOC)
Chesapeake
HCC1
| SUBJECT* |
| WATER PURFICATION SYSTEM HCC1 LAB |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 23667 |
| SOLICITATION NUMBER* |
| 36C24624Q0414 |
| RESPONSE DATE/TIME/ZONE |
| 03-06-2024 10am EASTERN TIME, NEW YORK, USA |
| ARCHIVE |
| 5 DAYS AFTER THE RESPONSE DATE |
SET-ASIDE
| PRODUCT SERVICE CODE* |
| 4610 |
| NAICS CODE* |
| 333310 |
PLACE OF PERFORMANCE
POSTAL CODE
COUNTRY
CONTACT INFORMATION
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs Network Contracting Office 6 100 Emancipation Drive Hampton VA 23667
POINT OF CONTACT*
Contract Specialist John P Harmsen john.harmsen@va.gov 757-722-7961
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
AGENCY CONTACT’S EMAIL ADDRESS
EMAIL DESCRIPTION
Presolicitation Notice Presolicitation Notice
DESCRIPTION
| *=Required Field |
| Presolicitation Notice |
Presolicitation Notice
Salient Characteristics
Water Purification System that is capable of producing not less than 200 liters of water per hour, fitted with deionization capabilities
MODEL MEDICA-R 200
· Make-up rate @15°C 200 l/hr
· Delivery flow rate – maximum115V, 60Hz 21 l/min @ 45psi (3bar) Inorganics (resistivity @25°C) >10 MΩ-cm (1)
· Organics (TOC) - typical <30ppb -500 PPB (1) Bacteria - typical <10 CFU/ml (2)
· Particles 0.2 μm filter Silica <0.05 mg/l
· Notes: 1 with Ion-exchange cylinder installed (Nuclear or Hypex grade resin), 2 subject to correct operating and maintenance procedures
· Feedwater Requirement:
· Source: Tap water as detailed below Conductivity: <1000 μS/cm
· Contaminant:
| · Measure: | Range: | Pretreatment: |
| · Calcium: Ca ppm as CaCO3 | <250 | None3 |
| · >250 | Softener or use very low RO recovery3 | |
| · Free chlorine: Cl2ppm | <0.1 | None |
| · 0.1 – 0.5 | 20 inch carbon block | |
| · >0.5 | Cylinder of carbon sized correctly to obtain <0.1 ppm | |
| · Chloramine: Cl2 ppm | <0.1 | None |
| · 0.1 – 0.5 | 20 inch high efficiency carbon (5) | |
| · >0.5 | Cylinder of carbon sized correctly to obtain <0.1 ppm | |
| · Silica: SiO2 ppm | <30 | None |
| · >30 | 20 inch cartridge depth filter or use very low RO recovery (4) | |
| · Fouling index: FI | <10 | None |
| · 10 to 20 | 20 inch cartridge depth filter | |
| · >20 | Back washable media filter with a minimum flow rate of 20 l/min | |
| · Iron/manganese: Fe/Mn ppm | <0.05 | None |
| · >0.05 | 20 inch cartridge depth filter (5) | |
| · >0.1 | Back-washable Fe filter (6) | |
| · Organics: TOC ppm C | <2 | None |
| · 2 to 3 | 20 inch carbon block (4) | |
| - | >3 | Cylinder of carbon sized correctly for TOC demand (4) Temperature: 4 - 40°C (Recommended 15 - 25°C) |
· Flowrate: (requirement at 15°C) 20 l/min
· Drain requirements: (gravity fall with air gap) 45 l/min
· Feedwater pressure: 60psi (4bar) maximum, 30psi (2bar) minimum
· Notes: 3) Check LSI, increase frequency of acid cleaning, 4) Increase frequency of alkaline cleaning,
· 5) Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter), 6) Increase frequency of acid cleaning Wall mount canister filter will need (Wall backing required)
· Must include: Installation, Commissioning, Training
· Electrical Requirements Mains Input: 115V, 60Hz
· System control voltage (not including pumps and UV) 24V dc Power consumption (peak demand): 2000VA
· Electrical protection rating 20 amps Noise level during recirculation <70dBA
· ***Note: System requires approximately 4' x 6' of floor space for tanks, filtration Unit, and filters.
· -The floor drain must be a minimum of 3" and meet all applicable building codes and system requirements
· -Actual Unit measurement 32" Wide x35"Deep x72"High