Minimally, the proposed system must be able to do the following:
· Electronically record temperatures by accurately monitoring two (2) low and two (2) high preprogrammed threshold temperatures (between -30°C and +50°C), and will indicate time and duration of a temperature breach (up to 6 days).
· 8 Parameter Settings total: (2) Low Temp, (2) High Temp, (1) Excursion time up to 6 days (measured in days, hours, minutes), (1) Excursion time low up to 6 days (same measurement), (1) delay start-up 30 minutes, (1) Sampling interval 1 minutes.
· Accurate within ±0.5°C between -20°C and +30°C
· Must be programmed at Manufacture to SNS provided specifications in Celsius
| · High Temp 1 |
| · >8°C |
| · High Temp 2 |
| · >25°C |
| · Low Temp 1 |
| · N/A |
| · Low Temp 2 |
| · N/A |
| · Excursion Time High 6 days, 19 hours, 21 minutes |
| · >48 Hours |
| · Excursion Time Low 6 days, 19 hours, 21 minutes |
| · N/A |
| · Delay Start Time (0-99 minutes) |
| · 30mins |
| · Sampling Interval (1-99 minutes) |
| · 1mins |
· Data recorded on temperature monitor devise must be tamper proof and permanent
· Weight ≤ 9 ounces
· Dimensions: ≤ 1.75" x 2.5" x 0.5" (W x H x Thickness)
· Resolution: ±0.1°C
· Battery Shelf Life ≥ 2 years (from time of programming)
· Expire date indicated on unit (defined by end of Month and Year)
· Must visually display an out of temperature notification (or an icon symbol) if the temperature exceeded the programmed parameters such as an “Ok” and/or an “Alarm”
· Must display out of time excursion defined in days, hours, minutes for both high and low settings
· Simple one or two button end user operation of monitor
· Must provide easy to follow, short and simple direction cards or sheets for end user to operate monitor and interpret results.
· Must provide end user operation instructions card or sheet for each unit purchased
· Must provide web site and phone contact information for end user for help assistance.
· With short DSNS notice, the Manufacture must provide knowledgeable operator help assistance 24/7 for the period of 2 weeks minimum to support end user questions. Minimum of 12 operator lines.
· Must include statement on end user directions for safe disposal and any associated material risks by end user
· No Chemical indicator devices accepted
· Single use, disposable
· Must describe their QA/QC Validation Approach per industry standard and traceable to NIST standards
· Must describe their General Qualification Acceptance Criteria
· Must describe their testing philosophy approach and methodology (Software and Hardware, Factory Acceptance Testing)
· Must provide Qualification verification letter for all units sold
· Units must be serialized or Lot # controlled (defined on Quality Acceptance Verification letter)
· Must provide Five reference sites where the monitors has been successfully implemented for a minimum of two years
· Must indicate the varying service level agreements and associated fees
· Must define how future upgrades affect other listed requirements