STATEMENT OF WORK
ONE COMPLETE SPOT BENCHTOP ENERGY DISPERSIVE X-RAY FLUORESCENCE (EDXRF) SPECTROMETER SYSTEM (MOBILE) WITH INTEGRATED CAMERA AND VARIABLE SMALL SPOT ANALYSIS FOR THE U.S. CUSTOMS AND BORDER PROTECTION, LABORATORY AND SCIENTIFIC SERVICES (LSSD) IN SAN FRANCISCO.
I.
TABLE OF CONTENTS
I.
TABLE OF CONTENTS
II.
STATEMENT OF INTENTION
III.
STATEMENT OF EXPECTATIONS
IV.
SPECIFIC STATEMENT OF INSTRUMENTATION AND SPECIFICATIONS
V.
SERVICEABILITY AND SYSTEM STATUS
II.
STATEMENT OF INTENTION
It is the intention of the CBP Laboratory to purchase a small benchtop energy dispersive x-ray fluorescence (EDXRF) spectrometer system for the CBP, LSSD, Laboratory in San Francisco. The instrument is being purchased for field deployment to replace our more than 15 year old portable xrf.
III.
STATEMENT OF EXPECTATIONS
It is the expectation of the CBP, LSSD, that the purchased system will produce quality analyses for inorganic elements from sodium to uranium. The system will be able to handle different matrices such as solids, thin films, metals, liquids, slurries, pastes and powders, etc.
It is our expectation that we will be able to easily select our operating parameters. We expect that the system will be easy to operate and easy to maintain. Switching from one configuration to another will be simple and straightforward, requiring no significant modifications to the system.
It is our expectation that the system that we purchase will provide the widest analytical application with the smallest samples at reasonable accuracy and precision and still meet our portable requirements.
Service and parts will be readily available. Repair and maintenance by a qualified service engineer will be available within 5 business days of a service call, and all required parts will be available within 5 business days.
On-site training (5 days) covering the operation, application and routine maintenance of the system components will be provided at the time of the installation of the complete system.
IV.
GENERAL STATEMENT OF INSTRUMENTATION
The complete instrument system, all sub-systems, and all components, parts, and attachments shall be new. Proposals incorporating reconditioned or used instruments in part or in full are not acceptable and will be regarded as non-responsive. The installed instrument system shall be new, commercially available, and in commercial production as of the closing date of this solicitation. The completely installed instrument system in the U.S. Customs Service Laboratory in San Francisco shall meet all of the main and sub-system manufacturers published overall and sub-system performance specifications as installed in our laboratory on our site.
The instrument system containing all listed sub-systems and major components shall be provided as part of the complete instrument system and each shall meet their individually described detailed technical and performance specifications. The instrument system and sub-systems will be purchased from a single vendor. Thus, any vendor not responding to all minimum requirements will be considered non-responsive. No prototypes, one-of-a-kind, or two-of-a-kind systems will be considered.
The instrument will be used to analyze for inorganic elements with masses between sodium and uranium. These elements may be contained in different matrices such as petroleum crude oil, minerals, steel discs, powders, thin films, etc. Each major specific sub-system is described below, followed by detailed technical and performance specifications. The instrument will feature state-of-the-art analog and digitally controlled components. These systems shall be completely integrated.
The provided instrument shall consist of the following listed major sub-systems and their listed major components:
A.
The benchtop energy dispersive x-ray fluorescence spectrometer system must be compact and be able to analyze for the elemental range from sodium (Na) to uranium (U) and in the concentration range from PPM (part per million) to 100%.
B.
The system must come with integrated high resolution camera and variable small spot analysis (1, 3, and 10 mm Spot Size Automatic Collimators).
C.
The system must be one unit with no operating attachments other than the associated laptop computer. The unit must be able to be moved and set up by one person of medium height, weight and strength.
D.
The system must come with a hard plastic travel case with integrated rollaway wheels and retractable handles. This case must allow the unit to be portable. It has to be easily moved by one person and set up for operation by one person. The interior must have customized interior for the protection of the system and parts (not including the computer or printer). The container will allow shipping via Federal Express priority service fully loaded.
E.
The unit must meet all safety regulations when operating and when inoperative (shut down). It must be able to be safely shipped by air or packed into a mini-van.
F.
Electrical Specifications
1.
Single phase AC 100/220V
2.
1.5A, 50/60 Hz
3.
Must meet adequate safety protections
G.
Detector
High performance, fast silicon drift detector
Peltier electronic cooling
Automated or user configurable shaping times for optimum analytical performance
4.
Digital signal processor
5.
Selectable Shaping Times H.
X-ray tube
X-ray tube with Ag anode (include X-ray enable key switch)
60 kV max voltage, 1 mA max current
Max. tube power 12W
Automated 7 position primary filter wheel I.
Sample Handling
1.
Must be able to handle powders (loose and pressed), fused beads, metals, liquids, thin films, slurries, pastes 2.
Sample cups must be provided with plastic film inserts that allow for 2nd tier leakage protection when measuring liquid samples (ex. S in oil)
3.
32 mm cup sample ring, 10 and 27 mm flat sample apertures 4.
32 mm single position sample spinner
5.
Sample chamber cannot exceed the dimensions of 105 X305 X 305 mm (HxWxD)
J.
The system must come with helium purge option.
K.
The system must come with a Laptop with Microsoft Windows 10 or higher operating system.
L.
Software
Software must be include the following attributes:
Software for qualitative and quantitative analysis.
Menu-based software for control of spectrometer functions and data analysis
Simple flow bar wizard to create new methods and profiles
Empirical calibration with overlap and matrix compensation
5.
Data download via USB V.
SERVICEABILITY AND SYSTEM STATUS
The system will pass the manufacturers specifications in the available work space of the laboratory receiving the instrument system, the CBP, in San Francisco (15th Floor) without additional customer cost or modifications to the facility. Power in the San Francisco laboratory is 120/208 volts, 60 Hz.
A.
The system will be commercially available for sale and in production as of the closing date of the solicitation.
B.
The supplier shall install the system and provide five (5) day on-site training.
C.
Technical representatives (English speaking) shall be available by telephone who are experts with the system operation and technical applications.
D.
Service and parts will be readily available. Repair and maintenance by a qualified service engineer (US citizen and English speaking) will be available within 5 business days of a service call, and all required parts will be available within 5 business days.
E.
At minimum, two (2) year warranty will be provided covering all parts and labor on all of the hardware and software purchased. The warranty shall include all parts, labor, travel/perdiem and shipping costs.The warranty shall start after system installation and complete acceptance of the system by the U.S. Customs Service.
F.
Point of delivery:
Inside delivery required to 14th floor:
U.S. Customs and Border Protection, Laboratory
630 Sansome St., Room 1450 San Francisco, CA 94111
G.
Scientists from the Customs and Border Protection Laboratory will evaluate the instrumentation over the first 30 days of the completion of installation and training, to determine that it is capable of meeting all of the requirements outlined in the Statement of Work. If it is not, the vendor will fix the deficiencies within the next 30 days, or the instrumentation will be returned for a full refund.
Invoices can be sent through IPP System after 30 days from the comprehensive training.
The government reserves the right to award a contract to the contractor who submits a proposal that provides the best overall value in terms of each of the evaluation categories identified above.