Solicitation Questions v.1.docx
DOCX document 13 KB Posted
- Attached to
- Scanning Electron Microscope System (SEM) Federal contract opportunity
- Solicitation number
- 1605C1-21-Q-00017
- Issued by
- Not on record
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Solicitation Questions.docx | DOCX document | |
| Sol 1605C1-21-Q-00017 SEM.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Solicitation SEM 1605C1-21-Q-00017 – Questions
1. Reference “2.1.13. The SEM-EDS system will have zero-magnification photographic sample mapping.” Is this for sample navigation, auto montage or something else?
The zero-magnification feature is required for sample mapping and navigation. Our analytical process includes the investigation of large samples, as well as multiple samples on the stage. Auto-montage is applicable mostly to materials failure investigations, where the analyst may require a complete image of the sample for their final report.
2. On the first page the Offer Due Date is given as 6/9. In the Schedule on page 2 the Close of Solicitation is given as 6/11. Could you please confirm the close date for when a reply to this from Thermo Fisher Scientific is due?
June 9, 2021 in block 8 is a typo. The solicitation deadline is June 11, 2021 in accordance with all the other references.
File details come from the government source that posted it. Updated .