This document contains the technical specifications for an Elemental Scientific HydrideFAST autosampler and related equipment required for a U.S. Geological Survey (USGS) contract opportunity. The key requirements include a hydride gas liquid separator sample introduction system compatible with a Perkin Elmer Nexion 2000 ICP-MS, a compact footprint, a robust syringe-driven system, automatic calibration curve generation, the ability to produce detection limits less than 0.0006 ppb for selenium, and the capability to switch between high sensitivity hydride generating elements and non-hydride forming elements using a nebulizer. The USGS is soliciting one system with delivery required no later than June 1, 2024 to their facility in Moffett Field, California. This is a small business set-aside opportunity under NAICS code 333516 with a firm, fixed-price quote requested. Responses are due to the Contracting Officer by March 25, 2024.

View the file

Other files for this federal contract opportunity

Other files attached to AUTOSAMPLER AND HYDRIDEFAST, newest first.
File Type Posted
Sol_140G0324Q0097_Amd_0001.pdf PDF
Combined_Synopsis_Solicitation_Attachments.pdf PDF
Sol_140G0324Q0097.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Required Specifications for Elemental Scientific Brand Name or Equal

One (1) 4DXCi and one (1) HFX-64:

Required Specifications:

1. Hydride gas liquid separator sample introduction system compatible with Perkin Elmer

Nexion 2000 ICP-MS (only known device is Elemental Scientific Instrument S31

HydrideFAST system).

2. Compact small footprint.

3. Robust syringe driven system.

4. Automatic calibration curve generation.

5. Capable of producing detection limit < 0.0006 ppb for selenium.

6. Autosampler capable of switching between high sensitivity hydride generating elements and non-hydride forming elements using a nebulizer.

7. Technical support readily available.

8. Four additional years of extended service agreements.

Quantity:

One system required.

Need Delivered No Later Than:

June 1, 2024

Delivery Address:

US Geological Survey

Building 800

450 Severyns Avenue

Moffett Field, CA 94035