Salient Characteristics S19172

DEPTH AND LEVEL PRESSURE TRANSMITTER

ALL PART ARE BRAND NAME OR EQUAL

SALIENT CHARACTERISTICS:

21 EA, Depth and Level Pressure Transmitter:

1. Shall have diameter no larger than 0.39”.

2. Shall have silicon, piezoresistive sensor.

3. Shall have accuracy at ±0.1% FS.

4. Shall have overpressure at 3 x range.

5. Shall have operating temperature range at 25° to 125°F (-4° to +50°C).

6. Shall have compensated temperature range at 30° to 85°F (-2° to +30°C).

7. Shall have compensated temperature effects at ±1.5% total for 25° to 125° F, over 50 psi FS and ±2% total for 25° to 125° F, up to 50 psi FS.

8. Shall have long term stability at 0.1% FS per year for pressure ranges greater than 30 psi FS.

9. 4-20 mA (2-wire) electrical configuration with in-line converter and operate from 9-33 VDC supply.

10. Shall have Polyurethane molded, Kevlar reinforced, vented, 6 conductors, cable.

11. Shall have fully potted and molded high integrity cable sea.

12. Shall include anti-shock protective nose cone at 0.39”.

13. Shall have fully welded Titanium housing.

14. Shall weigh no more than 1 oz. excluding cable.

15. Shall have four (4) with pressure range from 0 to 20 ft. water column with cable length from sensor to convertor at 20 ft and cable length from converter to termination at 3 ft. Shall have four (4) CABLES.

16. Shall have five (5) with pressure range from 0 to 40 ft. water column with cable length from sensor to convertor at 46 ft and cable length from converter to termination at 3 ft. Shall have five (5) CABLES.

17. Shall have two (2) with pressure range from 0 to 25 ft. water column with cable length from sensor to convertor at 25 ft and cable length from converter to termination at 3 ft. Shall have two (2) CABLES.

18. Shall have three (3) with pressure range from 0 to 40 ft. water column with cable length from sensor to convertor at 40 ft and cable length from converter to termination at 3 ft. Shall have three (3) CABLES.

19. Shall have two (2) with pressure range from 0 to 20 ft. water column with cable length from sensor to convertor at 25 ft and cable length from converter to termination at 3 ft. Shall have two (2) CABLES.

20. Shall have three (3) with pressure range from 0 to 30 ft. water column with cable length from sensor to convertor at 30 ft and cable length from converter to termination at 3 ft. Shall have three (3) CABLES.

21. Shall have two (2) with pressure range from 0 to 30 ft. water column with cable length from sensor to convertor at 36 ft and cable length from converter to termination at 3 ft. Shall have two (2) CABLES.

42 EA, Moisture Protection Vent Filter:

22. Shall measure 6” x 3/4” diameter

23. Shall contain a desiccant that changes color when maintenance is required.

25 EA, Galvanized Steel Cable Hanger:

24. Shall allow for installation after the cable termination has been made

25. Shall provide support for transmitters hanging in wells or tanks.