Questions and Answers_Final .docx
DOCX document 13 KB Posted
- Attached to
- Combustion Ion Chromatography System Federal contract opportunity
- Solicitation number
- 1333ND23QNB730551
About this file
This document contains questions and answers related to solicitation number 1333ND23QNB730551 from the Department of Commerce National Institute of Standards and Technology for a combustion ion chromatography system. The questions and answers clarify that solid and liquid samples would be analyzed separately and stored either in autosampler racks fitted onto the autosampler or in racks ready to load onto the bench. Additionally, the document specifies that instrument method settings for flow and temperature of combustion conditions would be adjusted per each method rather than on the fly.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Applicable Provisions and Clauses Final .pdf | ||
| Requirements Document Final.pdf | ||
| Combined Solicitation Synopsis Final.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
RFQ 1333ND23QNB730551 Combustion Ion Chromatography System- Questions and Answers
(1) Req 1.3 for autosampler to store 25 solid and 100 liquid samples simultaneously: both solids and liquids together at the same time?
Answer: Solid and liquid samples would be analyzed separately.
(2) Same 1.3: store all liquid and solid samples in autosampler racks fitted onto autosampler or in racks ready to load on the bench acceptable?
Answer: Racks ready to load
(3) Req. 1.4: do you mean the sensor system must adjust flow and temp of combustion conditions on the fly or per each instrument method?
Answer: Per each instrument method