SPECIFICATIONS

The USDA, ARS, WBSC intends to procure an HPLC system with a flow-through Radio Detector (Agilent 1260 Infinity II LC System with LabLogic Beta-RAM model 5B Radio Flow Detector for HPLC or equivalent) for the Crop Production Systems Research Unit in Stoneville, MS. The HPLC/Radio Detector system will be used to analyze the fate of herbicides in crops and weeds, and the environment (soil and water) in research projects related to weed and herbicide resistant management.

The following item with noted capabilities is required for procurement:

Quantity Description Instrument

1 Flow-through Radio detector must have following features:

Standard HPLC, UPLC, UHPLC, Rapid Resolution LC, Fast LC and

Micro-Flow Methods Micro/Macro Liquid Scintillator Pump Automatic Flow Cell Identification Leak Detection System High Efficiency 1.25” Photomultiplier Tubes Chemiluminescence and Cross-Talk Correction Two Channel Digital USB Connection Liquid

1 Mixing Flow Cell 500μL Volume 1 Flow-Logic U 4X4L Bottles

Acquisition Software 1 Instrument Control and Data Analysis Radio-Chromatography Software

Acquisition Software Modules 1 HPLC System Control Module w/regression analysis

HPLC System 1 Quaternary Pump VL, maximum pressure 400bar. Includes quaternary pump with integrated 4- channel degassing unit, column, connecting capillaries, solvent cabinet, solvent bottles and CAN cable.

1 HPLC System Tool 1 Max Uptime Kit. Includes stainless steel capillaries, PEEK tubing and fittings, cutter, PTFE frits, solvent inlet filter 1 Active Seal Wash 1 Active Inlet Valve (AIV) 1 Poroshell 120 EC-C18 4.6x150mm, 4μm 1 Vialsampler for use up to 600 bar. Includes 100 uL metering device and a 100 uL sample loop plus integrated needle flush port. Includes 2 drawers each for 66 vials (2 mL).

1 Standard drawer (6x11 vials) 1 Multiple draw option. Includes loop capillaries (400 uL and 1400 uL) to inject up to 500 uL or 1500 uL volume

1 Variable Wavelength Detector. For fast programmable single (up to 120 Hz) and dual wavelength detection. RFID tracking technology for flow cells and UV lamp. Flow cell must be ordered as an option.

1 Standard flow cell for VWD. Includes stainless steel flow cell with RFID tag for identification, 10 mm path length, 14 microliter cell volume.

1 Fraction Collector analytical scale. For flow rates up to 10 ml/min. Allows collection in containers up to 48 mm in height, with semi-prep needle up to 75mm height. Includes delay sensor. No tray included.

Support and Services 1 Installation and training